510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Alaris System with MMS, Model 8XXX

K072105510(k) SpecialCleared Dec 28, 2007150-day reviewClass II21 CFR 880.5725Recalls 5 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cardinal Health 303,Inc. · FRN Pump, Infusion

Cleared on Dec 28, 2007 as substantially equivalent to K030459 (Alaris Medical Systems, Inc., 2003) after a 150-day review. Its family tree reaches back 1 generation to 2003. 3 of the 6 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 5 times, including 3 Class I recalls.

Decided
Dec 28, 2007
Review
150 days
Class
Class II
Product code
FRN
Recalls
5 (3 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K030459 (2003). 4 later devices descend from it, 1 citing it directly. 3 of the 6 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2014: 1 descendant cleared, 1 with a Class I recall12022: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant clearedK030459 · Medley System with Medication Management System (MMS) · Alaris Medical Systems, Inc. · 2003200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 1 descendant K030459 · Medley System with Medication Management System (MMS) · Alaris Medical Systems, Inc. · 20030
  1. Root · 1 generation backK030459 Medley System with Medication Management System (MMS)Alaris Medical Systems, Inc. · 2003
  2. This deviceK072105 Alaris System with MMS, Model 8XXXCardinal Health 303,Inc. · 2007
  3. Descendant · cites it directly · 3 of its ownK133532 Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Recalled (Class I)
  4. Descendant · 2 generations on · 1 of its ownK211218 BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · Recalled (Class I)
  5. Descendant · 3 generations on · 0 of its ownK243855 BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc. · 2025
  6. Descendant · 2 generations on · 0 of its ownK210074 Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia…Fresenius Kabi AG · 2022

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · cites it directly3descendants
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 2 generations on1descendant
  • K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc. · 2025 · 3 generations on0descendants
  • K210074Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link, Agilia Duo…Fresenius Kabi AG · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K030459Medley System with Medication Management System (MMS)Alaris Medical Systems, Inc.2003—
0 (this device)K072105Alaris System with MMS, Model 8XXXCardinal Health 303,Inc.2007—
+1K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc.2014Class I
+2K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc.2023Class I
+3K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc.2025—
+2K210074Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link…Fresenius Kabi AG2022—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201411
202210
202311
202510
FIG. 1 — Lineage of K072105, 2003–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030459Medley System with Medication Management System (MMS)
Summary page 1
…Pump, Infusion and accessories, FRN (880.5725) PREDICATE DEVICE: Alaris® System with MMS (a.k.a. MedleyTM System with MMS), K030459 DEVICE DESCRIPTION As with the predicate device, the Alaris® System with MMS assists our customers in reducing the number of man…
Alaris Medical Systems, Inc.2003—

Also cleared against K030459: 8 other devices →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133532Alaris System with Guardrails Suite MX Class ICarefusion 303, Inc.20143

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 11, 2026Class IIZ-1645-2026Under Investigation by firmDue to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units
Sep 15, 2023Class IZ-0185-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Sep 15, 2023Class IZ-0184-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Sep 15, 2023Class IZ-0186-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Jan 24, 2023Class IIZ-1192-2023Software Design ChangeInfusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems.…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 128 Class I recalls
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 17 Class I recalls

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K133532 (Carefusion 303, Inc., 2014) has gone furthest, with 3 descendants of its own.
  • 8 other devices also cite its predicate K030459.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Alaris® System with Medication Management System (also known as the Medley™ System with MMS and the Alaris® System with MMS) is intended for use in today's growing professional healthcare environment for facilities that utilize infusion devices for the delivery of fluids, medications, blood and blood products.

The Alaris® System with MMS is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data.' All data entry and validation of infusion parameters is performed by the trained healthcare professional according to a physician's order.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072105: Alaris System with MMS, Model 8XXX.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072105

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026