510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Hamilton-T1,Hamilton-C1

K140939510(k) TraditionalCleared Sep 2, 2015509-day reviewClass II21 CFR 868.5895Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hamilton Medical AG · CBK Ventilator, Continuous, Facility Use

Cleared on Sep 2, 2015 as substantially equivalent to K120670 (Hamilton Medical AG, 2012) and 5 other predicates after a 509-day review. Its family tree reaches back 14 generations to 1991. 16 of the 124 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled once.

Decided
Sep 2, 2015
Review
509 days
Class
Class II
Product code
CBK
Recalls
1
Predicates
6

Its longest line of descent runs back 14 generations to K902859 (1991). 121 ancestors in all, the oldest K791045 (1979). 2 later devices descend from it, 2 citing it directly. 16 of the 124 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor, 1 with a Class I recall12012: 1 other ancestor2008: 5 other ancestors2006: 6 other ancestors, 1 with a Class I recall12011: 1 other ancestor, 1 with a Class I recall12010: 3 other ancestors, 1 with a Class I recall12007: 1 other ancestor2004: 10 other ancestors, 1 with a Class I recall12002: 5 other ancestors, 1 with a Class I recall12001: 6 other ancestors1999: 9 other ancestors1995: 6 other ancestors2005: 3 other ancestors2003: 4 other ancestors, 1 with a Class I recall12000: 4 other ancestors, 1 with a Class I recall11998: 10 other ancestors, 1 with a Class I recall11994: 5 other ancestors1991: 6 other ancestors1990: 6 other ancestors1997: 4 other ancestors1996: 3 other ancestors1979: 1 other ancestor1993: 1 other ancestor1992: 2 other ancestors1989: 1 other ancestor1988: 1 other ancestor1984: 2 other ancestors2019: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared, 1 with a Class I recall1K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK961687 · EVITA 4 · Drager, Inc. · 1996K970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK984535 · Puritan-Bennett 840 Ventilatory System with Bilevel Option · Puritan Bennett Corp. · 1998 · Class I recallK022679 · Raphael · Hamilton Medical AG · 2003 · recalledK040574 · Galileo Gold · Hamilton Medical, Inc. · 2004K051263 · Evitaxl with Option Smartcare · Drager Medical AG & Co. Kgaa · 2005 · recalledK061090 · Galileo Gold Asv · Hamilton Medical AG · 2006K070513 · Hamilton-G5 · Hamilton Medical AG · 2007 · recalledK081521 · Modification to Hamilton-G5 · Hamilton Medical AG · 2008 · Class I recallK102775 · Hamilton-C2 · Hamilton Medical AG · 2011×2: 2 generations cleared close together, K112006 (2012), K120670 (2012) K121225 · Hamilton-C2 · Hamilton Medical AG · 2012 · Class I recall19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors, 2 with a Class I recall22 generations back: 10 other ancestors, 2 with a Class I recall23 generations back: 24 other ancestors, 2 with a Class I recall24 generations back: 22 other ancestors5 generations back: 24 other ancestors, 1 with a Class I recall16 generations back: 15 other ancestors, 1 with a Class I recall17 generations back: 4 other ancestors, 1 with a Class I recall18 generations back: 4 other ancestors 1 generation forward: 2 descendants, 2 with a Class I recall2 K121225 · Hamilton-C2 · Hamilton Medical AG · 2012 · Class I recallK120670 · Hamilton-T1 · Hamilton Medical AG · 2012 · Class I recallK112006 · Hamilton-TC1 · Hamilton Medical AG · 2012K102775 · Hamilton-C2 · Hamilton Medical AG · 2011K081521 · Modification to Hamilton-G5 · Hamilton Medical AG · 2008 · Class I recallK070513 · Hamilton-G5 · Hamilton Medical AG · 2007 · recalledK061090 · Galileo Gold Asv · Hamilton Medical AG · 2006K051263 · Evitaxl with Option Smartcare · Drager Medical AG & Co. Kgaa · 2005 · recalledK040574 · Galileo Gold · Hamilton Medical, Inc. · 2004K022679 · Raphael · Hamilton Medical AG · 2003 · recalledK984535 · Puritan-Bennett 840 Ventilatory System with Bilevel Option · Puritan Bennett Corp. · 1998 · Class I recallK970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK961687 · EVITA 4 · Drager, Inc. · 1996K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalled−10−50
  1. Root · 14 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 10 generations between
  3. 3 generations backK112006 Hamilton-TC1Hamilton Medical AG · 2012
  4. 2 generations backK120670 Hamilton-T1Hamilton Medical AG · 2012 · Recalled (Class I)
  5. 1 generation backK121225 Hamilton-C2Hamilton Medical AG · 2012 · Recalled (Class I)
  6. This deviceK140939 Hamilton-T1,Hamilton-C1Hamilton Medical AG · 2015
  7. Descendant · cites it directly · 0 of its ownK181216 Hamilton-T1, Hamilton-C1Hamilton Medical AG · 2019 · Recalled (Class I)
  8. Descendant · cites it directly · 0 of its ownK153046 Hamilton-MR1Hamilton Medical AG · 2016 · Recalled (Class I)
Show all 14 generations
  1. Root · 14 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 13 generations backK961687 EVITA 4Drager, Inc. · 1996
  3. 12 generations backK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · Recalled (Class I)
  4. 11 generations backK984535 Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · Recalled (Class I)
  5. 10 generations backK022679 RaphaelHamilton Medical AG · 2003 · Recalled
  6. 9 generations backK040574 Galileo GoldHamilton Medical, Inc. · 2004
  7. 8 generations backK051263 Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa · 2005 · Recalled
  8. 7 generations backK061090 Galileo Gold AsvHamilton Medical AG · 2006
  9. 6 generations backK070513 Hamilton-G5Hamilton Medical AG · 2007 · Recalled
  10. 5 generations backK081521 Modification to Hamilton-G5Hamilton Medical AG · 2008 · Recalled (Class I)
  11. 4 generations backK102775 Hamilton-C2Hamilton Medical AG · 2011
  12. 3 generations backK112006 Hamilton-TC1Hamilton Medical AG · 2012
  13. 2 generations backK120670 Hamilton-T1Hamilton Medical AG · 2012 · Recalled (Class I)
  14. 1 generation backK121225 Hamilton-C2Hamilton Medical AG · 2012 · Recalled (Class I)
  15. This deviceK140939 Hamilton-T1,Hamilton-C1Hamilton Medical AG · 2015

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K181216Hamilton-T1, Hamilton-C1Hamilton Medical AG · 2019 · cites it directly0descendants
  • K153046Hamilton-MR1Hamilton Medical AG · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−14K902859Servo Ventilator 300Siemens Life Support Systems19913
−13K961687EVITA 4Drager, Inc.1996—
−12K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp.1998Class I
−11K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp.1998Class I
−10K022679RaphaelHamilton Medical AG20033
−9K040574Galileo GoldHamilton Medical, Inc.2004—
−8K051263Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa20051
−7K061090Galileo Gold AsvHamilton Medical AG2006—
−6K070513Hamilton-G5Hamilton Medical AG20071
−5K081521Modification to Hamilton-G5Hamilton Medical AG2008Class I
−4K102775Hamilton-C2Hamilton Medical AG2011—
−3K112006Hamilton-TC1Hamilton Medical AG2012—
−2K120670Hamilton-T1Hamilton Medical AG2012Class I
−1K121225Hamilton-C2Hamilton Medical AG2012Class I
0 (this device)K140939Hamilton-T1,Hamilton-C1Hamilton Medical AG2015—
+1K181216Hamilton-T1, Hamilton-C1Hamilton Medical AG2019Class I
+1K153046Hamilton-MR1Hamilton Medical AG2016Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198420
198810
198910
199060
199160
199220
199310
199450
199560
199630
199740
1998101
199990
200041
200160
200251
200341
2004101
200530
200661
200710
200850
201031
201111
201210
201311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201611
201911
FIG. 1 — Lineage of K140939, 1991–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

16 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K120670Hamilton-T1
Summary page 5
…me: 21 CFR 868.5895 Continuous Ventilator Device Classification and Product Code: Class II CBK (incl. Class II DQA) III. PREDICATE DEVICES HAMILTON-T1 (K120670) 21 CFR 868.5895 Continuous Ventilator, Class II, CBK HAMILTO…
Hamilton Medical AG2012Class I
*K120574Hamilton-C1
Summary page 5
…VICES HAMILTON-T1 (K120670) 21 CFR 868.5895 Continuous Ventilator, Class II, CBK HAMILTON-C1 (K120574) 21 CFR 868.5895 Continuous Ventilator, Class II, CBK HAMILTON-C2 (K121225) for the 21 CFR 868.589…
Hamilton Medical AG2012—
*K121225Hamilton-C2
Summary page 5
…I, CBK HAMILTON-C1 (K120574) 21 CFR 868.5895 Continuous Ventilator, Class II, CBK HAMILTON-C2 (K121225) for the 21 CFR 868.5895 Continuous Ventilator, Class II, neonatal modes. CBK HAMILTO…
Hamilton Medical AG2012Class I
*K131774Hamilton-G5
Summary page 5
…I, CBK HAMILTON-C2 (K121225) for the 21 CFR 868.5895 Continuous Ventilator, Class II, neonatal modes. CBK HAMILTON-G5 (K131774) for the 21 CFR 868.5895 Continuous Ventilator, Class II, Masimo SpO2 board and sensors CBK, (incl. Class…
Hamilton Medical AG2013Class I
*K073179Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option, Model 6675585
Summary page 5
…(K131774) for the 21 CFR 868.5895 Continuous Ventilator, Class II, Masimo SpO2 board and sensors CBK, (incl. Class II DQA) MAQUET Servo-i (K073179) for the 21 CFR 868.5895 Continuous Ventilator, Class II, nCPAP and nCPAP-PC modes. CBK Dräger Oxy…
Maquet Critical Care AB2008—
*K062267Oxylog 3000
Summary page 5
…MAQUET Servo-i (K073179) for the 21 CFR 868.5895 Continuous Ventilator, Class II, nCPAP and nCPAP-PC modes. CBK Dräger Oxylog 3000 (K062267) for 21 CFR 868.5895 Continuous Ventilator, Class II, O2 consumption CBK No reference…
Drager Medical B.V.2006Class I

Also cleared against K120670: 2 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181216Hamilton-T1, Hamilton-C1 Class IHamilton Medical AG20190
K153046Hamilton-MR1 Class IHamilton Medical AG20160

Has anything in its family been recalled? (16 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 9, 2018Class IIZ-2795-2018Nonconforming Material/ComponentAn out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K131774Hamilton-G5Hamilton Medical AG · 2013 · Direct predicate · ◆ in the 2013 ancestors bar of FIG. 12 Class I recalls
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K120670Hamilton-T1Hamilton Medical AG · 2012 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K062267Oxylog 3000Drager Medical B.V. · 2006 · Direct predicate · ◆ in the 2006 ancestors bar of FIG. 13 Class I recalls
  • K103803Hamilton-G5Hamilton Medical AG · 2011 · 2 generations back · ◆ in the 2011 ancestors bar of FIG. 11 Class I recall
  • K093632Infinity Acute Care System Workstation Neonatal CareDrager Medical AG & Co. Kgaa · 2010 · 2 generations back · ◆ in the 2010 ancestors bar of FIG. 14 Class I recalls
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 3 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K013642AVEABird Products Corp. · 2002 · 3 generations back · ◆ in the 2002 ancestors bar of FIG. 113 Class I recalls
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 5 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall

Nearest 12 of 16 shown.

What does this lineage say?

  • The line of descent spans 24 years and passes through 6 companies; a new generation was cleared every 1.1 years on median.
  • 38 of the 121 ancestors have a recall of any class; 14 have a Class I recall.
  • Of its descendants, K181216 (Hamilton Medical AG, 2019) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K120670.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The HAMILTON-T1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates. Intended areas of use: - In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room - For emergency medical care - During transport within and outside the hospital - During transfer by rescue vehicles, fixed wing aircraft, helicopter or ship…

Read the full statement (182 words)

The HAMILTON-T1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates.

Intended areas of use:

- In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room

- For emergency medical care

- During transport within and outside the hospital

- During transfer by rescue vehicles, fixed wing aircraft, helicopter or ship

The HAMILTON-T1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

The HAMILTON-C1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates.

Intended areas of use:

- In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room

- During transfer of ventilated patients within the hospital

The HAMILTON-C1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140939: Hamilton-T1,Hamilton-C1.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K140939

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026