510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

ABL800 Flex with Flexq Module, Model ABL8XX

K043218510(k) TraditionalCleared May 10, 2005169-day reviewClass II21 CFR 862.1120Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer Medical Aps · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on May 10, 2005 as substantially equivalent to K041874 (Radiometer Medical Aps, 2004) and 1 other predicate after a 169-day review. Its family tree reaches back 5 generations to 1986. 4 later devices descend from it. It has been recalled 3 times.

Decided
May 10, 2005
Review
169 days
Class
Class II
Product code
CHL
Recalls
3
Predicates
2

Its longest line of descent runs back 5 generations to K861087 (1986). 16 ancestors in all. 4 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1998: 2 other ancestors1987: 1 other ancestor1986: 1 other ancestor1997: 1 other ancestor1996: 3 other ancestors1995: 1 other ancestor1992: 1 other ancestor2015: 1 descendant cleared2008: 1 descendant cleared2006: 1 descendant cleared2011: 1 descendant clearedK861087 · Avl Model 983 Electrolyte Analyzer · Avl Scientific Corp. · 1986K961458 · Avl 9180 Electrolyte Analyzer · Avl Scientific Corp. · 1996 · recalledK972673 · Avl 9181 Electrolyte Analyzer · Avl Scientific Corp. · 1997 · recalledK992859 · ABL700 Series Analyzer with Autocheck Module · Radiometer America, Inc. · 2000 · recalledK041874 · ABL800 Flex · Radiometer Medical Aps · 2004 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 4 other ancestors4 generations back: 3 other ancestors5 generations back: 3 other ancestors 1 generation forward: 3 descendants2 generations forward: 1 descendant K041874 · ABL800 Flex · Radiometer Medical Aps · 2004 · recalledK992859 · ABL700 Series Analyzer with Autocheck Module · Radiometer America, Inc. · 2000 · recalledK972673 · Avl 9181 Electrolyte Analyzer · Avl Scientific Corp. · 1997 · recalledK961458 · Avl 9180 Electrolyte Analyzer · Avl Scientific Corp. · 1996 · recalledK861087 · Avl Model 983 Electrolyte Analyzer · Avl Scientific Corp. · 1986−50
  1. Root · 5 generations backK861087 Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp. · 1986
  2. 4 generations backK961458 Avl 9180 Electrolyte AnalyzerAvl Scientific Corp. · 1996 · Recalled
  3. 3 generations backK972673 Avl 9181 Electrolyte AnalyzerAvl Scientific Corp. · 1997 · Recalled
  4. 2 generations backK992859 ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc. · 2000 · Recalled
  5. 1 generation backK041874 ABL800 FlexRadiometer Medical Aps · 2004 · Recalled
  6. This deviceK043218 ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps · 2005
  7. Descendant · cites it directly · 1 of its ownK073537 Neonatal Bilirubin on Rapidlab Models 1245 and 1265Siemens Medical Solutions Diagnostics · 2008 · Recalled
  8. Descendant · cites it directly · 0 of its ownK151639 ABL80 FLEX and ABL80 FLEX CO-OXSendx Medical, Inc. · 2015
  9. Descendant · 2 generations on · 0 of its ownK110277 Rapidpoint 405 Neonatal BilirubinSiemens Healthcare Diagnostics · 2011 · Recalled
  10. Descendant · cites it directly · 0 of its ownK051968 ABL8X7 FlexRadiometer Medical Aps · 2006 · Recalled

Most influential descendants

Showing 4 of 4 · See all
  • K073537Neonatal Bilirubin on Rapidlab Models 1245 and 1265Siemens Medical Solutions Diagnostics · 2008 · cites it directly1descendant
  • K151639ABL80 FLEX and ABL80 FLEX CO-OXSendx Medical, Inc. · 2015 · cites it directly0descendants
  • K110277Rapidpoint 405 Neonatal BilirubinSiemens Healthcare Diagnostics · 2011 · 2 generations on0descendants
  • K051968ABL8X7 FlexRadiometer Medical Aps · 2006 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K861087Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp.1986—
−4K961458Avl 9180 Electrolyte AnalyzerAvl Scientific Corp.19961
−3K972673Avl 9181 Electrolyte AnalyzerAvl Scientific Corp.19977
−2K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc.20004
−1K041874ABL800 FlexRadiometer Medical Aps200422
0 (this device)K043218ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps2005—
+1K073537Neonatal Bilirubin on Rapidlab Models 1245 and 1265Siemens Medical Solutions Diagnostics20087
+1K151639ABL80 FLEX and ABL80 FLEX CO-OXSendx Medical, Inc.2015—
+2K110277Rapidpoint 405 Neonatal BilirubinSiemens Healthcare Diagnostics20115
+1K051968ABL8X7 FlexRadiometer Medical Aps20064

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198710
199210
199510
199630
199710
199820
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200810
201110
201510
FIG. 1 — Lineage of K043218, 1986–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.2 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041874ABL800 Flex
Summary page 1
…Classification Name: Blood gases and blood pH test system (21 CFR Section 862.1120) Predicate Devices RADIOMETER ABL800 FLEX (K041874) and AVL 9181 Electrolyte Analyzer (K972673). Device Description The ABL800 FLEX with FLEXO Module is an ABL800 FLEX Analyzer w…
Radiometer Medical Aps200422
*K972673Avl 9181 Electrolyte Analyzer
Summary page 1
…and blood pH test system (21 CFR Section 862.1120) Predicate Devices RADIOMETER ABL800 FLEX (K041874) and AVL 9181 Electrolyte Analyzer (K972673). Device Description The ABL800 FLEX with FLEXO Module is an ABL800 FLEX Analyzer with the added optional capability of automat…
Avl Scientific Corp.19977

Also cleared against K041874: 5 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K073537Neonatal Bilirubin on Rapidlab Models 1245 and 1265Siemens Medical Solutions Diagnostics20081
K151639ABL80 FLEX and ABL80 FLEX CO-OXSendx Medical, Inc.20150
K051968ABL8X7 FlexRadiometer Medical Aps20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 8, 2012Class IIZ-2303-2013Process controlRADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
Apr 7, 2011Not classifiedZ-2571-2011PendingBased on internal testing and user/customer complaint reports received, the firm determined that wrinkles may arise on the outer membrane of the calcium membrane units upon installation. Retention of fluids between the wrinkles can cause…
Jan 16, 2007Not classifiedZ-0705-2007OtherIn some situations pre-registered patient data is not reset when leaving the pre-registration mode in the software. This occurs when a Safe Pico blood sampler associated with a specific patient is pre-registered in the analyzer by…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 3 companies; a new generation was cleared every 2.4 years on median.
  • 9 of the 16 ancestors have a recall of any class.
  • Of its descendants, K073537 (Siemens Medical Solutions Diagnostics, 2008) has gone furthest, with 1 descendant of its own.
  • 5 other devices also cite its predicate K041874.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

The ABL800 FLEX with FLEXQ Module is intended for in vitro testing of samples of whole blood for the parameters pH, P0 2, pCO 2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) as well as for in vitro testing of samples of expired air for the parameters PO2 and pCO 2. Indication for use information…

Read the full statement (463 words)

The ABL800 FLEX with FLEXQ Module is intended for in vitro testing of samples of whole blood for the parameters pH, P0 2, pCO 2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) as well as for in vitro testing of samples of expired air for the parameters PO2 and pCO 2.

Indication for use information for the analytes measured by the ABL800 FLEX:

pH: pH is the indispensable measure of acidemia or alkalemia and is therefore an essential part of the pH/blood gas measurement. The normal function of many metabolic processes requires a pH to be within a relatively narrow range.

PO2: The arterial oxygen tension is an indicator of the oxygen uptake in the lungs.

PC0 2: pCO 2 is a direct reflection of the adequacy of alveolar ventilation in relation to the metabolic rate.

Potassium (cK4): The measurements of the concentration of potassium ions in plasma are used to monitor the electrolyte balance.

Sodium (cNa'): The measurements of the concentration of sodium ions in plasma are used to monitor the electrolyte balance.

Calcium (CCa 2 l): The measurements of the concentration of calcium ions in plasma are used to monitor the electrolyte balance.

Chloride (cClj: The measurements of the concentration of chloride ions in plasma are used to monitor the electrolyte balance.

Glucose (cGlu): The glucose measurements measure the concentration of glucose in plasma. The glucose measurements are used to screen for, diagnose and monitor diabetes, pre-diabetes, and hyper- and hypoglycemia.

Lactate (cLac): The lactate measurements measure the concentration of lactate in plasma. Lactate measurements serve as a marker of critical imbalance between tissue oxygen demand and oxygen supply.

Blifrubin (ctBiI): The bilirubin measurements measure the total concentration of bilirubin in plasma. ctBil is used to assess the risk of hyperbilirubinemia.

Total Hemoglobin (ctHb): ctHb is a measure of the potential oxygen-carrying capacity of the blood.

Oxygen Saturation (sO 2): sO2 is the percentage of oxygenated hemoglobin in relation to the amount of hemoglobin capable of carrying oxygen. sO2 allows evaluation of oxygenation.

Fraction of Oxyhemoglobin (FO2Hb): FO2Hb is a measure of the utilization of the potential oxygen transport capacity; that is the fraction of oxyhemoglobin in relation to all hemoglobins present (tHb) including dyshemoglobins.

Fraction of Carboxyhemoglobin (FCOHb): FCOI-b is the fraction of carboxyhemoglobin. It is incapable of transporting oxygen.

Fraction of Methemoglobin (FMetHb): FMetHb is the fraction of methemoglobin. It is incapable of transporting oxygen.

Fraction of Deoxyhemoglobin in Total Hemoglobin (FHHb): FHHb is the fraction of deoxyhemoglobin in total hemoglobin. It can bind oxygen then forming oxyhemoglobin.

Fraction of Fetal Hemoglobin (FHbF): Fetal hemoglobin consists of two a-chains and two P-chains, and has a higher oxygen affinity than adult Hb.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K043218: ABL800 Flex with Flexq Module, Model ABL8XX.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K043218

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026