Recalled · Class I · 2026Medley Syringe Pump Module System, Model 8110
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Alaris Medical Systems, Inc. · FRN Pump, Infusion
Cleared on Dec 19, 2002 as substantially equivalent to K974332 (Alaris Medical Systems, Inc., 1998) and 1 other predicate after an 80-day review. Its family tree reaches back 2 generations to 1994. 7 of the 29 devices in its family, counting itself, have a Class I recall. 21 later devices descend from it. It has been recalled 7 times, including a Class I recall.
- Decided
- Dec 19, 2002
- Review
- 80 days
- Class
- Class II
- Product code
- FRN
- Recalls
- 7 (1 Class I)
- Predicates
- 2
Its longest line of descent runs back 2 generations to K940147 (1994). 7 ancestors in all, the oldest K890120 (1989). 21 later devices descend from it, 4 citing it directly. 7 of the 29 devices in this family, counting this one, have a Class I recall.
- Root · 2 generations backK940147 Auto Syringe AS40A Infusion PumpBaxter Corp. · 1994 · Recalled
- 1 generation backK974332 Variable Pressure Syringe PumpAlaris Medical Systems, Inc. · 1998
- This deviceK023264 Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002
- Descendant · cites it directly · 6 of its ownK053554 CompufloMilestone Scientific, Inc. · 2006
- Descendant · 2 generations on · 5 of its ownK101194 Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011
- Descendant · 3 generations on · 4 of its ownK112067 Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011
- Descendant · cites it directly · 4 of its ownK111386 Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2…Smiths Medical MD, Inc. · 2011 · Recalled (Class I)
- Descendant · cites it directly · 4 of its ownK062699 Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc. · 2006
Showing 5 of 21 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 21 · See all- K053554CompufloMilestone Scientific, Inc. · 2006 · cites it directly6descendants
- K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011 · 2 generations on5descendants
- K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011 · 3 generations on4descendants
- K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety…Smiths Medical MD, Inc. · 2011 · cites it directly4descendants
- K062699Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc. · 2006 · cites it directly4descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K940147 | Auto Syringe AS40A Infusion Pump | Baxter Corp. | 1994 | 1 |
| −1 | K974332 | Variable Pressure Syringe Pump | Alaris Medical Systems, Inc. | 1998 | — |
| 0 (this device) | K023264 | Medley Syringe Pump Module System, Model 8110 | Alaris Medical Systems, Inc. | 2002 | — |
| +1 | K053554 | Compuflo | Milestone Scientific, Inc. | 2006 | — |
| +2 | K101194 | Navigator Delivery System (or Navigator DS) Model: Nav-010 | Carticept Medical, Inc. | 2011 | — |
| +3 | K112067 | Navigator Delivery System (Navigator DS) | Carticept Medical, Inc. | 2011 | — |
| +1 | K111386 | Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication… | Smiths Medical MD, Inc. | 2011 | Class I |
| +1 | K062699 | Perfusor Space Infusion Syringe Pump System | B.Braun Medical, Inc. | 2006 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1989 | 1 | 0 |
| 1990 | 1 | 0 |
| 1995 | 1 | 0 |
| 1996 | 1 | 0 |
| 1999 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 2 | 0 |
| 2009 | 1 | 0 |
| 2011 | 3 | 1 |
| 2012 | 3 | 0 |
| 2013 | 1 | 0 |
| 2014 | 1 | 1 |
| 2015 | 1 | 0 |
| 2016 | 2 | 0 |
| 2018 | 1 | 0 |
| 2020 | 1 | 1 |
| 2022 | 2 | 1 |
| 2023 | 1 | 1 |
| 2024 | 1 | 1 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FRN device
- #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K974332 | Variable Pressure Syringe PumpSummary page 2…510(k) No. Date Cleared Model P7000 Variable Syringe ALARIS Medical K974332 12/29/98 Pump Systems, Inc. Medex 3000 Series Syringe Medex, Inc.… | 1998 |
| * | K982640 | Medex 3000 Series Syringe Infusion PumpSummary page 2…ALARIS Medical K974332 12/29/98 Pump Systems, Inc. Medex 3000 Series Syringe Medex, Inc. K982640 4/6/99 Infusion Pump, Model 30 1Oa INTENDED USE The MEDLEYTM Syringe Pump is intended for use in today’s gr… | 1999 |
What descends from it? (21)
See all 214 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (7 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 11, 2026 | Class II | Z-1645-2026 | Under Investigation by firm | Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units |
| Nov 17, 2017 | Class II | Z-0322-2018 | Nonconforming Material/Component | The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed. |
| Aug 8, 2016 | Class II | Z-2879-2016 | Software design | A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another. |
| Jul 21, 2015 | Class I | Z-2362-2015 | Process control | Channel Error code is displayed on the PC unit with an audio and visual alarm, and on the syringe module. After the error is cleared on the PCU, the syringe pump is unresponsive to key presses until the next power cycle, or the module is… |
| Jun 12, 2009 | Not classified | Z-1794-2009 | Device Design | There is a potential risk of IUI connector failure in the Alaris System when these connectors become corroded or contaminated, which may result in an interruption of infusion with an audible and visual alarm. IUI connectors can become… |
| Mar 5, 2008 | Not classified | Z-1712-2008 | Device Design | A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failure of the U9 Socketed Integrated Circuits (U9 IC) on the display board of the… |
| Mar 5, 2008 | Not classified | Z-1711-2008 | Device Design | A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors with the Alaris System due to the Inter-Unit Interface (IUI) connectors on all Alaris System… |
The 5 latest of 7 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 128 Class I recalls
- K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety Software and Supported…Smiths Medical MD, Inc. · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 16 Class I recalls
- K191910SpaceStation MRIB. Braun Melsugen AG · 2020 · 2 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls
- K211125Novum IQ Syringe PumpBaxter Healthcare Corporation · 2022 · 2 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall
- K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 17 Class I recalls
- K242390Novum IQ Syringe Pump (40800BAXUS)Baxter Healthcare Corporation · 2024 · 3 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
Nearest 5 of 6; all are on the web page.
What does this lineage say?
- The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 4.7 years on median.
- 3 of the 7 ancestors have a recall of any class.
- Of its descendants, K053554 (Milestone Scientific, Inc., 2006) has gone furthest, with 6 descendants of its own.
How does its category compare?
All FRN devicesFRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FRN
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FRN
172in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FRN
7in 2025
Class I (37 since 2007) Class II/III not classified
Adverse event reports per year · FRN
52,792in 2025
MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 102 | 15 | 1,086 |
| 2008 | 8 | 89 | 24 | 8,261 |
| 2009 | 9 | 66 | 102 | 1,952 |
| 2010 | 3 | 93 | 37 | 6,133 |
| 2011 | 3 | 103 | 69 | 41,367 |
| 2012 | 4 | 217 | 42 (2) | 15,482 |
| 2013 | 6 | 312 | 22 (3) | 40,082 |
| 2014 | 6 | 172 | 4 (3) | 94,180 |
| 2015 | 3 | 232 | 5 | 26,742 |
| 2016 | 3 | 301 | 7 (1) | 16,560 |
| 2017 | 3 | 305 | 13 | 12,060 |
| 2018 | 3 | 267 | 1 | 16,591 |
| 2019 | 2 | — | 18 (1) | 15,850 |
| 2020 | 1 | — | 16 (5) | 94,186 |
| 2021 | 2 | — | 3 (2) | 199,342 |
| 2022 | 6 | 268 | 2 | 888,020 |
| 2023 | 7 | 265 | 16 (10) | 165,838 |
| 2024 | 5 | 103 | 9 (4) | 57,730 |
| 2025 | 8 | 172 | 7 (4) | 52,792 |
| 2026 | 2 | — | 11 (2) | — |
What is it indicated for?
The MEDLEY Syringe Pump is intended for use in today's growing professional healthcare environment for facilities that utilize syringe pumps for the delivery of fluids, medications, blood and blood products.
The MEDLEY Syringe Pump is indicated for use on adults, pediatrics and neonates for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural, enteral, or irrigation of fluid spaces.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K023264 · printed Nov 6, 2026
Cite this page
510(k) Report. “K023264: Medley Syringe Pump Module System, Model 8110.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K023264