510(k) Report

510(k) · Cleared

Recalled · 2005

Servo-I Ventilator System, Model 64 87 800 E407E

K040221510(k) SpecialCleared Feb 20, 200418-day reviewClass II21 CFR 868.5895Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Maquet Critical Care AB · CBK Ventilator, Continuous, Facility Use

Cleared on Feb 20, 2004 as substantially equivalent to K970839 (Siemens Elema AB, 1997) and 2 other predicates after an 18-day review. Its family tree reaches back 3 generations to 1991. 29 of the 88 devices in its family, counting itself, have a Class I recall. 75 later devices descend from it. It has been recalled once.

Decided
Feb 20, 2004
Review
18 days
Class
Class II
Product code
CBK
Recalls
1
Predicates
3

Its longest line of descent runs back 3 generations to K902859 (1991). 12 ancestors in all. 75 later devices descend from it, 2 citing it directly. 29 of the 88 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor1998: 1 other ancestor2001: 3 other ancestors2000: 2 other ancestors, 1 with a Class I recall11999: 1 other ancestor1997: 1 other ancestor2004: 1 descendant cleared, 1 with a Class I recall12007: 3 descendants cleared2008: 6 descendants cleared, 1 with a Class I recall12014: 3 descendants cleared, 1 with a Class I recall12006: 1 descendant cleared2011: 7 descendants cleared, 6 with a Class I recall62010: 2 descendants cleared2009: 4 descendants cleared, 1 with a Class I recall12012: 6 descendants cleared, 2 with a Class I recall22015: 6 descendants cleared, 1 with a Class I recall12017: 6 descendants cleared, 3 with a Class I recall32013: 6 descendants cleared, 3 with a Class I recall32021: 5 descendants cleared, 2 with a Class I recall22019: 4 descendants cleared, 1 with a Class I recall12026: 3 descendants cleared2024: 3 descendants cleared2016: 2 descendants cleared, 2 with a Class I recall22020: 4 descendants cleared, 3 with a Class I recall32025: 1 descendant cleared2018: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall1K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 6 other ancestors, 1 with a Class I recall13 generations back: 1 other ancestor 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 10 descendants, 1 with a Class I recall13 generations forward: 14 descendants, 6 with a Class I recall64 generations forward: 20 descendants, 8 with a Class I recall85 generations forward: 15 descendants, 6 with a Class I recall66 generations forward: 7 descendants, 4 with a Class I recall47 generations forward: 6 descendants, 2 with a Class I recall28 generations forward: 1 descendant K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalled0+5
  1. Root · 3 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 2 generations backK960010 Servo Ventilator 300 Computer InterfaceSiemens Elema AB · 1996
  3. 1 generation backK970839 Servo Ventilator 300ASiemens Elema AB · 1997 · Recalled
  4. This deviceK040221 Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004
  5. Descendant · cites it directly · 73 of its ownK041223 Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Recalled (Class I)
  6. Descendant · 2 generations on · 61 of its ownK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
  7. Descendant · 2 generations on · 37 of its ownK081842 Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008
  8. Descendant · 3 generations on · 28 of its ownK073179 Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option…Maquet Critical Care AB · 2008
  9. Descendant · 2 generations on · 26 of its ownK070513 Hamilton-G5Hamilton Medical AG · 2007 · Recalled

Showing 5 of 75 descendants. See all descendants

Descendants cleared per year

2004: 12006: 12007: 32008: 62009: 42010: 22011: 72012: 62013: 62014: 32015: 62016: 22017: 62018: 12019: 42020: 42021: 52023: 12024: 32025: 12026: 3
2004high 7 (2011) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 75 · See all
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · cites it directly73descendants
  • K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007 · 2 generations on61descendants
  • K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008 · 2 generations on37descendants
  • K073179Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option, Model 6675585Maquet Critical Care AB · 2008 · 3 generations on28descendants
  • K070513Hamilton-G5Hamilton Medical AG · 2007 · 2 generations on26descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K902859Servo Ventilator 300Siemens Life Support Systems19913
−2K960010Servo Ventilator 300 Computer InterfaceSiemens Elema AB1996—
−1K970839Servo Ventilator 300ASiemens Elema AB19974
0 (this device)K040221Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB2004—
+1K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB2004Class I
+2K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB2007—
+2K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc.2008—
+3K073179Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option, Model 6675585Maquet Critical Care AB2008—
+2K070513Hamilton-G5Hamilton Medical AG20071

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199810
199910
200021
200130
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200411
200610
200730
200861
200941
201020
201176
201262
201363
201431
201561
201622
201763
201810
201941
202043
202152
202311
202430
202510
202630
FIG. 1 — Lineage of K040221, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

29 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K970839Servo Ventilator 300A
Summary page 1
…s to which equivalence is being claimed 510(k) # Servo Ventilator 300 A, K970839 Servo-i Ventilatory System K022132 Evita 4,…
Siemens Elema AB19974
*K022132Servo Ventilator System, Model 64 87 800 E407E
Summary page 1
…, K970839 Servo-i Ventilatory System K022132 Evita 4, K980642 Maquet Critical Care AB…
Siemens Medical Solutions USA, Inc.2002—
*K980642EVITA 4 (8411900)
Summary page 1
…stem K022132 Evita 4, K980642 Maquet Critical Care AB MM 0053-01 special 510k servoi lmv…
Drager, Inc.1998—

Also cleared against K970839: 7 other devices →

What descends from it? (75)

See all 75

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407E Class IMaquet Critical Care AB200473
K062793Servo-I Ventilator System, Model 64 87 800 E407E; Nava OptionMaquet Critical Care AB20070

Has anything in its family been recalled? (29 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 24, 2005Not classifiedZ-0878-05OtherComplaints about batteries indicating a satisfying remaining time but after a short time suddenly shut off without giving alarms to indicate short remaining battery back-up time. (Power off alarm will be given).

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 2 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K111116Engstrom Carestation, Engstrom Pro, Models 1505-9000-000, 1505-9003-000Datex-Ohmeda · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 13 Class I recalls
  • K103803Hamilton-G5Hamilton Medical AG · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K142679Carescape R860Datex-Ohmeda, Inc. · 2015 · 3 generations on · ◆ in the 2015 descendants bar of FIG. 13 Class I recalls
  • K131774Hamilton-G5Hamilton Medical AG · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · 4 generations on · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls

Nearest 12 of 29 shown.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 5.5 years on median.
  • 5 of the 12 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K041223 (Maquet Critical Care AB, 2004) has gone furthest, with 73 descendants of its own.
  • 7 other devices also cite its predicate K970839.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Servo-i Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency. Servo-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities and for in-hospital transport.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040221: Servo-I Ventilator System, Model 64 87 800 E407E.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K040221

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026