510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

O-Two E700,O-Two E600,O-Two E500

K141595510(k) TraditionalCleared Mar 16, 2015276-day reviewClass II21 CFR 868.5925Recalls 3 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

O-Two Medical Technologies, Inc. · BTL Ventilator, Emergency, Powered (Resuscitator)

Cleared on Mar 16, 2015 as substantially equivalent to K062267 (Drager Medical B.V., 2006) and 1 other predicate after a 276-day review. Its family tree reaches back 6 generations to 1997. 2 of the 25 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 3 times, including 3 Class I recalls.

Decided
Mar 16, 2015
Review
276 days
Class
Class II
Product code
BTL
Recalls
3 (3 Class I)
Predicates
2

Its longest line of descent runs back 6 generations to K962603 (1997). 23 ancestors in all, the oldest K860555 (1986). 1 later device descends from it, 1 citing it directly. 2 of the 25 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2004: 1 other ancestor2001: 2 other ancestors1999: 3 other ancestors1995: 2 other ancestors1986: 1 other ancestor2000: 1 other ancestor1998: 4 other ancestors1993: 1 other ancestor1996: 1 other ancestor2023: 1 descendant clearedK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K990966 · Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric Disposable Sensor, Masimo Lnop -PDT-Pediatric/Small Dis · Masimo Corp. · 1999K021981 · Ivent 201 Portable Ventilator · Versamed, Ltd. · 2003K042468 · Ivent 201 Portable Ventilator · Versamed Medical Systems, Inc. · 2004K062267 · Oxylog 3000 · Drager Medical B.V. · 2006 · Class I recall1990200020102020
  1. Root · 6 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 5 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 4 generations backK990966 Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric…Masimo Corp. · 1999
  4. 3 generations backK021981 Ivent 201 Portable VentilatorVersamed, Ltd. · 2003
  5. 2 generations backK042468 Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2004
  6. 1 generation backK062267 Oxylog 3000Drager Medical B.V. · 2006 · Recalled (Class I)
  7. This deviceK141595 O-Two E700,O-Two E600,O-Two E500O-Two Medical Technologies, Inc. · 2015
  8. Descendant · cites it directly · 0 of its ownK221634 Oxymag - Transport and Emergency VentilatorMagnamed Tecnologia Medica S/A · 2023

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K221634Oxymag - Transport and Emergency VentilatorMagnamed Tecnologia Medica S/A · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−5K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−4K990966Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric Disposable…Masimo Corp.1999—
−3K021981Ivent 201 Portable VentilatorVersamed, Ltd.2003—
−2K042468Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2004—
−1K062267Oxylog 3000Drager Medical B.V.2006Class I
0 (this device)K141595O-Two E700,O-Two E600,O-Two E500O-Two Medical Technologies, Inc.2015—
+1K221634Oxymag - Transport and Emergency VentilatorMagnamed Tecnologia Medica S/A2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199310
199520
199610
199840
199930
200010
200120
200410
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
FIG. 1 — Lineage of K141595, 1997–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BTL device

  1. #1 K242769 VentStar Resus heated (N) (MP17030); VentStar…score 11.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062267Oxylog 3000
Summary page 4
…esiology Predicate Devices: Oxylog 3000  Manufactured Draeger medical GmbH  510(k) number K062267 Carevent PAR  Manufactured by O-Two Medical Technologies Inc.…
Drager Medical B.V.2006Class I
*K081330Carevent Par (Public Access Resuscitator)
Summary page 4
…Carevent PAR  Manufactured by O-Two Medical Technologies Inc.  510(k) number K081330 Device Description: The proposed o_two e700, o_two e600 and o_two e500 are time-cycled, volume-constant and pressure controlle…
O-Two Medical Technologies, Inc.2009—

Also cleared against K062267: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K221634Oxymag - Transport and Emergency VentilatorMagnamed Tecnologia Medica S/A20230

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 12, 2019Class IZ-0976-2019Process controlThere is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main…
Feb 12, 2019Class IZ-0978-2019Process controlThere is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main…
Feb 12, 2019Class IZ-0977-2019Process controlThere is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K062267Oxylog 3000Drager Medical B.V. · 2006 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 18 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 2 of the 23 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K221634 (Magnamed Tecnologia Medica S/A, 2023) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K062267.

How does its category compare?

All BTL devices

BTL Ventilator, Emergency, Powered (Resuscitator): FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · BTL

1in 2025

2007: 52008: 42009: 52010: 32011: 22013: 12014: 72015: 12017: 22018: 12019: 12020: 12021: 12023: 12025: 1
2007high 7 (2014) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BTL

203in 2014

2007: 122 days median2008: 198 days median2009: 173 days median2010: 140 days median2014: 203 days median
2007high 203 (2014) · low 122 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BTL

2in 2024

2007: 12011: 12012: 2, 2 Class I2013: 3, 1 Class I2015: 3, 2 Class I2019: 3, 3 Class I2021: 12024: 2, 2 Class I
2007high 3 (2013) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · BTL

165in 2025

2007: 52008: 372009: 692010: 962011: 752012: 1152013: 1032014: 702015: 792016: 802017: 1012018: 1972019: 3412020: 4542021: 5062022: 7552023: 7312024: 6672025: 165
2007high 755 (2022) · low 5 (2007)2025

MAUDE reports naming product code BTL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007512215
20084198037
20095173069
20103140096
20112—175
20120—2 (2)115
20131—3 (1)103
20147203070
20151—3 (2)79
20160—080
20172—0101
20181—0197
20191—3 (3)341
20201—0454
20211—1506
20220—0755
20231—0731
20240—2 (2)667
20251—0165
20260—0—

What is it indicated for?

o_two e700, e600 and e500 are a time-cycled, volume-constant and pressure-controlled (only e700) emergency and

transport ventilator designed for use in the pre-hospital, intra-hospital, inter-hospital and transport settings. It is intended

for use with adult, child, infant patients with a tidal volume from 50 ml (100 ml for e500) upwards who are in respiratory

and/or cardiac arrest or respiratory distress and who require ventilatory support.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141595: O-Two E700,O-Two E600,O-Two E500.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141595

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.