510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

BD Alaris System with Guardrails Suite MX v12.1.2

K211218510(k) TraditionalCleared Jul 21, 2023819-day reviewClass II21 CFR 880.5725Recalls 16 (7 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carefusion 303, Inc. · FRN Pump, Infusion

Cleared on Jul 21, 2023 as substantially equivalent to K133532 (Carefusion 303, Inc., 2014) after an 819-day review. Its family tree reaches back 4 generations to 1994. 7 of the 19 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 16 times, including 7 Class I recalls.

Decided
Jul 21, 2023
Review
819 days
Class
Class II
Product code
FRN
Recalls
16 (7 Class I)
Predicates
1

Its longest line of descent runs back 4 generations to K940147 (1994). 17 ancestors in all, the oldest K890120 (1989). 1 later device descends from it, 1 citing it directly. 7 of the 19 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2007: 1 other ancestor, 1 with a Class I recall12005: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor2001: 2 other ancestors, 1 with a Class I recall11995: 2 other ancestors, 1 with a Class I recall12000: 1 other ancestor1999: 1 other ancestor1996: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor2025: 1 descendant clearedK940147 · Auto Syringe AS40A Infusion Pump · Baxter Corp. · 1994 · recalledK974332 · Variable Pressure Syringe Pump · Alaris Medical Systems, Inc. · 1998K023264 · Medley Syringe Pump Module System, Model 8110 · Alaris Medical Systems, Inc. · 2002 · Class I recallK133532 · Alaris System with Guardrails Suite MX · Carefusion 303, Inc. · 2014 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 6 other ancestors, 4 with a Class I recall43 generations back: 3 other ancestors4 generations back: 4 other ancestors 1 generation forward: 1 descendant K133532 · Alaris System with Guardrails Suite MX · Carefusion 303, Inc. · 2014 · Class I recallK023264 · Medley Syringe Pump Module System, Model 8110 · Alaris Medical Systems, Inc. · 2002 · Class I recallK974332 · Variable Pressure Syringe Pump · Alaris Medical Systems, Inc. · 1998K940147 · Auto Syringe AS40A Infusion Pump · Baxter Corp. · 1994 · recalled0
  1. Root · 4 generations backK940147 Auto Syringe AS40A Infusion PumpBaxter Corp. · 1994 · Recalled
  2. 3 generations backK974332 Variable Pressure Syringe PumpAlaris Medical Systems, Inc. · 1998
  3. 2 generations backK023264 Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Recalled (Class I)
  4. 1 generation backK133532 Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Recalled (Class I)
  5. This deviceK211218 BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023
  6. Descendant · cites it directly · 0 of its ownK243855 BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc. · 2025

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K940147Auto Syringe AS40A Infusion PumpBaxter Corp.19941
−3K974332Variable Pressure Syringe PumpAlaris Medical Systems, Inc.1998—
−2K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc.2002Class I
−1K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc.2014Class I
0 (this device)K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc.2023—
+1K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199010
199521
199610
199910
200010
200121
200310
200511
200711
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K211218, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133532Alaris System with Guardrails Suite MX
Summary page 7
…1 Predicate Device The predicate device is the Alaris System with Guardrails Suite MX cleared under K133532 (August 21, 2014). Device Description The BD Alaris System with Guardrails Suite MX v12 is a modular infusion and monitoring sys…
Carefusion 303, Inc.2014Class I

Also cleared against K133532: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc.20250

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 11, 2026Class IIZ-1645-2026Under Investigation by firmDue to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units
Nov 6, 2025Class IIZ-1217-2026Under Investigation by firmDue to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices…
Oct 17, 2025Class IZ-0430-2026Device DesignIf infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed…
Jul 8, 2025Class IZ-2179-2025Under Investigation by firmInfusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Feb 18, 2025Class IZ-1304-2025Under Investigation by firmSoftware issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).
Feb 18, 2025Class IZ-1305-2025Under Investigation by firmSoftware issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).
Jan 23, 2025Class IIZ-1231-2025Under Investigation by firmDue to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is…
Aug 15, 2024Class IIZ-3139-2024Process controlDue to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
Aug 15, 2024Class IIZ-3138-2024Process controlDue to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
Aug 15, 2024Class IIZ-3140-2024Process controlDue to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
Aug 15, 2024Class IIZ-3141-2024Process controlDue to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
Aug 15, 2024Class IIZ-3142-2024Process controlDue to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
May 21, 2024Class IIZ-2262-2024Under Investigation by firmDue to a software issue the PC unit may not connect to the server which could impact wireless data transmission to and from the server.
Sep 15, 2023Class IZ-0185-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Sep 15, 2023Class IZ-0184-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Sep 15, 2023Class IZ-0186-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Direct predicate · ◆ on the line of descent in FIG. 128 Class I recalls
  • K072105Alaris System with MMS, Model 8XXXCardinal Health 303,Inc. · 2007 · 2 generations back · ◆ in the 2007 ancestors bar of FIG. 13 Class I recalls
  • K051641Alaris System PC Unit, Model 8001Cardinal Health, Alaris Products · 2005 · 2 generations back · ◆ in the 2005 ancestors bar of FIG. 110 Class I recalls
  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K012383Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem Iii Infusion (Msiii)…Alaris Medical Systems, Inc. · 2001 · 2 generations back · ◆ in the 2001 ancestors bar of FIG. 13 Class I recalls
  • K950419The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · 2 generations back · ◆ in the 1995 ancestors bar of FIG. 14 Class I recalls

What does this lineage say?

  • The line of descent spans 29 years and passes through 3 companies; a new generation was cleared every 8.9 years on median.
  • 11 of the 17 ancestors have a recall of any class; 6 have a Class I recall.
  • Of its descendants, K243855 (Carefusion 303, Inc., 2025) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K133532.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. See Pediatric*, Neonate**, and Adult Patient Population Tables for the module-specific variations. Administered…

Read the full statement (179 words)

The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. See Pediatric*, Neonate**, and Adult Patient Population Tables for the module-specific variations. Administered fluids include pharmaceutical drugs, red blood cells, and other blood components (platelets and fresh frozen plasma) as required for patient therapy. The BD Alaris System is an interoperable system capable of communicating and exchanging data with compatible information technology systems.

The BD Alaris System includes the PC Unit (PCU) and one or more of the following: Pump Module, Syringe Module, End-Tidal CO2 (EtCO2) Module, Auto-ID Module, Patient-Controlled Analgesia (PCA) Module, and associated software applications. The EtCO2 Module is a capnograph that continuously monitors end-tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).

The BD Alaris Pump Module, Syringe Module, and the Alaris PCA Module are indicated for varying patient populations, routes of administration, and infusates.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K211218: BD Alaris System with Guardrails Suite MX v12.1.2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K211218

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026