GE Datex-Ohmeda Aestiva/5 Mri Anesthesia System

K050055510(k) SpecialCleared Feb 4, 200524-day reviewClass II21 CFR 868.5160Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Ohmeda, Inc. · BSZ Gas-Machine, Anesthesia

Cleared on Feb 4, 2005 after a 24-day review. The FDA summary names its predicate without a number. 26 of the 65 devices in its family, counting itself, have a Class I recall. 64 later devices descend from it. It has been recalled 2 times.

Decided
Feb 4, 2005
Review
24 days
Class
Class II
Product code
BSZ
Recalls
2
Predicates
0

64 later devices descend from it, 4 citing it directly. 26 of the 65 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2007: 1 descendant cleared2013: 7 descendants cleared, 3 with a Class I recall32017: 6 descendants cleared, 3 with a Class I recall32008: 4 descendants cleared, 1 with a Class I recall12010: 1 descendant cleared2016: 2 descendants cleared, 2 with a Class I recall22014: 3 descendants cleared, 1 with a Class I recall12015: 6 descendants cleared, 1 with a Class I recall12009: 3 descendants cleared, 1 with a Class I recall12011: 6 descendants cleared, 6 with a Class I recall62021: 5 descendants cleared, 2 with a Class I recall22024: 3 descendants cleared2012: 3 descendants cleared, 1 with a Class I recall12019: 4 descendants cleared, 1 with a Class I recall12026: 3 descendants cleared2020: 4 descendants cleared, 3 with a Class I recall32018: 1 descendant cleared2025: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall1Reference device: K973896 · Ohmeda Excel 3000 Anesthesia System · Ohmeda Medical · 1998 · recalledReference device: K993410 · Dates-Ohmeda Aestiva Smartvent Mri · Datex-Ohmeda, Inc. · 2000 · recalledReference device: K023366 · Datex-Ohmeda 7900 Ventilator Enhancements · Datex-Ohmeda, Inc. · 2003 · Class I recall20052010201520202025
  1. This deviceK050055 GE Datex-Ohmeda Aestiva/5 Mri Anesthesia SystemDatex-Ohmeda, Inc. · 2005
  2. Descendant · cites it directly · 61 of its ownK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
  3. Descendant · 2 generations on · 37 of its ownK081842 Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008
  4. Descendant · 2 generations on · 28 of its ownK073179 Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option…Maquet Critical Care AB · 2008
  5. Descendant · 2 generations on · 21 of its ownK093886 Engstrom Carestation, Engstrom ProDatex-Ohmeda Inc., A Division of Instrumentarium C · 2010
  6. Descendant · 3 generations on · 14 of its ownK103803 Hamilton-G5Hamilton Medical AG · 2011 · Recalled (Class I)

Showing 5 of 64 descendants. See all descendants

Descendants cleared per year

2007: 12008: 42009: 32010: 12011: 62012: 32013: 72014: 32015: 62016: 22017: 62018: 12019: 42020: 42021: 52023: 12024: 32025: 12026: 3
2007high 7 (2013) · low 0 (2022)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 64 · See all
  • K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007 · cites it directly61descendants
  • K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008 · 2 generations on37descendants
  • K073179Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option, Model 6675585Maquet Critical Care AB · 2008 · 2 generations on28descendants
  • K093886Engstrom Carestation, Engstrom ProDatex-Ohmeda Inc., A Division of Instrumentarium C · 2010 · 2 generations on21descendants
  • K103803Hamilton-G5Hamilton Medical AG · 2011 · 3 generations on14descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K050055GE Datex-Ohmeda Aestiva/5 Mri Anesthesia SystemDatex-Ohmeda, Inc.2005—
+1K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB2007—
+2K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc.2008—
+2K073179Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option, Model 6675585Maquet Critical Care AB2008—
+2K093886Engstrom Carestation, Engstrom ProDatex-Ohmeda Inc., A Division of Instrumentarium C2010—
+3K103803Hamilton-G5Hamilton Medical AG2011Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200841
200931
201010
201166
201231
201373
201431
201561
201622
201763
201810
201941
202043
202152
202311
202430
202510
202630
FIG. 1 — Lineage of K050055, 2005–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

26 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Datex-Ohmeda Aestiva SmartVent MRI Anesthesia System
Named without a submission number
—
†K973896Ohmeda Excel 3000 Anesthesia SystemOhmeda Medical19981
†K993410Dates-Ohmeda Aestiva Smartvent MriDatex-Ohmeda, Inc.20001
†K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc.2003Class I

What descends from it? (64)

See all 64

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB200761
K122438Hamilton-MR1Hamilton Medical AG20131
K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire… Class IDatex-Ohmeda, Inc.20170
K072884Fabius Mri Class IDraeger Medical AG & Co. KG20080

Has anything in its family been recalled? (26 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 19, 2019Class IIZ-0114-2020Software in the Use EnvironmentCertain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that…
Dec 15, 2015Class IIZ-0679-2016Nonconforming Material/ComponentThe Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900…Datex-Ohmeda, Inc. · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 110 Class I recalls
  • K072884Fabius MriDraeger Medical AG & Co. KG · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K153046Hamilton-MR1Hamilton Medical AG · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K131774Hamilton-G5Hamilton Medical AG · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · 3 generations on · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls
  • K103803Hamilton-G5Hamilton Medical AG · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K111116Engstrom Carestation, Engstrom Pro, Models 1505-9000-000, 1505-9003-000Datex-Ohmeda · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 13 Class I recalls
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K103211Avea VentilatorCare Fusion · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 114 Class I recalls
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall

Nearest 12 of 26 shown.

What does this lineage say?

  • Of its descendants, K063404 (Maquet Critical Care AB, 2007) has gone furthest, with 61 descendants of its own.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The Aestiva/5 MRI Anesthesia System provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva/5 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent along with the Aestiva breathing circuit. The Aestiva/5 MRI Anesthesia System performed to specifications when tested directly next to 1.5 and 3.0 Tesla shielded MRI devices in a field strength that did not exceed 300 gauss.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050055: GE Datex-Ohmeda Aestiva/5 Mri Anesthesia System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050055

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026