510(k) Report

510(k) · Cleared

Recalled · Class I · 2021

GE 3.0T Signa Infinity Twinspeed with Excite MR System

K030874510(k) SpecialCleared Mar 28, 20038-day reviewClass II21 CFR 892.1000Recalls 9 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

General Electric Co. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Mar 28, 2003 as substantially equivalent to K022397 (GE Medical Systems, 2002) and 1 other predicate after an 8-day review. Its family tree reaches back 2 generations to 1999. 2 of the 105 devices in its family, counting itself, have a Class I recall. 100 later devices descend from it. It has been recalled 9 times, including a Class I recall.

Decided
Mar 28, 2003
Review
8 days
Class
Class II
Product code
LNH
Recalls
9 (1 Class I)
Predicates
2

Its longest line of descent runs back 2 generations to K990550 (1999). 4 ancestors in all. 100 later devices descend from it, 1 citing it directly. 2 of the 105 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors2004: 1 descendant cleared, 1 with a Class I recall12005: 3 descendants cleared2014: 6 descendants cleared2012: 5 descendants cleared2008: 3 descendants cleared2009: 2 descendants cleared2007: 1 descendant cleared2011: 4 descendants cleared2010: 3 descendants cleared2013: 7 descendants cleared2020: 8 descendants cleared2015: 6 descendants cleared2016: 5 descendants cleared2018: 3 descendants cleared2021: 5 descendants cleared2022: 8 descendants cleared2025: 6 descendants cleared2017: 6 descendants cleared2023: 7 descendants cleared2019: 5 descendants cleared2024: 2 descendants cleared2026: 4 descendants clearedK990550 · 3.0T Signa VH/I Magnetic Resonance System · Ge Medical Systems, Inc. · 1999K022397 · Signa 3.0T Magnetic Resonance System · GE Medical Systems · 2002 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 3 descendants3 generations forward: 8 descendants4 generations forward: 13 descendants5 generations forward: 17 descendants6 generations forward: 11 descendants7 generations forward: 8 descendants8 generations forward: 8 descendants9 generations forward: 7 descendants10 generations forward: 8 descendants11 generations forward: 5 descendants12 generations forward: 2 descendants13 generations forward: 5 descendants14 generations forward: 4 descendants K022397 · Signa 3.0T Magnetic Resonance System · GE Medical Systems · 2002 · recalledK990550 · 3.0T Signa VH/I Magnetic Resonance System · Ge Medical Systems, Inc. · 19990+5+10
  1. Root · 2 generations backK990550 3.0T Signa VH/I Magnetic Resonance SystemGe Medical Systems, Inc. · 1999
  2. 1 generation backK022397 Signa 3.0T Magnetic Resonance SystemGE Medical Systems · 2002 · Recalled
  3. This deviceK030874 GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co. · 2003
  4. Descendant · cites it directly · 99 of its ownK041476 GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · Recalled (Class I)
  5. Descendant · 2 generations on · 90 of its ownK052293 GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005 · Recalled
  6. Descendant · 3 generations on · 70 of its ownK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
  7. Descendant · 3 generations on · 69 of its ownK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
  8. Descendant · 4 generations on · 59 of its ownK091536 Optima MR450WGe Medical Systems, LLC · 2009 · Recalled

Showing 5 of 100 descendants. See all descendants

Descendants cleared per year

2004: 12005: 32007: 12008: 32009: 22010: 32011: 42012: 52013: 72014: 62015: 62016: 52017: 62018: 32019: 52020: 82021: 52022: 82023: 72024: 22025: 62026: 4
2004high 8 (2020) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 100 · See all
  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · cites it directly99descendants
  • K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005 · 2 generations on90descendants
  • K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · 3 generations on70descendants
  • K083147Discovery MR450Ge Medical Systems, LLC · 2008 · 3 generations on69descendants
  • K091536Optima MR450WGe Medical Systems, LLC · 2009 · 4 generations on59descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K9905503.0T Signa VH/I Magnetic Resonance SystemGe Medical Systems, Inc.1999—
−1K022397Signa 3.0T Magnetic Resonance SystemGE Medical Systems20021
0 (this device)K030874GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co.2003—
+1K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC2004Class I
+2K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC200521
+3K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC200811
+3K083147Discovery MR450Ge Medical Systems, LLC200810
+4K091536Optima MR450WGe Medical Systems, LLC20098

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200120

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200411
200530
200710
200830
200920
201030
201140
201250
201370
201460
201560
201650
201760
201830
201950
202080
202150
202280
202370
202420
202560
202640
FIG. 1 — Lineage of K030874, 1999–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022397Signa 3.0T Magnetic Resonance System
Summary page 1
…892.1000, 90-LNH Marketed Device: The [email protected] Infinity Twinspeed with Excite MR System is substantially equivalent to the currently marketed [email protected] MR system (K022397) with the main differences being an addition of eight independent receive channels and an update to the magnet design to increa…
GE Medical Systems20021
*K003575Modification to 3.0T Signa VH/I Magnetic Resonance System
Summary page 2
…st in diagnosis. Comparison with Predicate Device: The [email protected] Infinity Twinspeed with Excite MR System is a modification of the Signa@ 3.OT MR system (K022397) and (K003575) with the main differences being the addition of eight independent receive channels, the addition of the dual mode gradient coil,…
GE Medical Systems2001—

Also cleared against K022397: 1 other device →

What descends from it? (100)

See all 100

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR System Class IGe Medical Systems, LLC200499

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 23, 2021Class IIZ-0613-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Jun 7, 2018Class IIZ-0112-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jun 7, 2018Class IIZ-0110-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jun 7, 2018Class IIZ-0111-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jul 10, 2015Class IIZ-2339-2015Nonconforming Material/ComponentA gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with incorrect orientation…
Mar 9, 2015Class IIZ-1305-2015Software Manufacturing/Software DeploymentGE Healthcare has become aware of a potential safety issue involving MRI systems due to software versions not being maintained properly at some sites.
Jan 7, 2015Class IZ-1035-2015Employee errorAt certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be…
Nov 21, 2011Class IIZ-1943-2013Software designScreen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system…
Nov 9, 2010Not classifiedZ-1431-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the gradient cables are…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 4 years and passes through 3 companies; a new generation was cleared every 3.3 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K041476 (Ge Medical Systems, LLC, 2004) has gone furthest, with 99 descendants of its own.
  • 1 other device also cites its predicate K022397.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The GE Signa@Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa@Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head…

Read the full statement (118 words)

The GE Signa@Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa@Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa@Infinity systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K030874: GE 3.0T Signa Infinity Twinspeed with Excite MR System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K030874

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.