510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Ohmeda Medical Giraffe Incubator

K010222510(k) TraditionalCleared Mar 21, 200156-day reviewClass II21 CFR 880.5400Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ohmeda Medical · FMZ Incubator, Neonatal

Cleared on Mar 21, 2001 after a 56-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 of the 11 devices in its family, counting itself, have a Class I recall. 10 later devices descend from it. It has been recalled 3 times.

Decided
Mar 21, 2001
Review
56 days
Class
Class II
Product code
FMZ
Recalls
3
Predicates
0

10 later devices descend from it, 3 citing it directly. 2 of the 11 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2015: 1 descendant cleared, 1 with a Class I recall12017: 1 descendant cleared2002: 1 descendant cleared2007: 1 descendant cleared2022: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared2018: 1 descendant cleared2026: 2 descendants cleared2025: 1 descendant cleared200020052010201520202025
  1. This deviceK010222 Ohmeda Medical Giraffe IncubatorOhmeda Medical · 2001
  2. Descendant · cites it directly · 5 of its ownK020547 Ohmeda Medical Giraffe IncubatorOhmeda Medical · 2002 · Recalled
  3. Descendant · 2 generations on · 4 of its ownK072512 Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda…Datex-Ohmeda, Inc. · 2007
  4. Descendant · cites it directly · 3 of its ownK162821 Babyleo TN500Draegerwerk AG & CO Kgaa · 2017
  5. Descendant · cites it directly · 3 of its ownK152809 Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · Recalled (Class I)
  6. Descendant · 2 generations on · 2 of its ownK213551 Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc. · 2022 · Recalled (Class I)

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2002: 12007: 12015: 12017: 12018: 12019: 12022: 12025: 12026: 2
2002high 2 (2026) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 10 · See all
  • K020547Ohmeda Medical Giraffe IncubatorOhmeda Medical · 2002 · cites it directly5descendants
  • K072512Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda Uninterruptible…Datex-Ohmeda, Inc. · 2007 · 2 generations on4descendants
  • K162821Babyleo TN500Draegerwerk AG & CO Kgaa · 2017 · cites it directly3descendants
  • K152809Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · cites it directly3descendants
  • K213551Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K010222Ohmeda Medical Giraffe IncubatorOhmeda Medical2001—
+1K020547Ohmeda Medical Giraffe IncubatorOhmeda Medical20024
+2K072512Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda…Datex-Ohmeda, Inc.2007—
+1K162821Babyleo TN500Draegerwerk AG & CO Kgaa2017—
+1K152809Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc.2015Class I
+2K213551Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc.2022Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200710
201511
201710
201810
201910
202211
202510
202620
FIG. 1 — Lineage of K010222, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMZ device

  1. #1 K251663 Giraffe OmniBed Carestation (CS1); Giraffe…score 14.4 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (10)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K020547Ohmeda Medical Giraffe IncubatorOhmeda Medical20025
K162821Babyleo TN500Draegerwerk AG & CO Kgaa20173
K152809Giraffe Incubator Carestation CS1 Class IOhmeda Medical, A Division of Datex-Ohmeda, Inc.20153

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 28, 2020Class IIZ-2236-2020Device DesignPotential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
Nov 18, 2016Class IIZ-1502-2017Under Investigation by firmCertain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where…
Apr 25, 2005Not classifiedZ-1047-05OtherNeonatal incubators and beds may unintentionally move from a safe position following an interruption in the power supply.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K152809Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K213551Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on · ◆ in the 2022 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K020547 (Ohmeda Medical, 2002) has gone furthest, with 5 descendants of its own.

How does its category compare?

All FMZ devices

FMZ Incubator, Neonatal: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FMZ

2in 2025

2007: 32008: 12010: 42011: 12014: 12015: 12016: 12017: 12018: 22019: 22022: 12025: 22026 (partial year): 2
2007high 4 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMZ

60in 2010

2007: 81 days median2010: 60 days median
2007high 81 (2007) · low 60 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMZ

2in 2024

2008: 42011: 52013: 1, 1 Class I2016: 12018: 32019: 12023: 8, 8 Class I2024: 2, 2 Class I2026 (partial year): 3
2007high 8 (2023) · low 0 (2007)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · FMZ

250in 2025

2007: 142008: 102009: 82010: 112011: 102012: 122013: 112014: 32015: 92016: 122017: 42018: 182019: 242020: 82021: 152022: 272023: 192024: 1182025: 250
2007high 250 (2025) · low 3 (2014)2025

MAUDE reports naming product code FMZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007381014
20081—410
20090—08
2010460011
20111—510
20120—012
20130—1 (1)11
20141—03
20151—09
20161—112
20171—04
20182—318
20192—124
20200—08
20210—015
20221—027
20230—8 (8)19
20240—2 (2)118
20252—0250
20262—3—
Cite this page

510(k) Report. “K010222: Ohmeda Medical Giraffe Incubator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K010222

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026