510(k) · Cleared
Recalled · 2009Intera-Sensation Family
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Medical Systems · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Sep 10, 2003 after a 90-day review. The FDA summary names its predicate without a number. 5 of the 68 devices in its family, counting itself, have a Class I recall. 67 later devices descend from it. It has been recalled 3 times.
- Decided
- Sep 10, 2003
- Review
- 90 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 3
- Predicates
- 0
67 later devices descend from it, 2 citing it directly. 5 of the 68 devices in this family, counting this one, have a Class I recall.
- This deviceK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003
- Descendant · cites it directly · 65 of its ownK043147 Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Recalled (Class I)
- Descendant · cites it directly · 58 of its ownK041602 Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Recalled (Class I)
- Descendant · 2 generations on · 57 of its ownK052078 Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Recalled (Class I)
- Descendant · 2 generations on · 53 of its ownK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Recalled (Class I)
- Descendant · 2 generations on · 34 of its ownK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · Recalled
Showing 5 of 67 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 67 · See all- K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · cites it directly65descendants
- K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · cites it directly58descendants
- K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · 2 generations on57descendants
- K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · 2 generations on53descendants
- K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · 2 generations on34descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K031815 | Intera-Sensation Family | Philips Medical Systems | 2003 | — |
| +1 | K043147 | Achieva Family | Philips Medical Systems North America Co. | 2005 | Class I |
| +1 | K041602 | Panorama 1.0T | Philips Medical Systems North America, Inc. | 2004 | Class I |
| +2 | K052078 | Achieva, Intera and Panorama 1.0 Release 2-Series | Philips Medical Systems Nederland B.V. | 2005 | Class I |
| +2 | K063559 | Achieva, Intera and Panorama 1.0T, Release 2.5 | Philips Medical Systems | 2007 | Class I |
| +2 | K110151 | Achieva R4 1.5T and Achieva R4 3.0T | Philips Health Care | 2011 | 20 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 1 | 1 |
| 2005 | 3 | 3 |
| 2007 | 1 | 1 |
| 2008 | 2 | 0 |
| 2010 | 2 | 0 |
| 2011 | 4 | 0 |
| 2012 | 4 | 0 |
| 2013 | 1 | 0 |
| 2014 | 2 | 0 |
| 2015 | 4 | 0 |
| 2016 | 3 | 0 |
| 2017 | 3 | 0 |
| 2018 | 10 | 0 |
| 2019 | 2 | 0 |
| 2020 | 3 | 0 |
| 2021 | 4 | 0 |
| 2022 | 8 | 0 |
| 2023 | 5 | 0 |
| 2025 | 3 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Gyroscan INTERA family release 10-series Named without a submission number |
What descends from it? (67)
See all 672 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (5 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 12, 2009 | Not classified | Z-0794-2011 | Nonconforming Material/Component | The Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system. |
| May 1, 2009 | Not classified | Z-0469-2011 | Other | Combined use of the Synergy Flex-M / Shoulder Coil 1.5T increases the chance of RF interaction and heating up of the coil. This may result in possible burns of the patient. |
| Apr 30, 2007 | Not classified | Z-0983-2007 | Other | Potential for gaseous helium to improperly vent, resulting in rupture of the magnet vessel and venting of helium into the MRI site. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Cites it directly · ◆ in the 2005 descendants bar of FIG. 13 Class I recalls
- K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
- K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · 2 generations on · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
- K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · 2 generations on · ◆ in the 2005 descendants bar of FIG. 11 Class I recall
- K052013Achieva 1.5T & Inera 1.5T FamilyPhilips Medical Systems Nederland B.V. · 2005 · 2 generations on · ◆ in the 2005 descendants bar of FIG. 13 Class I recalls
What does this lineage say?
- Of its descendants, K043147 (Philips Medical Systems North America Co., 2005) has gone furthest, with 65 descendants of its own.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The indication for use for the INTERA SENSATION Release 1-series remains the same as the previous released predicate device(s), i.e. the capability as a diagnostic device that produces transverse, sagittal, coronal and oblique cross-sectional images based upon 'H metabolites, and that displays the internal structure of the whole body. These images when interpreted by a trained physician, yield information that may assist in diagnosis
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K031815 · printed Nov 3, 2026
Cite this page
510(k) Report. “K031815: Intera-Sensation Family.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K031815
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026