510(k) · Cleared
Recalled · 2006Knightstar 330 Ventilator
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Mallinckrodt, Inc. · MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator
Cleared on Dec 11, 2001 after a 434-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 37 of the 130 devices in its family, counting itself, have a Class I recall. 129 later devices descend from it. It has been recalled once.
- Decided
- Dec 11, 2001
- Review
- 434 days
- Class
- Class II
- Product code
- MNT
- Recalls
- 1
- Predicates
- 0
129 later devices descend from it, 3 citing it directly. 37 of the 130 devices in this family, counting this one, have a Class I recall.
- This deviceK003075 Knightstar 330 VentilatorMallinckrodt, Inc. · 2001
- Descendant · cites it directly · 102 of its ownK022132 Servo Ventilator System, Model 64 87 800 E407ESiemens Medical Solutions USA, Inc. · 2002
- Descendant · 2 generations on · 75 of its ownK040221 Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Recalled
- Descendant · 3 generations on · 73 of its ownK041223 Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Recalled (Class I)
- Descendant · 4 generations on · 61 of its ownK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
- Descendant · 2 generations on · 53 of its ownK040574 Galileo GoldHamilton Medical, Inc. · 2004
Showing 5 of 129 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 129 · See all- K022132Servo Ventilator System, Model 64 87 800 E407ESiemens Medical Solutions USA, Inc. · 2002 · cites it directly102descendants
- K040221Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 2 generations on75descendants
- K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 3 generations on73descendants
- K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007 · 4 generations on61descendants
- K040574Galileo GoldHamilton Medical, Inc. · 2004 · 2 generations on53descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K003075 | Knightstar 330 Ventilator | Mallinckrodt, Inc. | 2001 | — |
| +1 | K022132 | Servo Ventilator System, Model 64 87 800 E407E | Siemens Medical Solutions USA, Inc. | 2002 | — |
| +2 | K040221 | Servo-I Ventilator System, Model 64 87 800 E407E | Maquet Critical Care AB | 2004 | 1 |
| +3 | K041223 | Modification to Servo-I Ventilator System, Model 64 87 800 E407E | Maquet Critical Care AB | 2004 | Class I |
| +4 | K063404 | Servo-I Ventilator System, Models 64, 87 and 800 | Maquet Critical Care AB | 2007 | — |
| +2 | K040574 | Galileo Gold | Hamilton Medical, Inc. | 2004 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2002 | 1 | 0 |
| 2004 | 7 | 2 |
| 2005 | 2 | 0 |
| 2006 | 5 | 1 |
| 2007 | 4 | 0 |
| 2008 | 10 | 2 |
| 2009 | 5 | 1 |
| 2010 | 5 | 1 |
| 2011 | 11 | 6 |
| 2012 | 12 | 4 |
| 2013 | 8 | 4 |
| 2014 | 3 | 1 |
| 2015 | 6 | 1 |
| 2016 | 4 | 2 |
| 2017 | 7 | 3 |
| 2018 | 1 | 0 |
| 2019 | 6 | 2 |
| 2020 | 4 | 3 |
| 2021 | 7 | 3 |
| 2022 | 1 | 0 |
| 2023 | 4 | 1 |
| 2024 | 5 | 0 |
| 2025 | 6 | 0 |
| 2026 | 5 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
37 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MNT device
- #1 K233643 F&P Airvo 3 NIV (PT311US)score 13.5 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (129)
See all 1293 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (37 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 31, 2006 | Not classified | Z-0444-2007 | Other | Directions for Use-Puritan Bennett has determined that a service interval should be defined for the device. The firm has also determined that an exhalation filter should be employed when the device is in use. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K034040Esprit Ventilator, Model V1000Respironics, Inc. · 2004 · 2 generations on · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
- K073069Modification to Avea VentilatorViasys Respiratory Care, Inc. · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 113 Class I recalls
- K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 3 generations on · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
- K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 4 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
- K053388Model 840 Ventilator with Proportional Assist Ventilation Plus Software OptionPuritan Bennett Corp. · 2006 · 4 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
- K093632Infinity Acute Care System Workstation Neonatal CareDrager Medical AG & Co. Kgaa · 2010 · 4 generations on · ◆ in the 2010 descendants bar of FIG. 14 Class I recalls
- K111116Engstrom Carestation, Engstrom Pro, Models 1505-9000-000, 1505-9003-000Datex-Ohmeda · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 13 Class I recalls
- K103803Hamilton-G5Hamilton Medical AG · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
- K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013 · 5 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
- K121225Hamilton-C2Hamilton Medical AG · 2012 · 5 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
- K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
- K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
Nearest 12 of 37 shown.
Nearest 5 of 37; all are on the web page.
What does this lineage say?
- Of its descendants, K022132 (Siemens Medical Solutions USA, Inc., 2002) has gone furthest, with 102 descendants of its own.
How does its category compare?
All MNT devicesMNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MNT
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MNT
Too few MNT 510(k)s in any year since 2007 to give a median.
Recalls per year · MNT
5in 2022
Class I (6 since 2007) Class II/III not classified
Adverse event reports per year · MNT
3,545in 2025
MAUDE reports naming product code MNT, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 1 | 59 |
| 2008 | 1 | — | 0 | 76 |
| 2009 | 2 | — | 1 | 29 |
| 2010 | 0 | — | 1 | 63 |
| 2011 | 2 | — | 0 | 170 |
| 2012 | 2 | — | 0 | 257 |
| 2013 | 1 | — | 0 | 322 |
| 2014 | 0 | — | 0 | 887 |
| 2015 | 0 | — | 0 | 1,657 |
| 2016 | 0 | — | 0 | 2,515 |
| 2017 | 1 | — | 0 | 2,878 |
| 2018 | 0 | — | 0 | 2,203 |
| 2019 | 0 | — | 1 (1) | 9,350 |
| 2020 | 1 | — | 33 | 4,170 |
| 2021 | 0 | — | 0 | 3,790 |
| 2022 | 1 | — | 5 (5) | 3,277 |
| 2023 | 1 | — | 0 | 4,565 |
| 2024 | 2 | — | 0 | 4,808 |
| 2025 | 1 | — | 0 | 3,545 |
| 2026 | 1 | — | 0 | — |
https://510kreport.com/device/K003075 · printed Nov 6, 2026