510(k) Report

510(k) · Cleared

Recalled · 2006

Knightstar 330 Ventilator

K003075510(k) TraditionalCleared Dec 11, 2001434-day reviewClass II21 CFR 868.5895Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Mallinckrodt, Inc. · MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Cleared on Dec 11, 2001 after a 434-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 37 of the 130 devices in its family, counting itself, have a Class I recall. 129 later devices descend from it. It has been recalled once.

Decided
Dec 11, 2001
Review
434 days
Class
Class II
Product code
MNT
Recalls
1
Predicates
0

129 later devices descend from it, 3 citing it directly. 37 of the 130 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2004: 7 descendants cleared, 2 with a Class I recall22006: 5 descendants cleared, 1 with a Class I recall12002: 1 descendant cleared2010: 5 descendants cleared, 1 with a Class I recall12008: 10 descendants cleared, 2 with a Class I recall22005: 2 descendants cleared2007: 4 descendants cleared2009: 5 descendants cleared, 1 with a Class I recall12011: 11 descendants cleared, 6 with a Class I recall62014: 3 descendants cleared, 1 with a Class I recall12024: 5 descendants cleared2021: 7 descendants cleared, 3 with a Class I recall32012: 12 descendants cleared, 4 with a Class I recall42013: 8 descendants cleared, 4 with a Class I recall42023: 4 descendants cleared, 1 with a Class I recall12019: 6 descendants cleared, 2 with a Class I recall22015: 6 descendants cleared, 1 with a Class I recall12017: 7 descendants cleared, 3 with a Class I recall32025: 6 descendants cleared2026: 5 descendants cleared2016: 4 descendants cleared, 2 with a Class I recall22022: 1 descendant cleared2020: 4 descendants cleared, 3 with a Class I recall32018: 1 descendant cleared200020052010201520202025
  1. This deviceK003075 Knightstar 330 VentilatorMallinckrodt, Inc. · 2001
  2. Descendant · cites it directly · 102 of its ownK022132 Servo Ventilator System, Model 64 87 800 E407ESiemens Medical Solutions USA, Inc. · 2002
  3. Descendant · 2 generations on · 75 of its ownK040221 Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Recalled
  4. Descendant · 3 generations on · 73 of its ownK041223 Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Recalled (Class I)
  5. Descendant · 4 generations on · 61 of its ownK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
  6. Descendant · 2 generations on · 53 of its ownK040574 Galileo GoldHamilton Medical, Inc. · 2004

Showing 5 of 129 descendants. See all descendants

Descendants cleared per year

2002: 12004: 72005: 22006: 52007: 42008: 102009: 52010: 52011: 112012: 122013: 82014: 32015: 62016: 42017: 72018: 12019: 62020: 42021: 72022: 12023: 42024: 52025: 62026: 5
2002high 12 (2012) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 129 · See all
  • K022132Servo Ventilator System, Model 64 87 800 E407ESiemens Medical Solutions USA, Inc. · 2002 · cites it directly102descendants
  • K040221Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 2 generations on75descendants
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 3 generations on73descendants
  • K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007 · 4 generations on61descendants
  • K040574Galileo GoldHamilton Medical, Inc. · 2004 · 2 generations on53descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K003075Knightstar 330 VentilatorMallinckrodt, Inc.2001—
+1K022132Servo Ventilator System, Model 64 87 800 E407ESiemens Medical Solutions USA, Inc.2002—
+2K040221Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB20041
+3K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB2004Class I
+4K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB2007—
+2K040574Galileo GoldHamilton Medical, Inc.2004—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200472
200520
200651
200740
2008102
200951
201051
2011116
2012124
201384
201431
201561
201642
201773
201810
201962
202043
202173
202210
202341
202450
202560
202650
FIG. 1 — Lineage of K003075, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

37 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MNT device

  1. #1 K233643 F&P Airvo 3 NIV (PT311US)score 13.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (129)

See all 129

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K022132Servo Ventilator System, Model 64 87 800 E407ESiemens Medical Solutions USA, Inc.2002102
K053607Breas Vivo 40 SystemBreas Medical AB200625
K031896Nasal-Air Critical CareInnomed Technologies20040

Has anything in its family been recalled? (37 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 31, 2006Not classifiedZ-0444-2007OtherDirections for Use-Puritan Bennett has determined that a service interval should be defined for the device. The firm has also determined that an exhalation filter should be employed when the device is in use.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K034040Esprit Ventilator, Model V1000Respironics, Inc. · 2004 · 2 generations on · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
  • K073069Modification to Avea VentilatorViasys Respiratory Care, Inc. · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 113 Class I recalls
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 3 generations on · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 4 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K053388Model 840 Ventilator with Proportional Assist Ventilation Plus Software OptionPuritan Bennett Corp. · 2006 · 4 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K093632Infinity Acute Care System Workstation Neonatal CareDrager Medical AG & Co. Kgaa · 2010 · 4 generations on · ◆ in the 2010 descendants bar of FIG. 14 Class I recalls
  • K111116Engstrom Carestation, Engstrom Pro, Models 1505-9000-000, 1505-9003-000Datex-Ohmeda · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 13 Class I recalls
  • K103803Hamilton-G5Hamilton Medical AG · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013 · 5 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 5 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls

Nearest 12 of 37 shown.

What does this lineage say?

  • Of its descendants, K022132 (Siemens Medical Solutions USA, Inc., 2002) has gone furthest, with 102 descendants of its own.

How does its category compare?

All MNT devices

MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MNT

1in 2025

2008: 12009: 22011: 22012: 22013: 12017: 12020: 12022: 12023: 12024: 22025: 12026 (partial year): 1
2007high 2 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MNT

Too few MNT 510(k)s in any year since 2007 to give a median.

Recalls per year · MNT

5in 2022

2007: 12009: 12010: 12019: 1, 1 Class I2020: 332022: 5, 5 Class I
2007high 33 (2020) · low 0 (2008)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · MNT

3,545in 2025

2007: 592008: 762009: 292010: 632011: 1702012: 2572013: 3222014: 8872015: 1,6572016: 2,5152017: 2,8782018: 2,2032019: 9,3502020: 4,1702021: 3,7902022: 3,2772023: 4,5652024: 4,8082025: 3,545
2007high 9,350 (2019) · low 29 (2009)2025

MAUDE reports naming product code MNT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—159
20081—076
20092—129
20100—163
20112—0170
20122—0257
20131—0322
20140—0887
20150—01,657
20160—02,515
20171—02,878
20180—02,203
20190—1 (1)9,350
20201—334,170
20210—03,790
20221—5 (5)3,277
20231—04,565
20242—04,808
20251—03,545
20261—0—
Cite this page

510(k) Report. “K003075: Knightstar 330 Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K003075

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026