510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Artiste MV SA

K072485510(k) TraditionalCleared Dec 27, 2007114-day reviewClass II21 CFR 892.5050Recalls 24

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IYE Accelerator, Linear, Medical

Cleared on Dec 27, 2007 as substantially equivalent to K060226 (Siemens Medical Solutions USA, Inc., 2006) after a 114-day review. Its family tree reaches back 8 generations to 1992. 11 later devices descend from it. It has been recalled 24 times.

Decided
Dec 27, 2007
Review
114 days
Class
Class II
Product code
IYE
Recalls
24
Predicates
1

Its longest line of descent runs back 8 generations to K914698 (1992). 29 ancestors in all, the oldest K862339 (1986). 11 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 2 other ancestors1999: 1 other ancestor2001: 2 other ancestors1997: 1 other ancestor1995: 2 other ancestors1990: 2 other ancestors1988: 1 other ancestor1986: 1 other ancestor1992: 1 other ancestor1991: 1 other ancestor1998: 2 other ancestors1996: 1 other ancestor1994: 2 other ancestors1993: 1 other ancestor1989: 1 other ancestor2010: 2 descendants cleared2011: 1 descendant cleared2009: 3 descendants cleared2013: 1 descendant cleared2014: 2 descendants cleared2012: 1 descendant cleared2018: 1 descendant clearedK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K031764 · Oncor Avant-Garde with Coherence Workspaces · Siemens Medical Solutions USA, Inc. · 2003 · recalledK060226 · Oncor Expression · Siemens Medical Solutions USA, Inc. · 2006 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 3 other ancestors3 generations back: 7 other ancestors4 generations back: 2 other ancestors5 generations back: 1 other ancestor6 generations back: 4 other ancestors7 generations back: 4 other ancestors 1 generation forward: 6 descendants2 generations forward: 3 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant K060226 · Oncor Expression · Siemens Medical Solutions USA, Inc. · 2006 · recalledK031764 · Oncor Avant-Garde with Coherence Workspaces · Siemens Medical Solutions USA, Inc. · 2003 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992−50
  1. Root · 8 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 7 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 6 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
  4. 5 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
  5. 4 generations backK013112 Focal SimComputerized Medical Systems, Inc. · 2001 · Recalled
  6. 3 generations backK022036 Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc. · 2002
  7. 2 generations backK031764 Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  8. 1 generation backK060226 Oncor ExpressionSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  9. This deviceK072485 Artiste MV SASiemens Medical Solutions USA, Inc. · 2007
  10. Descendant · cites it directly · 9 of its ownK082775 Prescision OptionSiemens Medical Solutions USA, Inc. · 2009
  11. Descendant · cites it directly · 7 of its ownK090683 Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · Recalled
  12. Descendant · cites it directly · 5 of its ownK102671 Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc. · 2010
  13. Descendant · cites it directly · 4 of its ownK103606 Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc. · 2011 · Recalled
  14. Descendant · cites it directly · 4 of its ownK092145 160MLC OptionSiemens Medical Solutions USA, Inc. · 2009 · Recalled

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

2009: 32010: 22011: 12012: 12013: 12014: 22018: 1
2009high 3 (2009) · low 0 (2015)2026

Most influential descendants

Showing 5 of 11 · See all
  • K082775Prescision OptionSiemens Medical Solutions USA, Inc. · 2009 · cites it directly9descendants
  • K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · cites it directly7descendants
  • K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc. · 2010 · cites it directly5descendants
  • K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc. · 2011 · cites it directly4descendants
  • K092145160MLC OptionSiemens Medical Solutions USA, Inc. · 2009 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−7K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−6K990833Dose Calc, Model VS. 1.02Integrity Medical Software, Inc.1999—
−5K993923Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories2000—
−4K013112Focal SimComputerized Medical Systems, Inc.200110
−3K022036Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc.2002—
−2K031764Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc.200317
−1K060226Oncor ExpressionSiemens Medical Solutions USA, Inc.200672
0 (this device)K072485Artiste MV SASiemens Medical Solutions USA, Inc.2007—
+1K082775Prescision OptionSiemens Medical Solutions USA, Inc.2009—
+1K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc.20094
+1K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc.2010—
+1K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc.20117
+1K092145160MLC OptionSiemens Medical Solutions USA, Inc.200911

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198810
198910
199020
199110
199210
199310
199420
199520
199610
199710
199820
199910
200120
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200930
201020
201110
201210
201310
201420
201810
FIG. 1 — Lineage of K072485, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060226Oncor Expression
Summary page 1
…SATM Device classification: 892.5050/ IYE/Radiology Substantial equivalence claimed to: ONCOR EXPRESSION with COHERENCE Workspaces (K060226) Device Description: The ACCEL system is the internal name for the Siemens program which includes the ONCOR Linear Acce…
Siemens Medical Solutions USA, Inc.200672

Also cleared against K060226: 6 other devices →

What descends from it? (11)

See all 11

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K082775Prescision OptionSiemens Medical Solutions USA, Inc.20099
K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc.20097
K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc.20105
K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc.20114
K092145160MLC OptionSiemens Medical Solutions USA, Inc.20094
K100096Linac Collimator Assembly Housing (Lcah) Model LcahartIntegra Radionics, Inc.20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 18, 2020Class IIZ-1677-2020Use errorPotential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
Mar 20, 2019Class IIZ-1204-2019Use errorAdvisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field…
Feb 14, 2017Class IIZ-1488-2017Software designSoftware update
Jan 25, 2017Class IIZ-1212-2017Device DesignSiemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension.
Dec 19, 2014Class IIZ-1059-2015Component design/selectionA hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke…
Dec 2, 2014Class IIZ-0980-2015Component design/selectionUse of any unauthorized third-party components on Siemens LINACs may lead to patient or operator injury, damage to the LINAC, and can potentially compromise customer warranty. The dark current dose must be considered in the overall dose…
Jun 17, 2014Class IIZ-2180-2014Software designSiemens became aware of a potential issue with the syngo RT Therapist 4.3.SP1 and syngo RT Oncologist 4.3SP1 software applications running on Siemens brand LINAC systems that may experience sporadic system crashes that will cause the…
May 28, 2014Class IIZ-2174-2014Use errorA safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision…
Apr 1, 2014Class IIZ-1652-2014Labeling designSiemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models.
Apr 1, 2014Class IIZ-1645-2014Software designTable may lose calibration during patient treatment, which may cause the possibility of mistreatment.
Oct 23, 2013Class IIZ-0438-2014Device DesignA safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there…
Jan 22, 2012Not classifiedZ-1092-2012Software designSoftware upgrade to installed base to the latest release of the RT Therapist version 4.1.144 which corrects multiple safety issues that could lead to patient mistreatment.
Jun 3, 2011Not classifiedZ-3003-2011Software designA potential safety issue was discovered when a pause for an imaging segment is inserted (where an imaging segment does not exist) with the ARTISTE and RT Therapist 4.1 system will continue to the next image.
Dec 3, 2010Not classifiedZ-1422-2011OtherThe firm become aware of a potential issue with ARTISTE system that may result in patient injury should the positioner belt break, and the cam disc (secondary safety mechanism) is not installed and if the flat panel is over the patient.
Jun 29, 2010Not classifiedZ-1523-2011OtherA software problem that has the potential for data loss during the transfer of treatment records to the OIS, which may not be recorded in MOSAIQ, and subsequently may lead to mistreatment.
Jan 22, 2010Not classifiedZ-0449-2011OtherThe robotic lift armature shaft may break and allow the Flat Panel Positioner with the digital flat panel imager to fall into the Flat Panel cradle. If the positioner is over a patient, there is a potential for injury.
Sep 10, 2009Not classifiedZ-0121-2010Software designA failed data transfer is possible that may result in data not being recorded in the interfaced system. Siemens Medical Solutions, Inc. USA became aware of this potential issue with its ARTISTE Linear Accelerator's Syngo RT Therapist v4.1…
Aug 21, 2009Not classifiedZ-1970-2011Software designWhen editing the note type of a previously entered note, then reverting to the original type, data may be lost.
Jul 30, 2009Not classifiedZ-1064-2011Software designPotential for Mistreatment-- when treatment plans intended for non-ARTISTE machinese are loaded, if the plans contain unspecified beam parameters, attributes or unknown block codes, the therapy system may remove those parameters.
Apr 9, 2009Not classifiedZ-1888-2009Software designUnexpected treatment plan alteration-- Software issue may result in unintended modifications to treatment
Dec 20, 2008Not classifiedZ-0949-2009Device DesignThe range of tolerance is to high for the accessory holder, therefore, Wedges and other accessories do not latch to the locking mechanism correctly.
Oct 23, 2008Not classifiedZ-0857-2009Labeling designUnder certain conditions, the flat panel may move unexpectedly and collide with the patient.
Oct 23, 2008Not classifiedZ-0927-2009Software designIf a plan contains a sequence of beams, including at least 1 ARC treatment beam and if during the setup stage of a beam following an ARC beam in the delivery sequence a termination occurs, and the user selects the immediate resumption…
Jul 4, 2008Not classifiedZ-0116-2009Software change controlThree issues: unexpected rotation of gantry, unexpected movement of table between beams during patient setup, and single exposure images will be overwritten by double exposure images.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 6 companies; a new generation was cleared every 1.7 years on median.
  • 12 of the 29 ancestors have a recall of any class.
  • Of its descendants, K082775 (Siemens Medical Solutions USA, Inc., 2009) has gone furthest, with 9 descendants of its own.
  • 6 other devices also cite its predicate K060226.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The intended use of the ARTISTE MV SAT M linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer. The ARTISTE MV SATM includes an Electronic Portal Imaging Device (EPID), a 160 leaf multi-leaf collimator, and the syngoTM RT Therapist Express Workspace with MVisionTM. The syngoTM RT Therapist Express Workspace is a software application that uses syngoTM based applications with these embedded features: • patient data…

Read the full statement (126 words)

The intended use of the ARTISTE MV SAT M linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.

The ARTISTE MV SATM includes an Electronic Portal Imaging Device (EPID), a 160 leaf multi-leaf collimator, and the syngoTM RT Therapist Express Workspace with MVisionTM.

The syngoTM RT Therapist Express Workspace is a software application that uses syngoTM based applications with these embedded features:

• patient data management, selection and setup by use of a local, on-board database,

• treatment delivery/verification and treatment recording by use of on-board applications,

• Megavoltage Cone Beam acquisition method for acquiring projection data for 3D reconstruction.

• Patient positioning verification by use of the OPTIVUE imaging system, including MVisionTM(Cone Beam acquisition and reconstruction and the Adaptive Targeting application).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072485: Artiste MV SA.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K072485

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026