K060226510(k) TraditionalCleared Mar 15, 200644-day reviewClass II21 CFR 892.5050Recalls 72

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IYE Accelerator, Linear, Medical

Cleared on Mar 15, 2006 as substantially equivalent to K031764 (Siemens Medical Solutions USA, Inc., 2003) and 3 other predicates after a 44-day review. Its family tree reaches back 7 generations to 1992. 13 later devices descend from it. It has been recalled 72 times.

Decided
Mar 15, 2006
Review
44 days
Class
Class II
Product code
IYE
Recalls
72
Predicates
4

Its longest line of descent runs back 7 generations to K914698 (1992). 28 ancestors in all, the oldest K862339 (1986). 13 later devices descend from it, 7 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 2 other ancestors1999: 1 other ancestor2001: 2 other ancestors1997: 1 other ancestor1995: 2 other ancestors1990: 2 other ancestors1988: 1 other ancestor1986: 1 other ancestor1992: 1 other ancestor1991: 1 other ancestor1998: 2 other ancestors1996: 1 other ancestor1994: 2 other ancestors1993: 1 other ancestor1989: 1 other ancestor2010: 2 descendants cleared2009: 4 descendants cleared2007: 1 descendant cleared2013: 1 descendant cleared2011: 1 descendant cleared2014: 2 descendants cleared2012: 1 descendant cleared2018: 1 descendant clearedReference device: K010938 · Syngo Multimodality Workstation · Siemens Medical Solutions USA, Inc. · 2001 · recalledReference device: K972275 · Lantis Treatstation · Siemens Medical Solutions USA, Inc. · 1997 · recalledK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K031764 · Oncor Avant-Garde with Coherence Workspaces · Siemens Medical Solutions USA, Inc. · 2003 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 7 other ancestors3 generations back: 2 other ancestors4 generations back: 1 other ancestor5 generations back: 4 other ancestors6 generations back: 4 other ancestors 1 generation forward: 7 descendants2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant K031764 · Oncor Avant-Garde with Coherence Workspaces · Siemens Medical Solutions USA, Inc. · 2003 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992−50
  1. Root · 7 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 6 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 5 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
  4. 4 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
  5. 3 generations backK013112 Focal SimComputerized Medical Systems, Inc. · 2001 · Recalled
  6. 2 generations backK022036 Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc. · 2002
  7. 1 generation backK031764 Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  8. This deviceK060226 Oncor ExpressionSiemens Medical Solutions USA, Inc. · 2006
  9. Descendant · cites it directly · 11 of its ownK072485 Artiste MV SASiemens Medical Solutions USA, Inc. · 2007 · Recalled
  10. Descendant · cites it directly · 9 of its ownK082775 Prescision OptionSiemens Medical Solutions USA, Inc. · 2009
  11. Descendant · cites it directly · 7 of its ownK090683 Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · Recalled
  12. Descendant · cites it directly · 5 of its ownK102671 Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc. · 2010
  13. Descendant · 2 generations on · 4 of its ownK103606 Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc. · 2011 · Recalled

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2007: 12009: 42010: 22011: 12012: 12013: 12014: 22018: 1
2007high 4 (2009) · low 0 (2008)2026

Most influential descendants

Showing 5 of 13 · See all
  • K072485Artiste MV SASiemens Medical Solutions USA, Inc. · 2007 · cites it directly11descendants
  • K082775Prescision OptionSiemens Medical Solutions USA, Inc. · 2009 · cites it directly9descendants
  • K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · cites it directly7descendants
  • K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc. · 2010 · cites it directly5descendants
  • K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc. · 2011 · 2 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−6K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−5K990833Dose Calc, Model VS. 1.02Integrity Medical Software, Inc.1999—
−4K993923Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories2000—
−3K013112Focal SimComputerized Medical Systems, Inc.200110
−2K022036Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc.2002—
−1K031764Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc.200317
0 (this device)K060226Oncor ExpressionSiemens Medical Solutions USA, Inc.2006—
+1K072485Artiste MV SASiemens Medical Solutions USA, Inc.200724
+1K082775Prescision OptionSiemens Medical Solutions USA, Inc.2009—
+1K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc.20094
+1K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc.2010—
+2K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc.20117

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198810
198910
199020
199110
199210
199310
199420
199520
199610
199710
199820
199910
200120
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200940
201020
201110
201210
201310
201420
201810
FIG. 1 — Lineage of K060226, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031764Oncor Avant-Garde with Coherence Workspaces
Summary page 1
…t Code: lYE Substantial Equivalence Claimed To: PRODUCT Clearance Claim of Equivalence For: ONCOR Avant-Garde with K031764 ONCOR Expression linear accelerator with COHERENCE COHERENCE Workspaces Workspaces, includes the…
Siemens Medical Solutions USA, Inc.200317
*K993425Primus
Summary page 1
…PRIMEVIEW3i is the version for use with the PRIMUS product line. PRIMUS K993425 ONCOR Expression linear accelerator LEONARDO K040970 Syngo based applications on COHERENCE Wo…
Siemens Medical Solutions USA, Inc.199950
*K040970Leonardo
Summary page 1
…product line. PRIMUS K993425 ONCOR Expression linear accelerator LEONARDO K040970 Syngo based applications on COHERENCE Workspaces Siemens Siremobile Iso-C 3D K040347 3D Cone Beam Reconstruction…
Siemens Medical Systems, Inc.20041
*K040347Siremobil Iso-C 3D
Summary page 1
…ONCOR Expression linear accelerator LEONARDO K040970 Syngo based applications on COHERENCE Workspaces Siemens Siremobile Iso-C 3D K040347 3D Cone Beam Reconstruction Specifically 3D reconstruction application. Description Summary: Within the submission th…
Siemens Medical Systems, Inc.20041
†K010938Syngo Multimodality WorkstationSiemens Medical Solutions USA, Inc.20015
†K972275Lantis TreatstationSiemens Medical Solutions USA, Inc.19975

Also cleared against K031764: 2 other devices →

What descends from it? (13)

See all 13

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072485Artiste MV SASiemens Medical Solutions USA, Inc.200711
K082775Prescision OptionSiemens Medical Solutions USA, Inc.20099
K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc.20097
K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc.20105
K092145160MLC OptionSiemens Medical Solutions USA, Inc.20094
K123812Coherence RT Therapist Connect Workspace, V2.3, Aria Connectivity Option for Siemens…Siemens Medical Solutions USA, Inc.20133
K083502Panther Realart, Model 4.7Prowess, Inc.20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 10, 2019Class IIZ-2198-2019OtherPotentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability
Mar 20, 2019Class IIZ-1204-2019Use errorAdvisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field…
Feb 14, 2018Class IIZ-1400-2018Software designControl Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
Feb 14, 2018Class IIZ-1399-2018Software designControl Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
Feb 14, 2017Class IIZ-1491-2017Software designSoftware update
Jan 25, 2017Class IIZ-1214-2017Device DesignSiemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension.
Aug 5, 2015Class IIZ-2812-2015Software design (manufacturing process)A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients.
May 5, 2015Class IIZ-0168-2016Software designpotential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option.
Dec 19, 2014Class IIZ-1059-2015Component design/selectionA hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke…
Nov 19, 2014Class IIZ-0984-2015Software designThere may be an existing dark current phenomenon on ARTISTE LINAC in combination with IMRT or mARC treatments using unflat beams. Software issue.
Oct 23, 2013Class IIZ-0437-2014Device DesignA safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there…
May 29, 2013Class IIZ-1663-2013Process change controlThe information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in mistreatment of patients.
Apr 4, 2013Class IIZ-1182-2013Software designSoftware update to fix multiple safety related issues.
Apr 4, 2012Not classifiedZ-1494-2012Process controlThe firm became aware of a potential issue with Siemens-branded ONCOR or PRIMUS linear accelerator with an OPTIVUE or MVision imaging system that has the potential collision problem of the imaging panel if the panel does not retract…
Apr 3, 2012Not classifiedZ-1486-2012Software designSIEMENS has received reports where a dropping down of the 550 TxT" Treatment Table was observed while the table was being moved downwards during operation by means of the hand control.
Jan 22, 2012Not classifiedZ-1092-2012Software designSoftware upgrade to installed base to the latest release of the RT Therapist version 4.1.144 which corrects multiple safety issues that could lead to patient mistreatment.
Nov 9, 2011Not classifiedZ-0459-2012Component design/selectionReports of gantry rotation without receiving a movement command caused by an overload within the drive signal power line.
Jun 3, 2011Not classifiedZ-3003-2011Software designA potential safety issue was discovered when a pause for an imaging segment is inserted (where an imaging segment does not exist) with the ARTISTE and RT Therapist 4.1 system will continue to the next image.
May 25, 2011Not classifiedZ-2789-2011Software designPotential safety issue of possible data loss for the transfer of the same patients treatment data between the LANTIS OIS and the RTT v2.x software and the RT Therapist version 4.1 system
Mar 31, 2011Not classifiedZ-2216-2011Software designSiemens has become aware of a potential safety issue that may result in mistreatment when customers are configured with the ARTISTE Linac RT Therapist v4.2 in combination with RT Oncologist v.4.2 and use the remote "Online 2D Image Review.
Mar 31, 2011Not classifiedZ-2243-2011Software designSiemens found a potential safety issue related to patient treatment with table offsets on syngo RT Therapist (RTT) 4.1 and 4.2 which could result in dose being delivered to wrong location.
Mar 24, 2011Not classifiedZ-2190-2011PendingWhen using the ARTISTE Linac and RT Therapist v4.1 along with an imported cropped reference image, the image maybe processed incorrectly, thus resulting in a mistreatment.
Aug 21, 2009Not classifiedZ-1970-2011Software designWhen editing the note type of a previously entered note, then reverting to the original type, data may be lost.
Mar 6, 2009Not classifiedZ-2000-2011Software change controlThe product has a software problem; it does not calculate the Creatinine Clearance correctly for patients less than two years old.
May 15, 2008Not classifiedZ-2100-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2097-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2098-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2104-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2096-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2095-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2101-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2103-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2102-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
May 15, 2008Not classifiedZ-2099-2008Software designSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
Jan 11, 2008Not classifiedZ-1400-2008Software designLinear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
Dec 10, 2007Not classifiedZ-1761-2008Device DesignRadiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
Dec 10, 2007Not classifiedZ-1760-2008Device DesignRadiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
Dec 10, 2007Not classifiedZ-1762-2008Device DesignRadiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
Dec 10, 2007Not classifiedZ-1763-2008Device DesignRadiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
Nov 6, 2007Not classifiedZ-0099-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0092-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0089-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0091-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0093-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0094-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0097-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0090-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0095-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0098-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Nov 6, 2007Not classifiedZ-0096-2009Software designFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 6 companies; a new generation was cleared every 12 months on median.
  • 11 of the 28 ancestors have a recall of any class.
  • Of its descendants, K072485 (Siemens Medical Solutions USA, Inc., 2007) has gone furthest, with 11 descendants of its own.
  • 2 other devices also cite its predicate K031764.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The intended use of the ONCOR Expression TM linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer. The ONCOR Expression includes an Electronic Portal Imaging Device (EPID) that will be marketed as OPTIVUETM 10OOST. The indications for use for the OPTIVUE 1000ST is the acquisition of portal images for the purpose of verification of the treatment field and shielding blocks in relation to patient positioning markers…

Read the full statement (295 words)

The intended use of the ONCOR Expression TM linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.

The ONCOR Expression includes an Electronic Portal Imaging Device (EPID) that will be marketed as OPTIVUETM 10OOST. The indications for use for the OPTIVUE 1000ST is the acquisition of portal images for the purpose of verification of the treatment field and shielding blocks in relation to patient positioning markers and/or anatomical landmarks in radiotherapy treatment. OPTIVUE 10OOOST will also allow for verification of the exit dose in radiotherapy treatment. The indications for use remain unchanged from the previously cleared OPTIVUE EPID (K031764).

Additionally, the ONCOR Expression includes as a standard feature, an 82 leaf multi-leaf collimator that is marketed as OPTIFOCUSTM. The OPTIFOCUS MVLG is provided to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation. The indications for use of the OPTIFOCUS MVLC remains unchanged from the previously cleared OPTIFOCUS MVLC (K031764).

The COHERENCETM Workspaces:

The COHERENCE Therapist workspace includes a number of syngoTM based software applications who's indication for use include the viewing, processing, filming, and archiving of medical images. The COHERENCE Therapist Workspace also permits patient data management, patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording.

The Therapist Workspace in this release includes MVisionTM which consists of the Megavoltage Cone Beam imaging method for acquiring 20 projection data that is reconstructed by the Cone Beam reconstruction method into 3D data for use by the Adaptive TargetingTM application for advanced treatment localization and patient positioning. Adaptive Targeting can also use 3D data imported from other devices. The indications for use for the COHERENCE Therapist workspace remains unchanged from the previously cleared COHERENCE Therapist workspace (K031764).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060226: Oncor Expression.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K060226

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026