Medex 3000 Series MR Syringe Infusion Pump

K040899510(k) TraditionalCleared Apr 30, 200424-day reviewClass II21 CFR 880.5725Recalls 21 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medex, Inc. · FRN Pump, Infusion

Cleared on Apr 30, 2004 after a 24-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 4 of the 14 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 21 times, including 4 Class I recalls.

Decided
Apr 30, 2004
Review
24 days
Class
Class II
Product code
FRN
Recalls
21 (4 Class I)
Predicates
0

13 later devices descend from it, 3 citing it directly. 4 of the 14 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2006: 1 descendant cleared2011: 3 descendants cleared, 1 with a Class I recall12008: 1 descendant cleared2016: 2 descendants cleared2022: 1 descendant cleared, 1 with a Class I recall12024: 1 descendant cleared, 1 with a Class I recall12012: 2 descendants cleared2013: 1 descendant cleared2015: 1 descendant cleared200020052010201520202025
  1. This deviceK040899 Medex 3000 Series MR Syringe Infusion PumpMedex, Inc. · 2004
  2. Descendant · cites it directly · 6 of its ownK053554 CompufloMilestone Scientific, Inc. · 2006
  3. Descendant · 2 generations on · 5 of its ownK101194 Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011
  4. Descendant · 3 generations on · 4 of its ownK112067 Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011
  5. Descendant · cites it directly · 4 of its ownK111386 Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2…Smiths Medical MD, Inc. · 2011 · Recalled (Class I)
  6. Descendant · 4 generations on · 3 of its ownK120830 Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2012

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2006: 12008: 12011: 32012: 22013: 12015: 12016: 22022: 12024: 1
2006high 3 (2011) · low 0 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K053554CompufloMilestone Scientific, Inc. · 2006 · cites it directly6descendants
  • K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011 · 2 generations on5descendants
  • K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011 · 3 generations on4descendants
  • K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety…Smiths Medical MD, Inc. · 2011 · cites it directly4descendants
  • K120830Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2012 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K040899Medex 3000 Series MR Syringe Infusion PumpMedex, Inc.2004—
+1K053554CompufloMilestone Scientific, Inc.2006—
+2K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc.2011—
+3K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc.2011—
+1K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication…Smiths Medical MD, Inc.2011Class I
+4K120830Navigator Delivery System (Navigator DS)Carticept Medical, Inc.2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200810
201131
201220
201310
201510
201620
202211
202411
FIG. 1 — Lineage of K040899, 2004–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (13)

See all 13

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K053554CompufloMilestone Scientific, Inc.20066
K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety… Class ISmiths Medical MD, Inc.20114
K072603Abc Syringe Infusion Pump, Model 4100Elixir Corporation20080

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 19, 2023Class IZ-0863-2024Software Design ChangeMedfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusion Restarted with Incorrect Parameters, 3. Screen Lock, 4. Interruption of…
Dec 19, 2023Class IIZ-1540-2024Process controlMedfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the…
Dec 19, 2023Class IIZ-1542-2024Process controlMedfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the…
Aug 16, 2023Class IIZ-0078-2024Under Investigation by firmA force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will…
Aug 16, 2023Class IIZ-0076-2024Under Investigation by firmA force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will…
Aug 16, 2023Class IIZ-0077-2024Under Investigation by firmA force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will…
Apr 18, 2022Class IZ-1272-2022Device DesignMultiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues…
Jun 26, 2020Class IZ-2734-2020Software designInaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
Jun 26, 2020Class IZ-2735-2020Software designInaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
Apr 13, 2018Class IIZ-2662-2018Process change controlCertain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded syringe.
Nov 13, 2017Class IIZ-1136-2018Process change controlCertain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump)…
Nov 13, 2017Class IIZ-1135-2018Process change controlCertain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump)…
Aug 22, 2017Class IIZ-0314-2018Component design/selectionPotential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid Syringe Size" alarm.
Dec 3, 2015Class IIZ-0653-2022Component change controlThere have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall…
Jan 2, 2012Not classifiedZ-1206-2012Vendor change controlPumps rendered inoperable at power up (displaying a blank screen and alarm).
Jun 22, 2010Not classifiedZ-2121-2010Software designSmiths Medical is conducting a correction of Medfusion Syringe Infusion Pumps, Models 3010 and 3010a, with obsolete software versions 2.0.2, 2.0.3, and 2.0.4. Smiths Medical has found that pumps with obsolete software may continue to run…
May 14, 2010Class IIZ-0996-2013Device DesignAn issue with the plunger rod seal was identified where the seal had the potential to migrate inside the pump case top during slow rate/ high back pressure infusions.
Nov 16, 2009Not classifiedZ-0650-2010Device DesignSmiths Medical has become aware of an increased trend in reports of Motor Not Running (MNR) and Motor Rate Error (MRE) alarm message events with certain Medfusion Syringe Infusion Pumps, Models 3010A and 3500 (Pump). MNR/ MRE are high…
Nov 16, 2009Not classifiedZ-0651-2010Device DesignSmiths Medical has become aware of an increased trend in reports of Motor Not Running (MNR) and Motor Rate Error (MRE) alarm message events with certain Medfusion Syringe Infusion Pumps, Models 3010A and 3500 (Pump). MNR/ MRE are high…
Dec 18, 2008Not classifiedZ-1019-2009Nonconforming Material/ComponentDelay in therapy. Keypad issues with Medfusion Syringe Infusion Pump, model numbers 3500-306 and 3500-415 ("Pump") have occurred during implementation or shortly after first being placed into use. These issues can be intermittent, and may…
Dec 18, 2008Not classifiedZ-1018-2009Nonconforming Material/ComponentDelay in therapy. Keypad issues with Medfusion Syringe Infusion Pump, model numbers 3500-306 and 3500-415 ("Pump") have occurred during implementation or shortly after first being placed into use. These issues can be intermittent, and may…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety Software and Supported…Smiths Medical MD, Inc. · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 16 Class I recalls
  • K211125Novum IQ Syringe PumpBaxter Healthcare Corporation · 2022 · 2 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall
  • K242390Novum IQ Syringe Pump (40800BAXUS)Baxter Healthcare Corporation · 2024 · 3 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K053554 (Milestone Scientific, Inc., 2006) has gone furthest, with 6 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Medex 3000 Series MRI Syringe Infusion Pump indications for use are as follows: · in the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids. * in the following delivery routes: arterial, epidural, intravenous, spinal, subcutaneous, and enteral. * in the following delivery modes: continuous, volume/time, mass, body weight…

Read the full statement (158 words)

The Medex 3000 Series MRI Syringe Infusion Pump indications for use are as follows:

· in the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids.

* in the following delivery routes: arterial, epidural, intravenous, spinal, subcutaneous, and enteral.

* in the following delivery modes: continuous, volume/time, mass, body weight, custom dilution, intermittent and bolus.

· in critical care, anesthesia, neonatal and pediatric applications or other healthcare settings where the use of the syringe infusion pump can be monitored or supervised by a clinician.

* inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla or less.

The syringe infusion pump is contraindicated for use on the inlet side of Extracorporeal Membrane Oxygenation (ECMO) systems where the negative pressure is greater than -100mm Hg as the high negative pressures can result in uncontrolled fluid flow.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K040899: Medex 3000 Series MR Syringe Infusion Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K040899

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026