Alaris System with Guardrails Suite MX

K133532510(k) TraditionalCleared Aug 21, 2014276-day reviewClass II21 CFR 880.5725Recalls 35 (28 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carefusion 303, Inc. · FRN Pump, Infusion

Cleared on Aug 21, 2014 as substantially equivalent to K051641 (Cardinal Health, Alaris Products, 2005) and 6 other predicates after a 276-day review. Its family tree reaches back 3 generations to 1994. 7 of the 20 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 35 times, including 28 Class I recalls.

Decided
Aug 21, 2014
Review
276 days
Class
Class II
Product code
FRN
Recalls
35 (28 Class I)
Predicates
7

Its longest line of descent runs back 3 generations to K940147 (1994). 16 ancestors in all, the oldest K890120 (1989). 3 later devices descend from it, 2 citing it directly. 7 of the 20 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2007: 1 other ancestor, 1 with a Class I recall12005: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor2001: 2 other ancestors, 1 with a Class I recall11995: 2 other ancestors, 1 with a Class I recall12000: 1 other ancestor1999: 1 other ancestor1996: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor2022: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant clearedK940147 · Auto Syringe AS40A Infusion Pump · Baxter Corp. · 1994 · recalledK974332 · Variable Pressure Syringe Pump · Alaris Medical Systems, Inc. · 1998K023264 · Medley Syringe Pump Module System, Model 8110 · Alaris Medical Systems, Inc. · 2002 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 6 other ancestors, 4 with a Class I recall42 generations back: 3 other ancestors3 generations back: 4 other ancestors 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 1 descendant K023264 · Medley Syringe Pump Module System, Model 8110 · Alaris Medical Systems, Inc. · 2002 · Class I recallK974332 · Variable Pressure Syringe Pump · Alaris Medical Systems, Inc. · 1998K940147 · Auto Syringe AS40A Infusion Pump · Baxter Corp. · 1994 · recalled0
  1. Root · 3 generations backK940147 Auto Syringe AS40A Infusion PumpBaxter Corp. · 1994 · Recalled
  2. 2 generations backK974332 Variable Pressure Syringe PumpAlaris Medical Systems, Inc. · 1998
  3. 1 generation backK023264 Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Recalled (Class I)
  4. This deviceK133532 Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014
  5. Descendant · cites it directly · 1 of its ownK211218 BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · Recalled (Class I)
  6. Descendant · 2 generations on · 0 of its ownK243855 BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc. · 2025
  7. Descendant · cites it directly · 0 of its ownK210074 Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia…Fresenius Kabi AG · 2022

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · cites it directly1descendant
  • K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc. · 2025 · 2 generations on0descendants
  • K210074Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link, Agilia Duo…Fresenius Kabi AG · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K940147Auto Syringe AS40A Infusion PumpBaxter Corp.19941
−2K974332Variable Pressure Syringe PumpAlaris Medical Systems, Inc.1998—
−1K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc.2002Class I
0 (this device)K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc.2014—
+1K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc.2023Class I
+2K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc.2025—
+1K210074Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link…Fresenius Kabi AG2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199010
199521
199610
199910
200010
200121
200310
200511
200711
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
202311
202510
FIG. 1 — Lineage of K133532, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (7)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051641Alaris System PC Unit, Model 8001
Summary page 4
…umber Number Code, Regulation Alaris System Components Point-of-Care (PC) Unit Module 8015 80F--RN K051641* K091308 Large Volume Pump (LVP) Module 8100 21 CFR 880.5725 K950419* K012383…
Cardinal Health, Alaris Products2005Class I
*K091308Alaris PC Unit, Models 8000 and 8015
Summary page 4
…Number Code, Regulation Alaris System Components Point-of-Care (PC) Unit Module 8015 80F--RN K051641* K091308 Large Volume Pump (LVP) Module 8100 21 CFR 880.5725 K950419* K012383 Syringe Pump (SYR) Module…
Cardinal Health 303,Inc.2009—
*K950419The Imed Orion Infusion Pump and Administration Sets
Summary page 4
…-of-Care (PC) Unit Module 8015 80F--RN K051641* K091308 Large Volume Pump (LVP) Module 8100 21 CFR 880.5725 K950419* K012383 Syringe Pump (SYR) Module 8110 K023264 N/A Guar…
Imed Corp.1995Class I
*K012383Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem Iii…
Summary page 4
…8015 80F--RN K051641* K091308 Large Volume Pump (LVP) Module 8100 21 CFR 880.5725 K950419* K012383 Syringe Pump (SYR) Module 8110 K023264 N/A Guardrails Suite MX Component…
Alaris Medical Systems, Inc.2001Class I
*K023264Medley Syringe Pump Module System, Model 8110
Summary page 4
…Volume Pump (LVP) Module 8100 21 CFR 880.5725 K950419* K012383 Syringe Pump (SYR) Module 8110 K023264 N/A Guardrails Suite MX Components Alaris Guardrails Editor (GRE) 8961 80F--RN K07210…
Alaris Medical Systems, Inc.2002Class I
*K072105Alaris System with MMS, Model 8XXX
Summary page 4
…8110 K023264 N/A Guardrails Suite MX Components Alaris Guardrails Editor (GRE) 8961 80F--RN K072105* K030459 safety software 21 CFR 880.5725 *Primary…
Cardinal Health 303,Inc.2007Class I
*K030459Medley System with Medication Management System (MMS)
Summary page 4
…K023264 N/A Guardrails Suite MX Components Alaris Guardrails Editor (GRE) 8961 80F--RN K072105* K030459 safety software 21 CFR 880.5725 *Primary…
Alaris Medical Systems, Inc.2003—

Also cleared against K051641: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211218BD Alaris System with Guardrails Suite MX v12.1.2 Class ICarefusion 303, Inc.20231
K210074Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link, Agilia Duo…Fresenius Kabi AG20220

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 11, 2026Class IIZ-1645-2026Under Investigation by firmDue to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units
Sep 15, 2023Class IZ-0185-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Sep 15, 2023Class IZ-0184-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Sep 15, 2023Class IZ-0186-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Mar 26, 2021Class IZ-1544-2021Component design/selectionBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may…
Mar 25, 2021Class IZ-1651-2021Under Investigation by firmBezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free…
Mar 24, 2021Class IZ-1375-2021Component design/selectionAftermarket front bezel components were installed during service/repair, not by the original manufacturer, within infusion pump modules. Over time, the posts on the front bezel component may crack or separate. Separation of one or more…
Mar 3, 2021Class IZ-1309-2021Process designPump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in…
Feb 26, 2021Class IZ-1543-2021Component design/selectionBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may…
Feb 25, 2021Class IZ-2204-2021Component design/selectionBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may…
Feb 25, 2021Class IZ-1331-2021Component design/selectionBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may…
Feb 25, 2021Class IZ-1332-2021Component design/selectionBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may…
Aug 4, 2020Class IZ-2939-2020Device DesignPump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious…
Aug 4, 2020Class IZ-2909-2020Reprocessing ControlsThe pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk…
Aug 4, 2020Class IZ-2910-2020Reprocessing ControlsThe pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk…
Aug 4, 2020Class IZ-2882-2020Component design/selectionThe Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion.
Aug 4, 2020Class IIZ-2935-2020Vendor change controlInfusion pump component defect may result in interruption of patient monitoring
Jun 30, 2020Class IZ-2717-2020Device DesignCleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an…
Jun 30, 2020Class IZ-2716-2020Device DesignCleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an…
Jun 30, 2020Class IZ-2722-2020Device DesignCleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an…
Jun 30, 2020Class IZ-2736-2020Device DesignBroken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane…
Jun 30, 2020Class IZ-2741-2020Nonconforming Material/ComponentIf one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power…
Jun 30, 2020Class IZ-2740-2020Nonconforming Material/ComponentIf one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power…
Jun 30, 2020Class IZ-2721-2020Device DesignCleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an…
Jun 30, 2020Class IZ-2720-2020Device DesignCleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an…
Jun 30, 2020Class IZ-2718-2020Device DesignCleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an…
Jun 30, 2020Class IZ-2719-2020Device DesignCleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an…
Jun 30, 2020Class IZ-2723-2020Device DesignCleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an…
Jun 30, 2020Class IIZ-2822-2020Device DesignLED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.
Jun 30, 2020Class IIZ-2824-2020Device DesignLED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.
Jun 30, 2020Class IIZ-2823-2020Device DesignLED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.
Jun 30, 2020Class IIZ-2825-2020Device DesignLED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.
Jun 30, 2020Class IIZ-2826-2020Device DesignLED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.
Feb 4, 2020Class IZ-1359-2020Device Design1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion…
Apr 13, 2018Class IZ-1768-2019Component design/selectionThe bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e.…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K072105Alaris System with MMS, Model 8XXXCardinal Health 303,Inc. · 2007 · Direct predicate · ◆ in the 2007 ancestors bar of FIG. 13 Class I recalls
  • K051641Alaris System PC Unit, Model 8001Cardinal Health, Alaris Products · 2005 · Direct predicate · ◆ in the 2005 ancestors bar of FIG. 110 Class I recalls
  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K012383Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem Iii Infusion (Msiii)…Alaris Medical Systems, Inc. · 2001 · Direct predicate · ◆ in the 2001 ancestors bar of FIG. 13 Class I recalls
  • K950419The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · Direct predicate · ◆ in the 1995 ancestors bar of FIG. 14 Class I recalls
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · Cites it directly · ◆ in the 2023 descendants bar of FIG. 17 Class I recalls

What does this lineage say?

  • The line of descent spans 20 years and passes through 3 companies; a new generation was cleared every 4.7 years on median.
  • 10 of the 16 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K211218 (Carefusion 303, Inc., 2023) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K051641.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Alaris System with Guardrails Suite MX is intended for use in professional healthcare facilities that utilize infusion devices for the delivery of fluids, medications, blood and blood products. The Alaris System with Guardrails Suite MX is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and…

Read the full statement (137 words)

The Alaris System with Guardrails Suite MX is intended for use in professional healthcare facilities that utilize infusion devices for the delivery of fluids, medications, blood and blood products.

The Alaris System with Guardrails Suite MX is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by the trained healthcare professional according to a physician's order.

The Alaris System with Guardrails Suite MX is an interoperable system capable of communicating and exchanging data accurately, effectively, securely, and consistently with different information technology systems, sooftware applications, and networks, in various settings; and exchanging data such that the clinical or operational purpose and meaning of the data are preserved and unaltered.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133532: Alaris System with Guardrails Suite MX.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K133532

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026