510(k) · Cleared
Recalled · Class II · 2018Procare Monitor B20
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ge Medical Systems China Co., Ltd. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)
Cleared on Mar 22, 2013 as substantially equivalent to K120598 (Ge Medical Systems China Co., Ltd., 2012) after a 238-day review. Its family tree reaches back 8 generations to 1997. 3 of the 70 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times.
- Decided
- Mar 22, 2013
- Review
- 238 days
- Class
- Class II
- Product code
- MHX
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 8 generations to K962603 (1997). 69 ancestors in all, the oldest K896030 (1989). No later device cites it as a predicate yet. 3 of the 70 devices in this family, counting this one, have a Class I recall.
- Root · 8 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
- 7 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
- 6 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
- 5 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
- 4 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
- 3 generations backK022193 Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400NGe Medical Systems Information Technologies · 2002
- 2 generations backK073203 Carescape V100Ge Medical Systems Information Technologies · 2007
- 1 generation backK120598 Procare Monitor B40Ge Medical Systems China Co., Ltd. · 2012 · Recalled
- This deviceK122253 Procare Monitor B20Ge Medical Systems China Co., Ltd. · 2013
Class I recalls in this family
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −8 | K962603 | Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables | Masimo Corp. | 1997 | — |
| −7 | K973887 | Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables | Masimo Corp. | 1998 | — |
| −6 | K982255 | Model 2000 Pulse Oximeter | Ivy Biomedical Systems, Inc. | 1998 | — |
| −5 | K991823 | N-395 Pulse Oximeter, Model N-395 | Nellcor Puritan Bennett, Inc. | 1999 | — |
| −4 | K012891 | Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and… | Nellcor Puritan Bennett, Inc. | 2002 | 1 |
| −3 | K022193 | Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400N | Ge Medical Systems Information Technologies | 2002 | — |
| −2 | K073203 | Carescape V100 | Ge Medical Systems Information Technologies | 2007 | — |
| −1 | K120598 | Procare Monitor B40 | Ge Medical Systems China Co., Ltd. | 2012 | 2 |
| 0 (this device) | K122253 | Procare Monitor B20 | Ge Medical Systems China Co., Ltd. | 2013 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1989 | 1 | 0 |
| 1990 | 2 | 0 |
| 1991 | 4 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1994 | 3 | 0 |
| 1995 | 3 | 0 |
| 1996 | 4 | 0 |
| 1997 | 2 | 0 |
| 1998 | 1 | 0 |
| 1999 | 3 | 0 |
| 2000 | 8 | 0 |
| 2001 | 3 | 0 |
| 2002 | 8 | 1 |
| 2003 | 4 | 0 |
| 2004 | 5 | 2 |
| 2005 | 1 | 0 |
| 2006 | 3 | 0 |
| 2007 | 1 | 0 |
| 2009 | 2 | 0 |
| 2010 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MHX device
- #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K120598 | Procare Monitor B40Summary page 1…870.1025 Arrhythmia detector and alarm (including ST- segment measurement and alarm) Product Code: MHX Predicate Device s) K120598 PROCARE Monitor B40 Device Description: The PROCARE Monitor B20 is a multi-parameter patient mo… | 2012 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (3 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 11, 2018 | Class II | Z-0583-2019 | Software design | Patient monitors may restart due to network overload caused by network configuration. |
| Oct 11, 2013 | Class II | Z-0489-2014 | Software design | GE Healthcare has recently become aware of a potential safety issue due to ECG filter settings associated with the B20/B40 V1, B30 and B40 V2 Patient Monitors. The B20, B30 and B40 Patient Monitors set the ECG filter setting incorrectly… |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K043276Datex-Ohmeda S/5TM FM with L-FICU04 and L-FICU04A Software and N-Fcrec ModuleDatex-Ohmeda · 2004 · 2 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
- K040831Datex-Ohmeda Trusat Pulse Oximeter and Accessories, Model 3500Datex-Ohmeda, Inc. · 2004 · 3 generations back · ◆ in the 2004 ancestors bar of FIG. 13 Class I recalls
- K021955Trutrak Plus Enhancements to the Datex-Ohmeda 3800 Series & 3900 Series Oximeters & AccessoriesDatex-Ohmeda, Inc. · 2002 · 4 generations back · ◆ in the 2002 ancestors bar of FIG. 12 Class I recalls
What does this lineage say?
- The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 11 months on median.
- 20 of the 69 ancestors have a recall of any class; 3 have a Class I recall.
- 1 other device also cites its predicate K120598.
How does its category compare?
All MHX devicesMHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.
FDA decisions per year · MHX
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MHX
192in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MHX
21so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MHX
3,761in 2025
MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 15 | 96 | 3 | 131 |
| 2008 | 16 | 58 | 2 | 232 |
| 2009 | 13 | 57 | 7 | 382 |
| 2010 | 20 | 90 | 13 | 955 |
| 2011 | 12 | 81 | 16 | 949 |
| 2012 | 16 | 88 | 11 | 676 |
| 2013 | 21 | 120 | 14 | 1,997 |
| 2014 | 13 | 248 | 6 | 1,609 |
| 2015 | 14 | 203 | 5 | 1,284 |
| 2016 | 13 | 149 | 7 | 1,105 |
| 2017 | 14 | 134 | 6 | 1,115 |
| 2018 | 11 | 203 | 12 | 2,157 |
| 2019 | 7 | 243 | 6 | 1,462 |
| 2020 | 11 | 196 | 0 | 1,714 |
| 2021 | 5 | 195 | 0 | 2,423 |
| 2022 | 15 | 189 | 6 | 1,470 |
| 2023 | 8 | 229 | 3 | 2,254 |
| 2024 | 5 | 181 | 3 | 2,814 |
| 2025 | 8 | 192 | 15 | 3,761 |
| 2026 | 14 | 205 | 21 | — |
What is it indicated for?
The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The PROCARE Monitor B20 is intended for use uinder the direct supervision of a licensed health care practitioner. The PROCARE Monitor B20 is not intended for use during MIII. The PROCARE…
Read the full statement (139 words)
The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
The PROCARE Monitor B20 is intended for use uinder the direct supervision of a licensed health care practitioner.
The PROCARE Monitor B20 is not intended for use during MIII.
The PROCARE Monitor B20 monitors and displays oscillometric nion-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, EGG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring.
Esophagea]INasopharyngeaV/Tymnpanic/Rectal/Bladder/Mxillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, and functional oxygen saturation (Sp0 2 ) and pulse rate via continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K122253 · printed Nov 6, 2026
Cite this page
510(k) Report. “K122253: Procare Monitor B20.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K122253