K033737510(k) TraditionalCleared Dec 9, 200311-day reviewClass II21 CFR 892.1650Recalls 54 (13 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Dec 9, 2003 as substantially equivalent to K002016 (Philips Medical Systems North America, Inc., 2000) after an 11-day review. Its family tree reaches back 1 generation to 2000. 6 of the 28 devices in its family, counting itself, have a Class I recall. 26 later devices descend from it. It has been recalled 54 times, including 13 Class I recalls.

Decided
Dec 9, 2003
Review
11 days
Class
Class II
Product code
OWB
Recalls
54 (13 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K002016 (2000). 26 later devices descend from it, 5 citing it directly. 6 of the 28 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2013: 4 descendants cleared, 1 with a Class I recall12010: 1 descendant cleared2014: 4 descendants cleared2021: 1 descendant cleared2016: 5 descendants cleared, 1 with a Class I recall12018: 2 descendants cleared2025: 3 descendants cleared2017: 2 descendants cleared, 2 with a Class I recall22022: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall12026: 2 descendants clearedK002016 · Philips Integris Allura · Philips Medical Systems North America, Inc. · 2000 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 5 descendants, 1 with a Class I recall12 generations forward: 5 descendants3 generations forward: 3 descendants4 generations forward: 3 descendants5 generations forward: 2 descendants, 2 with a Class I recall26 generations forward: 2 descendants, 1 with a Class I recall17 generations forward: 2 descendants, 1 with a Class I recall18 generations forward: 2 descendants9 generations forward: 2 descendants K002016 · Philips Integris Allura · Philips Medical Systems North America, Inc. · 2000 · recalled0+5
  1. Root · 1 generation backK002016 Philips Integris AlluraPhilips Medical Systems North America, Inc. · 2000 · Recalled
  2. This deviceK033737 Allura Xper FD20Philips Medical Systems North America, Inc. · 2003
  3. Descendant · cites it directly · 24 of its ownK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  4. Descendant · cites it directly · 18 of its ownK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · Recalled
  5. Descendant · 2 generations on · 15 of its ownK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
  6. Descendant · 3 generations on · 13 of its ownK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · Recalled
  7. Descendant · 4 generations on · 8 of its ownK162148 Allura Xper R9Philips Medical Systems Nethrlands BV · 2016 · Recalled

Showing 5 of 26 descendants. See all descendants

Descendants cleared per year

2010: 12013: 42014: 42016: 52017: 22018: 22020: 12021: 12022: 12025: 32026: 2
2010high 5 (2016) · low 0 (2011)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 26 · See all
  • K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · cites it directly24descendants
  • K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · cites it directly18descendants
  • K133292Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · 2 generations on15descendants
  • K141979Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · 3 generations on13descendants
  • K162148Allura Xper R9Philips Medical Systems Nethrlands BV · 2016 · 4 generations on8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K002016Philips Integris AlluraPhilips Medical Systems North America, Inc.200013
0 (this device)K033737Allura Xper FD20Philips Medical Systems North America, Inc.2003—
+1K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201039
+1K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V.201339
+2K133292Allura Xper FD; or TablePhilips Medical Systems Nederland B.V.201445
+3K141979Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV201481
+4K162148Allura Xper R9Philips Medical Systems Nethrlands BV20163

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201341
201440
201651
201722
201820
202011
202110
202210
202530
202620
FIG. 1 — Lineage of K033737, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002016Philips Integris Allura
Summary page 1
…equivalent to the Integris Allura, both manufactured by Philips Medical Systems. The Philips Integris Allura received a 510(k) substantially equivalent determination in K002016 on September 6, 2000. Device description: The Allura Xper FD20 is a stationary fluoroscopic angiographic X-ray system for vascul…
Philips Medical Systems North America, Inc.200013

Also cleared against K002016: 2 other devices →

What descends from it? (26)

See all 26

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201024
K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V.201318
K1321472D PerfusionPhilips Medical Systems Nederland B.V.20132
K131263XperguidePhilips Medical Systems Nederland B.V.20131
K130842Allura Xper FD or Table Series; Allura Xper FD10 Series; Allura Xper FD20 Series; Allura… Class IPhilips Medical Systems Nederland B.V.20130

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 13, 2025Class IZ-1160-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1144-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1159-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1151-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1150-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1148-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1154-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1158-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1147-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1142-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1143-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1149-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1145-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IIZ-1080-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 2, 2024Class IIZ-1144-2024Process designFramegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may…
Feb 19, 2019Class IIZ-1141-2019Component design/selectionAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or…
Feb 19, 2019Class IIZ-1140-2019Component design/selectionAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or…
Jul 16, 2018Class IIZ-3222-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3224-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3221-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3223-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Mar 22, 2017Class IIZ-1822-2017Under Investigation by firmDue to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to…
Dec 9, 2016Class IIZ-1061-2017Device DesignSystem may lock up.
Dec 9, 2016Class IIZ-1062-2017Device DesignSystem may lock up.
Dec 1, 2016Class IIZ-0922-2017Process controlDue to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
Nov 3, 2015Class IIZ-0764-2016Device DesignNormally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped…
Jul 6, 2015Class IIZ-0810-2016Process controlThe actuator assembly became detached and the monitor carriage with a flexvision monitor dropped to its lowest position.
Feb 3, 2015Class IIZ-0346-2016Process controlMonitor Ceiling Suspension system may fall
Feb 3, 2015Class IIZ-0349-2016Process controlMonitor Ceiling Suspension system may fall
Feb 3, 2015Class IIZ-0354-2016Process controlMonitor Ceiling Suspension system may fall
Feb 3, 2015Class IIZ-0353-2016Process controlMonitor Ceiling Suspension system may fall
Feb 3, 2015Class IIZ-0348-2016Process controlMonitor Ceiling Suspension system may fall
Feb 3, 2015Class IIZ-0350-2016Process controlMonitor Ceiling Suspension system may fall
Feb 3, 2015Class IIZ-0351-2016Process controlMonitor Ceiling Suspension system may fall
Feb 3, 2015Class IIZ-0352-2016Process controlMonitor Ceiling Suspension system may fall
Feb 3, 2015Class IIZ-0347-2016Process controlMonitor Ceiling Suspension system may fall
Sep 24, 2014Class IIZ-1280-2016Device DesignLoss of key image functionality due to a bent pedal of the Footswitch.
Jul 8, 2014Class IIZ-0181-2015Software designFaulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.
Jul 8, 2014Class IIZ-0182-2015Software designFaulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.
Apr 2, 2012Not classifiedZ-1462-2012Process controlThe monitor Ceiling suspension may fail, which may cause the monitor to drop.
Jul 8, 2011Not classifiedZ-2976-2011Software designSystem freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.
Nov 4, 2010Not classifiedZ-1368-2011PendingAdjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch it does not release…
Feb 4, 2010Not classifiedZ-2448-2010Software designA lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem could only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator…
Feb 4, 2010Not classifiedZ-2449-2010Software designA lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator…
Dec 10, 2008Not classifiedZ-0348-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Dec 10, 2008Not classifiedZ-0349-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Oct 23, 2008Not classifiedZ-2475-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Oct 23, 2008Not classifiedZ-2476-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Sep 24, 2007Not classifiedZ-0329-2008Device DesignInaccurate anatomical measurements when using the optional SmartMask feature
Sep 24, 2007Not classifiedZ-0330-2008Device DesignInaccurate anatomical measurements when using the optional SmartMask feature

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K130842Allura Xper FD or Table Series; Allura Xper FD10 Series; Allura Xper FD20 Series; Allura Xper FD Biplane…Philips Medical Systems Nederland B.V. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 111 Class I recalls
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 7 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K102005 (Philips Medical Systems North America, Inc., 2010) has gone furthest, with 24 descendants of its own.
  • 2 other devices also cite its predicate K002016.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Allura Xper FD20 is intended for:

o Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis.

o Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomiss), pacemaker implantations, and electrophysiology (EP).

» Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K033737: Allura Xper FD20.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033737

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026