510(k) · Cleared
Recalled · Class II · 2024Brivo MR355
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ge Hangwei Medical Systems Co., Ltd. · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Mar 27, 2012 as substantially equivalent to K103330 (Ge Hangwei Medical Systems Co., Ltd., 2011) after a 13-day review. Its family tree reaches back 9 generations to 1997. 5 of the 26 devices in its family, counting itself, have a Class I recall. It has been recalled 4 times.
- Decided
- Mar 27, 2012
- Review
- 13 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 4
- Predicates
- 1
Its longest line of descent runs back 9 generations to K971684 (1997). 25 ancestors in all, the oldest K962061 (1996). No later device cites it as a predicate yet. 5 of the 26 devices in this family, counting this one, have a Class I recall.
- Root · 9 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
- 5 generations between
- 3 generations backK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
- 2 generations backK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
- 1 generation backK103330 Brivo MR355, Optima MR360Ge Hangwei Medical Systems Co., Ltd. · 2011 · Recalled
- This deviceK120778 Brivo MR355Ge Hangwei Medical Systems Co., Ltd. · 2012
Show all 9 generations
- Root · 9 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
- 8 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
- 7 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
- 6 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
- 5 generations backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
- 4 generations backK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
- 3 generations backK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
- 2 generations backK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
- 1 generation backK103330 Brivo MR355, Optima MR360Ge Hangwei Medical Systems Co., Ltd. · 2011 · Recalled
- This deviceK120778 Brivo MR355Ge Hangwei Medical Systems Co., Ltd. · 2012
Class I recalls in this family
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −9 | K971684 | Magnetom Project 047 System (Numaris 3.5 Software) | Siemens Medical Solutions USA, Inc. | 1997 | 4 |
| −8 | K032428 | Magnetom Avanto MR System | Siemens Medical Solutions USA, Inc. | 2003 | 4 |
| −7 | K041112 | Magnetom Espree System | Siemens Medical Solutions USA, Inc. | 2004 | 4 |
| −6 | K052164 | Magnetom Systems with Software Syngo MR 2006A | Siemens Medical Solutions USA, Inc. | 2005 | — |
| −5 | K062454 | Magnetom Systems - I-Class and T-Class Releases | Siemens Medi Cal Solutions, Inc. | 2006 | — |
| −4 | K072237 | Magnetom Verio | Siemens Medical Solutions USA, Inc. | 2007 | Class I |
| −3 | K081028 | Signa 3.0T MR750 System | Ge Medical Systems, LLC | 2008 | 11 |
| −2 | K083147 | Discovery MR450 | Ge Medical Systems, LLC | 2008 | 10 |
| −1 | K103330 | Brivo MR355, Optima MR360 | Ge Hangwei Medical Systems Co., Ltd. | 2011 | 6 |
| 0 (this device) | K120778 | Brivo MR355 | Ge Hangwei Medical Systems Co., Ltd. | 2012 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 1 | 0 |
| 1998 | 1 | 0 |
| 1999 | 2 | 0 |
| 2001 | 4 | 1 |
| 2002 | 2 | 1 |
| 2003 | 1 | 1 |
| 2004 | 1 | 1 |
| 2005 | 4 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K103330 | Brivo MR355, Optima MR360Summary page 1…c Resonance Imaging System Classification 21 C.F.R. 892. 1000 Magnetic Resonance Diagnostic Device Names:Product Code: LNH Predicate Device(s): K103330, GE 1.5T Brivo MR355/Optima MR360 Device Description: The modified Brivo MR355 adds (1) one dedicated coil: 4-ch… | 2011 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (5 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 14, 2024 | Class II | Z-2478-2024 | Device Design | GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2… |
| Jun 14, 2024 | Class II | Z-2477-2024 | Device Design | GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2… |
| Dec 23, 2021 | Class II | Z-0611-2022 | Process control | The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements. |
| May 11, 2017 | Class II | Z-3021-2017 | Mixed-up of materials/components | Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · 4 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
- K030874GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co. · 2003 · 5 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
- K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · 5 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
- K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · 6 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 1.1 years on median.
- 18 of the 25 ancestors have a recall of any class; 5 have a Class I recall.
- 1 other device also cites its predicate K103330.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The Brivo MR355 is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or flmctions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart…
Read the full statement (130 words)
The Brivo MR355 is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or flmctions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Brivo MR3 55 reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K120778 · printed Nov 6, 2026
Cite this page
510(k) Report. “K120778: Brivo MR355.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K120778