510(k) · Cleared
Recalled · Class II · 2013Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table Option
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ge Medical Systems, Inc. · OWB Interventional Fluoroscopic X-Ray System
Cleared on Oct 16, 2003 as substantially equivalent to K021021 (Siemens Medical Solutions USA, Inc., 2002) after a 9-day review. Its family tree reaches back 3 generations to 1989. 75 later devices descend from it. It has been recalled 8 times.
- Decided
- Oct 16, 2003
- Review
- 9 days
- Class
- Class II
- Product code
- OWB
- Recalls
- 8
- Predicates
- 1
Its longest line of descent runs back 3 generations to K890348 (1989). 6 ancestors in all. 75 later devices descend from it, 3 citing it directly.
- Root · 3 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
- 2 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
- 1 generation backK021021 Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc. · 2002 · Recalled
- This deviceK033244 Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt…Ge Medical Systems, Inc. · 2003
- Descendant · cites it directly · 74 of its ownK052412 Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova…GE Healthcare · 2005 · Recalled
- Descendant · cites it directly · 47 of its ownK060259 Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · Recalled
- Descendant · 2 generations on · 23 of its ownK081582 Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · Recalled
- Descendant · cites it directly · 22 of its ownK061163 Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova…Ge Medical Systems, LLC · 2006 · Recalled
- Descendant · 2 generations on · 21 of its ownK101868 Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C · 2010
Showing 5 of 75 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 75 · See all- K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare · 2005 · cites it directly74descendants
- K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · cites it directly47descendants
- K081582Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · 2 generations on23descendants
- K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ…Ge Medical Systems, LLC · 2006 · cites it directly22descendants
- K101868Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C · 2010 · 2 generations on21descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K890348 | LC, Cardiac Positioner #B5078A | General Electric Co. | 1989 | 5 |
| −2 | K993037 | LCV+ Version 2 | Ge Medical Systems, Inc. | 2000 | 6 |
| −1 | K021021 | Modular Angiography System Axiom Artis | Siemens Medical Solutions USA, Inc. | 2002 | 21 |
| 0 (this device) | K033244 | Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table Option | Ge Medical Systems, Inc. | 2003 | — |
| +1 | K052412 | Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQ | GE Healthcare | 2005 | 50 |
| +1 | K060259 | Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121 | Ge Medical Systems, LLC | 2006 | 4 |
| +2 | K081582 | Infx-8000V, Infinix-Cfi and Infinix Vfi | Toshiba America Medical Systems, In.C | 2008 | 1 |
| +1 | K061163 | Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova… | Ge Medical Systems, LLC | 2006 | 38 |
| +2 | K101868 | Infinix-I Interventional Angiography System Model Infx-8000V | Toshiba America Medical Systems, In.C | 2010 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1993 | 1 | 0 |
| 1995 | 1 | 0 |
| 2001 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 0 |
| 2006 | 2 | 0 |
| 2008 | 4 | 0 |
| 2009 | 3 | 0 |
| 2010 | 3 | 0 |
| 2011 | 4 | 0 |
| 2012 | 3 | 0 |
| 2013 | 2 | 0 |
| 2014 | 4 | 0 |
| 2015 | 6 | 0 |
| 2016 | 4 | 0 |
| 2017 | 4 | 0 |
| 2018 | 10 | 0 |
| 2020 | 3 | 0 |
| 2021 | 3 | 0 |
| 2022 | 4 | 0 |
| 2023 | 3 | 0 |
| 2024 | 5 | 0 |
| 2025 | 4 | 0 |
| 2026 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OWB device
- #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K021021 | Modular Angiography System Axiom ArtisSummary page 1…dical System’s Innova 4100, with a new Tilt table option for patient positioning, for diagnostic fluoroscopic imaging is substantially equivalent to Siemens Axiom Artis (K021021). This opinion is based on the information contained in the comparison table (Attachment 3), and the product data sheets (Attachm… | 2002 |
What descends from it? (75)
See all 753 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Dec 11, 2013 | Class II | Z-1163-2014 | Software design | GE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of real-time interventional… |
| Jan 21, 2013 | Class II | Z-0947-2013 | Nonconforming Material/Component | A locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems. |
| Oct 21, 2011 | Class II | Z-1142-2013 | Labeling design | As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff… |
| May 16, 2011 | Class II | Z-1616-2013 | Software design | GE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems. |
| Jun 3, 2010 | Not classified | Z-1922-2010 | Device Design | GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. |
| Jun 3, 2010 | Not classified | Z-1923-2010 | Device Design | GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. |
| Jun 3, 2010 | Not classified | Z-1921-2010 | Device Design | GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. |
| Apr 16, 2009 | Not classified | Z-0137-2010 | Device Design | Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room… |
The 5 latest of 8 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 2.3 years on median.
- 4 of the 6 ancestors have a recall of any class.
- Of its descendants, K052412 (GE Healthcare, 2005) has gone furthest, with 74 descendants of its own.
- 4 other devices also cite its predicate K021021.
How does its category compare?
All OWB devicesOWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · OWB
22in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OWB
148in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OWB
42so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OWB
14,008in 2025
MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 52 | 0 | 1 |
| 2008 | 4 | 135 | 1 | 0 |
| 2009 | 9 | 99 | 0 | 0 |
| 2010 | 7 | 66 | 0 | 0 |
| 2011 | 6 | 79 | 0 | 1 |
| 2012 | 6 | 37 | 0 | 1 |
| 2013 | 11 | 119 | 10 | 27 |
| 2014 | 15 | 158 | 9 | 103 |
| 2015 | 11 | 66 | 35 | 2,134 |
| 2016 | 16 | 103 | 13 | 4,074 |
| 2017 | 20 | 76 | 13 | 1,604 |
| 2018 | 23 | 54 | 6 | 1,449 |
| 2019 | 9 | 174 | 7 | 1,558 |
| 2020 | 16 | 53 | 8 | 1,581 |
| 2021 | 24 | 60 | 31 | 1,981 |
| 2022 | 21 | 90 | 24 | 2,932 |
| 2023 | 12 | 152 | 22 | 7,604 |
| 2024 | 13 | 126 | 11 | 7,942 |
| 2025 | 22 | 148 | 30 | 14,008 |
| 2026 | 15 | 221 | 42 | — |
What is it indicated for?
The Digital Fluoroscopic Imaging System is indicated for use in diagnostic and
interventional angiographic procedures of human anatomy. It is intended to replace image
intensifier fluoroscopic systems in all diagnostic or interventionnal procedures. This device is
not intended for mammography applications.
The new tilt table will support performing procedures like CO, studies, Venography.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K033244 · printed Nov 3, 2026
Cite this page
510(k) Report. “K033244: Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table Option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K033244