510(k) Report

510(k) · Cleared

Recalled · 2006

SE Infusion System with MMS

K043590510(k) TraditionalCleared Mar 25, 200587-day reviewClass II21 CFR 880.5725Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cardinal Health, Alaris Products · FRN Pump, Infusion

Cleared on Mar 25, 2005 as substantially equivalent to K030459 (Alaris Medical Systems, Inc., 2003) after an 87-day review. Its family tree reaches back 1 generation to 2003. It has been recalled once.

Decided
Mar 25, 2005
Review
87 days
Class
Class II
Product code
FRN
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K030459 (2003). No later device cites it as a predicate yet.

K030459 · Medley System with Medication Management System (MMS) · Alaris Medical Systems, Inc. · 20032000200220042006
  1. Root · 1 generation backK030459 Medley System with Medication Management System (MMS)Alaris Medical Systems, Inc. · 2003
  2. This deviceK043590 SE Infusion System with MMSCardinal Health, Alaris Products · 2005
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K030459Medley System with Medication Management System (MMS)Alaris Medical Systems, Inc.2003—
0 (this device)K043590SE Infusion System with MMSCardinal Health, Alaris Products2005—
FIG. 1 — Lineage of K043590, 2003–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030459Medley System with Medication Management System (MMS)
Summary page 2
…ce: Predicate Device Manufacturer 510(k) No. Date Cleared Medley System with Cardinal Health, Alaris K030459 April 4, 2003 MMS Products (previously known as A…
Alaris Medical Systems, Inc.2003—

Also cleared against K030459: 8 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 10, 2006Not classifiedZ-1484-06Software designThis recall was initiated because of a potential for over infusion with all models of the Alaris SE Pumps (formerly the Signature Edition Infusion Pumps) caused by key bounce.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 8 other devices also cite its predicate K030459.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The SE Infusion System with MMS is intended for use in today's

growing professional healthcare environment for facilities that utilize infusion devices for

the delivery of fluids, medications, blood and blood products.

The SE Infusion System with MMS is intended to provide trained healthcare caregivers a

way to automate the programming of infusion parameters, thereby decreasing the amount

of manual steps necessary to enter infusion data. All data entry and validation of infusion

parameters is performed by the trained healthcare professional according to a physician's

order.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K043590: SE Infusion System with MMS.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K043590

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026