V-Pro 60 Low Temperature Sterilization System

K140498510(k) TraditionalCleared Jul 24, 2014147-day reviewClass II21 CFR 880.6860

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · MLR Sterilizer, Chemical

Cleared on Jul 24, 2014 as substantially equivalent to K131120 (STERIS Corporation, 2013) after a 147-day review. Its family tree reaches back 7 generations to 2003.

Decided
Jul 24, 2014
Review
147 days
Class
Class II
Product code
MLR
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K020875 (2003). 12 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 7 generations backK020875 TSO3 Ozone Sterilizer, Model 125LTechnologies of Sterilization With Oxone, Tso3, Inc. · 2003
  2. 6 generations backK051595 TSO3 Ozone Sterilizer, Model 125LTso3, Inc. · 2006
  3. 5 generations backK090636 TSO3 Ozone Sterilizer, Model 125LTso3, Inc. · 2009
  4. 4 generations backK102330 Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation · 2011
  5. 3 generations backK112760 Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation · 2011
  6. 2 generations backK120632 Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation · 2012
  7. 1 generation backK131120 Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature…STERIS Corporation · 2013
  8. This deviceK140498 V-Pro 60 Low Temperature Sterilization SystemSTERIS Corporation · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K020875TSO3 Ozone Sterilizer, Model 125LTechnologies of Sterilization With Oxone, Tso3, Inc.2003—
−6K051595TSO3 Ozone Sterilizer, Model 125LTso3, Inc.2006—
−5K090636TSO3 Ozone Sterilizer, Model 125LTso3, Inc.2009—
−4K102330Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation2011—
−3K112760Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation2011—
−2K120632Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation2012—
−1K131120Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation2013—
0 (this device)K140498V-Pro 60 Low Temperature Sterilization SystemSTERIS Corporation2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200510
200710
200910
201120
FIG. 1 — Lineage of K140498, 2003–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MLR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MLR device

  1. #1 K250802 STERRAD® 100NX Sterilizer with ALLClear™…score 14.0 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K131120Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature Sterilization System
Summary page 2
…Product Code MLR Device Class: Class 11 2. Predicate Devices Amsco V-PRO maX Low Temperature Stetilization System (K131 120). Please note that the predicate device was originally cleared under K102330. A comparison between the proposed…
STERIS Corporation2013—

Also cleared against K131120: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
  • 3 of the 12 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K131120.

How does its category compare?

All MLR devices

MLR Sterilizer, Chemical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · MLR

1in 2025

2007: 22009: 12011: 62012: 22013: 12014: 22015: 12016: 42017: 22018: 22019: 42021: 22022: 22023: 52024: 32025: 12026 (partial year): 1
2007high 6 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MLR

216in 2024

2011: 214 days median2016: 183 days median2019: 90 days median2023: 222 days median2024: 216 days median
2007high 222 (2023) · low 90 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MLR

1in 2017

2007: 52008: 82009: 22010: 12011: 22012: 32013: 92014: 32017: 1
2007high 9 (2013) · low 0 (2015)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MLR

46in 2025

2007: 2722008: 5362009: 4332010: 2112011: 342012: 1012013: 5412014: 3992015: 4122016: 2862017: 2792018: 2892019: 2222020: 1742021: 2022022: 1882023: 1402024: 442025: 46
2007high 541 (2013) · low 34 (2011)2025

MAUDE reports naming product code MLR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—5272
20080—8536
20091—2433
20100—1211
20116214234
20122—3101
20131—9541
20142—3399
20151—0412
201641830286
20172—1279
20182—0289
20194900222
20200—0174
20212—0202
20222—0188
202352220140
20243216044
20251—046
20261—0—

What is it indicated for?

The V-PRO 60 Low Temperature Sterilization System, with VAPROX® HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in Healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical…

Read the full statement (398 words)

The V-PRO 60 Low Temperature Sterilization System, with VAPROX® HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in Healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

The V-PRO 60 Sterilizer's Lumen Cycle can sterilize:

Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:

a. Single or dual lumen devices with stainless steel lumens that are 0.77 mm (-1/32") internal diameter (ID) and 410 mm (16-9/64") in length

b. Triple lumen devices with stainless steel lumens that are 1.2 mm (-3/64") ID and 275 mm (-10-55/64") in length; 1.8 mm (-5/64") ID and 310 mm (-12-13/64") in length; or 2.8 mm (-7/64") ID and 317 mm (12-31/64") in length

The validation studies for all lumen configurations were conducted using a maximum of twelve (12) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of one instrument tray with instrument organizers and mat and two pouches for a total weight of 11 lbs (5.0 kg).

The V-PRO 60 Sterilizer's Non Lumen Cycle can sterilize:

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors.

The validation studies were performed using a validation load consisting of one instrument tray with instrument organizers and mat and one pouch for a total weight of 12 lbs (5.4 kg).

The V-PRO 60 Sterilizer's Flexible Cycle can sterilize:

One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a single or dual lumen device with lumens that are > 1 mm (-3/64") ID and < 990 mm (38-63/64") in length.

The validation studies were conducted with one flexible endoscope, packaged into a tray with silicone mat, instrument organizers and light cord (if not integral to scope) and no additional load.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140498: V-Pro 60 Low Temperature Sterilization System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K140498

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026