510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table Option

K033244510(k) TraditionalCleared Oct 16, 20039-day reviewClass II21 CFR 892.1650Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Oct 16, 2003 as substantially equivalent to K021021 (Siemens Medical Solutions USA, Inc., 2002) after a 9-day review. Its family tree reaches back 3 generations to 1989. 75 later devices descend from it. It has been recalled 8 times.

Decided
Oct 16, 2003
Review
9 days
Class
Class II
Product code
OWB
Recalls
8
Predicates
1

Its longest line of descent runs back 3 generations to K890348 (1989). 6 ancestors in all. 75 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1995: 1 other ancestor1993: 1 other ancestor2006: 2 descendants cleared2005: 1 descendant cleared2010: 3 descendants cleared2009: 3 descendants cleared2016: 4 descendants cleared2008: 4 descendants cleared2011: 4 descendants cleared2018: 10 descendants cleared2012: 3 descendants cleared2013: 2 descendants cleared2022: 4 descendants cleared2017: 4 descendants cleared2015: 6 descendants cleared2014: 4 descendants cleared2021: 3 descendants cleared2024: 5 descendants cleared2020: 3 descendants cleared2025: 4 descendants cleared2023: 3 descendants cleared2026: 3 descendants clearedK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK021021 · Modular Angiography System Axiom Artis · Siemens Medical Solutions USA, Inc. · 2002 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 8 descendants3 generations forward: 12 descendants4 generations forward: 13 descendants5 generations forward: 11 descendants6 generations forward: 8 descendants7 generations forward: 3 descendants8 generations forward: 5 descendants9 generations forward: 4 descendants10 generations forward: 3 descendants11 generations forward: 2 descendants12 generations forward: 2 descendants13 generations forward: 1 descendant K021021 · Modular Angiography System Axiom Artis · Siemens Medical Solutions USA, Inc. · 2002 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalled0+5+10
  1. Root · 3 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 2 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. 1 generation backK021021 Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc. · 2002 · Recalled
  4. This deviceK033244 Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt…Ge Medical Systems, Inc. · 2003
  5. Descendant · cites it directly · 74 of its ownK052412 Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova…GE Healthcare · 2005 · Recalled
  6. Descendant · cites it directly · 47 of its ownK060259 Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · Recalled
  7. Descendant · 2 generations on · 23 of its ownK081582 Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · Recalled
  8. Descendant · cites it directly · 22 of its ownK061163 Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova…Ge Medical Systems, LLC · 2006 · Recalled
  9. Descendant · 2 generations on · 21 of its ownK101868 Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C · 2010

Showing 5 of 75 descendants. See all descendants

Descendants cleared per year

2005: 12006: 22008: 42009: 32010: 32011: 42012: 32013: 22014: 42015: 62016: 42017: 42018: 102020: 32021: 32022: 42023: 32024: 52025: 42026: 3
2005high 10 (2018) · low 0 (2007)2026

Most influential descendants

Showing 5 of 75 · See all
  • K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare · 2005 · cites it directly74descendants
  • K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · cites it directly47descendants
  • K081582Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · 2 generations on23descendants
  • K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ…Ge Medical Systems, LLC · 2006 · cites it directly22descendants
  • K101868Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C · 2010 · 2 generations on21descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.19895
−2K993037LCV+ Version 2Ge Medical Systems, Inc.20006
−1K021021Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc.200221
0 (this device)K033244Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table OptionGe Medical Systems, Inc.2003—
+1K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare200550
+1K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC20064
+2K081582Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C20081
+1K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova…Ge Medical Systems, LLC200638
+2K101868Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199510
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200620
200840
200930
201030
201140
201230
201320
201440
201560
201640
201740
2018100
202030
202130
202240
202330
202450
202540
202630
FIG. 1 — Lineage of K033244, 1989–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K021021Modular Angiography System Axiom Artis
Summary page 1
…dical System’s Innova 4100, with a new Tilt table option for patient positioning, for diagnostic fluoroscopic imaging is substantially equivalent to Siemens Axiom Artis (K021021). This opinion is based on the information contained in the comparison table (Attachment 3), and the product data sheets (Attachm…
Siemens Medical Solutions USA, Inc.200221

Also cleared against K021021: 4 other devices →

What descends from it? (75)

See all 75

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare200574
K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC200647
K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ…Ge Medical Systems, LLC200622

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 11, 2013Class IIZ-1163-2014Software designGE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of real-time interventional…
Jan 21, 2013Class IIZ-0947-2013Nonconforming Material/ComponentA locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems.
Oct 21, 2011Class IIZ-1142-2013Labeling designAs part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff…
May 16, 2011Class IIZ-1616-2013Software designGE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.
Jun 3, 2010Not classifiedZ-1922-2010Device DesignGE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety.
Jun 3, 2010Not classifiedZ-1923-2010Device DesignGE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety.
Jun 3, 2010Not classifiedZ-1921-2010Device DesignGE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety.
Apr 16, 2009Not classifiedZ-0137-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 2.3 years on median.
  • 4 of the 6 ancestors have a recall of any class.
  • Of its descendants, K052412 (GE Healthcare, 2005) has gone furthest, with 74 descendants of its own.
  • 4 other devices also cite its predicate K021021.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Digital Fluoroscopic Imaging System is indicated for use in diagnostic and

interventional angiographic procedures of human anatomy. It is intended to replace image

intensifier fluoroscopic systems in all diagnostic or interventionnal procedures. This device is

not intended for mammography applications.

The new tilt table will support performing procedures like CO, studies, Venography.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K033244: Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table Option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K033244

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026