Recalled · Class I · 2026Achieva, Intera and Panorama 1.0 Release 2-Series
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Medical Systems Nederland B.V. · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Sep 7, 2005 as substantially equivalent to K043147 (Philips Medical Systems North America Co., 2005) and 1 other predicate after a 37-day review. Its family tree reaches back 2 generations to 2003. 4 of the 61 devices in its family, counting itself, have a Class I recall. 57 later devices descend from it. It has been recalled 12 times, including a Class I recall.
- Decided
- Sep 7, 2005
- Review
- 37 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 12 (1 Class I)
- Predicates
- 2
Its longest line of descent runs back 2 generations to K031815 (2003). 3 ancestors in all. 57 later devices descend from it, 2 citing it directly. 4 of the 61 devices in this family, counting this one, have a Class I recall.
- Root · 2 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
- 1 generation backK041602 Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Recalled (Class I)
- This deviceK052078 Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005
- Descendant · cites it directly · 53 of its ownK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Recalled (Class I)
- Descendant · 2 generations on · 34 of its ownK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · Recalled
- Descendant · 3 generations on · 19 of its ownK163116 Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · Recalled
- Descendant · 4 generations on · 15 of its ownK173079 Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · Recalled
- Descendant · 5 generations on · 13 of its ownK183063 Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · Recalled
Showing 5 of 57 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 57 · See all- K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · cites it directly53descendants
- K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · 2 generations on34descendants
- K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · 3 generations on19descendants
- K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · 4 generations on15descendants
- K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · 5 generations on13descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K031815 | Intera-Sensation Family | Philips Medical Systems | 2003 | 3 |
| −1 | K041602 | Panorama 1.0T | Philips Medical Systems North America, Inc. | 2004 | Class I |
| 0 (this device) | K052078 | Achieva, Intera and Panorama 1.0 Release 2-Series | Philips Medical Systems Nederland B.V. | 2005 | — |
| +1 | K063559 | Achieva, Intera and Panorama 1.0T, Release 2.5 | Philips Medical Systems | 2007 | Class I |
| +2 | K110151 | Achieva R4 1.5T and Achieva R4 3.0T | Philips Health Care | 2011 | 20 |
| +3 | K163116 | Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 | Philips Medical Systems Nederland B.V. | 2017 | 7 |
| +4 | K173079 | Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4 | Philips Medical Systems Nederland B.V. | 2018 | 11 |
| +5 | K183063 | Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems | Philips Medical Systems Nederland B.V. | 2019 | 25 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2007 | 1 | 1 |
| 2008 | 1 | 0 |
| 2010 | 2 | 0 |
| 2011 | 3 | 0 |
| 2012 | 2 | 0 |
| 2013 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 4 | 0 |
| 2016 | 3 | 0 |
| 2017 | 3 | 0 |
| 2018 | 9 | 0 |
| 2019 | 2 | 0 |
| 2020 | 3 | 0 |
| 2021 | 4 | 0 |
| 2022 | 8 | 0 |
| 2023 | 5 | 0 |
| 2025 | 3 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K043147 | Achieva FamilySummary page 1…ACHIEVA, INTERA and PANORAMA 1.OT Release 2-series are the successor of the predicate devices ACHIEVA, INTERA and PANORAMA L.OT release 1-series. (FDA references K043147, K043147 & K041602). Indications for use: The ACHIEVA, INTERA and PANORAMA 1.OT Release 2-series are magnetic resonance… | 2005 |
| * | K041602 | Panorama 1.0TSummary page 1…PANORAMA 1.OT Release 2-series are the successor of the predicate devices ACHIEVA, INTERA and PANORAMA L.OT release 1-series. (FDA references K043147, K043147 & K041602). Indications for use: The ACHIEVA, INTERA and PANORAMA 1.OT Release 2-series are magnetic resonance diagnostic devices… | 2004 |
What descends from it? (57)
See all 572 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 21, 2026 | Class II | Z-3234-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3240-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3233-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3232-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Nov 17, 2023 | Class I | Z-0505-2024 | Component design/selection | Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium… |
| Oct 23, 2023 | Class II | Z-0393-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0389-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0390-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0392-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Dec 21, 2018 | Class II | Z-0876-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0875-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Nov 20, 2014 | Class II | Z-1307-2015 | Component design/selection | VCG battery ignited in a VCG unit when recharging. |
The 5 latest of 12 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Direct predicate · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
- K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
- K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 2 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
- 3 of the 3 ancestors have a recall of any class; 2 have a Class I recall.
- Of its descendants, K063559 (Philips Medical Systems, 2007) has gone furthest, with 53 descendants of its own.
- 5 other devices also cite its predicate K043147.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The ACHIEVA, INTERA and PANORAMA 1.OT Release 2-series are magnetic resonance diagnostic devices that produce cross-sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body. These images when interpreted by a trained physician, yield information that may assist in diagnosis.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K052078 · printed Nov 6, 2026
Cite this page
510(k) Report. “K052078: Achieva, Intera and Panorama 1.0 Release 2-Series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052078
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026