510(k) Report

510(k) · Cleared

Cleared

Renaissance System

K110911510(k) SpecialCleared Apr 28, 201128-day reviewClass II21 CFR 882.4560

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Mazor Robotics, Ltd. · HAW Neurological Stereotaxic Instrument

Cleared on Apr 28, 2011 as substantially equivalent to K102130 (Mazor Surgical Technologies, Ltd., 2010) after a 28-day review. Its family tree reaches back 9 generations to 1994. 1 of the 57 devices in its family, counting itself, has a Class I recall. 24 later devices descend from it.

Decided
Apr 28, 2011
Review
28 days
Class
Class II
Product code
HAW
Recalls
None
Predicates
1

Its longest line of descent runs back 9 generations to K911783 (1994). 32 ancestors in all, the oldest K811452 (1981). 24 later devices descend from it, 2 citing it directly. 1 of the 57 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2006: 1 other ancestor2005: 2 other ancestors2004: 1 other ancestor2003: 2 other ancestors2000: 2 other ancestors1996: 1 other ancestor2002: 1 other ancestor2001: 4 other ancestors, 1 with a Class I recall11999: 1 other ancestor1997: 2 other ancestors1987: 1 other ancestor1981: 1 other ancestor1995: 3 other ancestors2011: 1 descendant cleared2012: 1 descendant cleared2014: 1 descendant cleared2015: 1 descendant cleared2017: 2 descendants cleared2018: 2 descendants cleared2020: 2 descendants cleared2023: 3 descendants cleared2021: 1 descendant cleared2024: 1 descendant cleared2026: 6 descendants cleared2025: 3 descendants clearedReference device: K033413 · Spineassist Device · Mazor Surgical Technologies, Ltd. · 2004Reference device: K051676 · Spineassist Device with Additional Accessories · Mazor Surgical Technologies, Ltd. · 2005Reference device: K063607 · Spineassist System · Mazor Surgical Technologies, Ltd. · 2007Reference device: K073467 · Spineassist System · Mazor Surgical Technologies, Ltd. · 2008Reference device: K081672 · C-Insight · Mazor Surgical Technologies, Ltd. · 2008K911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994K960714 · Isg Viewing Wand · I.S.G. Technologies, Inc. · 1996K964229 · Regulus Navigator · Compass Intl., Inc. · 1997K981315 · Navitrack · Orthosoft, Inc. · 1998K033413 · Spineassist Device · Mazor Surgical Technologies, Ltd. · 2004K051676 · Spineassist Device with Additional Accessories · Mazor Surgical Technologies, Ltd. · 2005K063607 · Spineassist System · Mazor Surgical Technologies, Ltd. · 2007K073467 · Spineassist System · Mazor Surgical Technologies, Ltd. · 2008K102130 · Tenzing System · Mazor Surgical Technologies, Ltd. · 201019801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 7 other ancestors4 generations back: 7 other ancestors, 1 with a Class I recall15 generations back: 5 other ancestors6 generations back: 1 other ancestor7 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 2 descendants5 generations forward: 1 descendant6 generations forward: 1 descendant7 generations forward: 1 descendant8 generations forward: 2 descendants9 generations forward: 5 descendants10 generations forward: 5 descendants11 generations forward: 2 descendants K102130 · Tenzing System · Mazor Surgical Technologies, Ltd. · 2010K073467 · Spineassist System · Mazor Surgical Technologies, Ltd. · 2008K063607 · Spineassist System · Mazor Surgical Technologies, Ltd. · 2007K051676 · Spineassist Device with Additional Accessories · Mazor Surgical Technologies, Ltd. · 2005K033413 · Spineassist Device · Mazor Surgical Technologies, Ltd. · 2004K981315 · Navitrack · Orthosoft, Inc. · 1998K964229 · Regulus Navigator · Compass Intl., Inc. · 1997K960714 · Isg Viewing Wand · I.S.G. Technologies, Inc. · 1996K911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994−50+5+10
  1. Root · 9 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 5 generations between
  3. 3 generations backK063607 Spineassist SystemMazor Surgical Technologies, Ltd. · 2007
  4. 2 generations backK073467 Spineassist SystemMazor Surgical Technologies, Ltd. · 2008
  5. 1 generation backK102130 Tenzing SystemMazor Surgical Technologies, Ltd. · 2010
  6. This deviceK110911 Renaissance SystemMazor Robotics, Ltd. · 2011
  7. Descendant · cites it directly · 23 of its ownK113228 Renaissance SystemMazor Robotics, Ltd. · 2011
  8. Descendant · 2 generations on · 21 of its ownK140167 Renaissance X SystemMazor Robotics, Ltd. · 2014
  9. Descendant · cites it directly · 21 of its ownK120812 RenaissanceMazor Robotics, Ltd. · 2012 · Recalled
  10. Descendant · 2 generations on · 20 of its ownK152041 Renaissance X SystemMazor Robotics, Ltd. · 2015
  11. Descendant · 3 generations on · 19 of its ownK163221 Mazor XMazor Robotics, Ltd. · 2017
Show all 9 generations
  1. Root · 9 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 8 generations backK960714 Isg Viewing WandI.S.G. Technologies, Inc. · 1996
  3. 7 generations backK964229 Regulus NavigatorCompass Intl., Inc. · 1997
  4. 6 generations backK981315 NavitrackOrthosoft, Inc. · 1998
  5. 5 generations backK033413 Spineassist DeviceMazor Surgical Technologies, Ltd. · 2004
  6. 4 generations backK051676 Spineassist Device with Additional AccessoriesMazor Surgical Technologies, Ltd. · 2005
  7. 3 generations backK063607 Spineassist SystemMazor Surgical Technologies, Ltd. · 2007
  8. 2 generations backK073467 Spineassist SystemMazor Surgical Technologies, Ltd. · 2008
  9. 1 generation backK102130 Tenzing SystemMazor Surgical Technologies, Ltd. · 2010
  10. This deviceK110911 Renaissance SystemMazor Robotics, Ltd. · 2011

Showing 5 of 24 descendants. See all descendants

Descendants cleared per year

2011: 12012: 12014: 12015: 12017: 22018: 22020: 22021: 12023: 32024: 12025: 32026: 6
2011high 6 (2026) · low 0 (2013)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 24 · See all
  • K113228Renaissance SystemMazor Robotics, Ltd. · 2011 · cites it directly23descendants
  • K140167Renaissance X SystemMazor Robotics, Ltd. · 2014 · 2 generations on21descendants
  • K120812RenaissanceMazor Robotics, Ltd. · 2012 · cites it directly21descendants
  • K152041Renaissance X SystemMazor Robotics, Ltd. · 2015 · 2 generations on20descendants
  • K163221Mazor XMazor Robotics, Ltd. · 2017 · 3 generations on19descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K911783Allegro Viewing WandI.S.G. Technologies, Inc.1994—
−8K960714Isg Viewing WandI.S.G. Technologies, Inc.1996—
−7K964229Regulus NavigatorCompass Intl., Inc.1997—
−6K981315NavitrackOrthosoft, Inc.1998—
−5K033413Spineassist DeviceMazor Surgical Technologies, Ltd.2004—
−4K051676Spineassist Device with Additional AccessoriesMazor Surgical Technologies, Ltd.2005—
−3K063607Spineassist SystemMazor Surgical Technologies, Ltd.2007—
−2K073467Spineassist SystemMazor Surgical Technologies, Ltd.2008—
−1K102130Tenzing SystemMazor Surgical Technologies, Ltd.2010—
0 (this device)K110911Renaissance SystemMazor Robotics, Ltd.2011—
+1K113228Renaissance SystemMazor Robotics, Ltd.2011—
+2K140167Renaissance X SystemMazor Robotics, Ltd.2014—
+1K120812RenaissanceMazor Robotics, Ltd.20121
+2K152041Renaissance X SystemMazor Robotics, Ltd.2015—
+3K163221Mazor XMazor Robotics, Ltd.2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198710
199530
199610
199720
199910
200020
200141
200210
200320
200410
200520
200610
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201210
201410
201510
201720
201820
202020
202110
202330
202410
202530
202660
FIG. 1 — Lineage of K110911, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K102130Tenzing System
Summary page 1
…Predicate Devices The Renaissance System is substantially equivalent to the following device: i\'lanfifacfufer o | Deviee . 510(k)° | DateCleared Mazor Robotics TenZing K102130 08/26/2010 613 3, Device Description The Renaissance system is a device modification of the TenZing system (K102…
Mazor Surgical Technologies, Ltd.2010—
†K033413Spineassist DeviceMazor Surgical Technologies, Ltd.2004—
†K051676Spineassist Device with Additional AccessoriesMazor Surgical Technologies, Ltd.2005—
†K063607Spineassist SystemMazor Surgical Technologies, Ltd.2007—
†K073467Spineassist SystemMazor Surgical Technologies, Ltd.2008—
†K081672C-InsightMazor Surgical Technologies, Ltd.2008—

Also cleared against K102130: 1 other device →

What descends from it? (24)

See all 24

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K113228Renaissance SystemMazor Robotics, Ltd.201123
K120812RenaissanceMazor Robotics, Ltd.201221

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 4 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 1.7 years on median.
  • 7 of the 32 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K113228 (Mazor Robotics, Ltd., 2011) has gone furthest, with 23 descendants of its own.
  • 1 other device also cites its predicate K102130.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

The Renaissance System is indicated for precise positioning of 'surgical instruments or implants during general spinal surgery. It may be used in either open or percutaneous procedures

Renalssance 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K110911: Renaissance System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K110911

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026