Panorama 1.0T

K041602510(k) TraditionalCleared Jul 26, 200442-day reviewClass II21 CFR 892.1000Recalls 7 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America, Inc. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Jul 26, 2004 as substantially equivalent to K031815 (Philips Medical Systems, 2003) after a 42-day review. Its family tree reaches back 1 generation to 2003. 3 of the 60 devices in its family, counting itself, have a Class I recall. 58 later devices descend from it. It has been recalled 7 times, including a Class I recall.

Decided
Jul 26, 2004
Review
42 days
Class
Class II
Product code
LNH
Recalls
7 (1 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K031815 (2003). 58 later devices descend from it, 2 citing it directly. 3 of the 60 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2007: 1 descendant cleared, 1 with a Class I recall12005: 1 descendant cleared, 1 with a Class I recall12008: 1 descendant cleared2019: 2 descendants cleared2017: 3 descendants cleared2010: 2 descendants cleared2011: 3 descendants cleared2018: 9 descendants cleared2012: 2 descendants cleared2015: 4 descendants cleared2016: 3 descendants cleared2013: 1 descendant cleared2020: 3 descendants cleared2014: 1 descendant cleared2026: 2 descendants cleared2021: 4 descendants cleared2022: 8 descendants cleared2025: 3 descendants cleared2023: 5 descendants clearedK031815 · Intera-Sensation Family · Philips Medical Systems · 2003 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants, 2 with a Class I recall22 generations forward: 8 descendants3 generations forward: 12 descendants4 generations forward: 10 descendants5 generations forward: 11 descendants6 generations forward: 5 descendants7 generations forward: 2 descendants8 generations forward: 3 descendants9 generations forward: 3 descendants10 generations forward: 2 descendants K031815 · Intera-Sensation Family · Philips Medical Systems · 2003 · recalled0+5+10
  1. Root · 1 generation backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
  2. This deviceK041602 Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004
  3. Descendant · cites it directly · 57 of its ownK052078 Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Recalled (Class I)
  4. Descendant · cites it directly · 53 of its ownK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Recalled (Class I)
  5. Descendant · 2 generations on · 34 of its ownK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · Recalled
  6. Descendant · 3 generations on · 19 of its ownK163116 Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · Recalled
  7. Descendant · 4 generations on · 15 of its ownK173079 Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · Recalled

Showing 5 of 58 descendants. See all descendants

Descendants cleared per year

2005: 12007: 12008: 12010: 22011: 32012: 22013: 12014: 12015: 42016: 32017: 32018: 92019: 22020: 32021: 42022: 82023: 52025: 32026: 2
2005high 9 (2018) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 58 · See all
  • K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · cites it directly57descendants
  • K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · cites it directly53descendants
  • K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · 2 generations on34descendants
  • K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · 3 generations on19descendants
  • K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · 4 generations on15descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K031815Intera-Sensation FamilyPhilips Medical Systems20033
0 (this device)K041602Panorama 1.0TPhilips Medical Systems North America, Inc.2004—
+1K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V.2005Class I
+1K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems2007Class I
+2K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care201120
+3K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V.20177
+4K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V.201811

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200511
200711
200810
201020
201130
201220
201310
201410
201540
201630
201730
201890
201920
202030
202140
202280
202350
202530
202620
FIG. 1 — Lineage of K041602, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031815Intera-Sensation Family
Summary page 1
…PANORAMA 1LOT is a magnetic resonance diagnostic device with a vertical field based on the same platform as its predicate device cylindrical INTERA ACHIEVA (ref. K0318 15). Indications for use: PANORAMA 1.OT is a Magnetic Resonance Diagnostic Device intended for general diagnostic use that…
Philips Medical Systems20033

Also cleared against K031815: 1 other device →

What descends from it? (58)

See all 58

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K052078Achieva, Intera and Panorama 1.0 Release 2-Series Class IPhilips Medical Systems Nederland B.V.200557
K063559Achieva, Intera and Panorama 1.0T, Release 2.5 Class IPhilips Medical Systems200753

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 17, 2023Class IZ-0505-2024Component design/selectionFirm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium…
Dec 21, 2018Class IIZ-0898-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0917-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Mar 16, 2018Class IIZ-1434-2018Nonconforming Material/ComponentPotential risk for helium gas inside the MR examination room during a magnet quench
Feb 5, 2010Not classifiedZ-1458-2011OtherIn the ExamCard MobiFlex the Patient Position is "Prone" which must be "Supine". The result is mislabeling of left and right side of the patient. If this is not detected by the operator there is a risk of wrong treatment or the need of…
May 12, 2009Not classifiedZ-0794-2011Nonconforming Material/ComponentThe Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system.
Apr 30, 2007Not classifiedZ-0984-2007OtherPotential for gaseous helium to improperly vent, resulting in rupture of the magnet vessel and venting of helium into the MRI site.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Cites it directly · ◆ in the 2005 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K052078 (Philips Medical Systems Nederland B.V., 2005) has gone furthest, with 57 descendants of its own.
  • 1 other device also cites its predicate K031815.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

PANORAMA 1.OT is a Magnetic Resonance Diagnostic Device intended for general diagnostic use that produces transverse, sagittal, coronal and oblique cross-sectional images based upon 'H metabolites, and that displays the internal structure of the whole body. These images when interpreted by a trained physician, yield information that may assist in diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041602: Panorama 1.0T.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041602

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026