510(k) Report

510(k) · Cleared

Cleared

Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test Dipcard

K142800510(k) TraditionalCleared Dec 11, 201473-day reviewClass II21 CFR 862.3100

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Co-Innovation Biotech Co., Ltd. · DKZ Enzyme Immunoassay, Amphetamine

Cleared on Dec 11, 2014 as substantially equivalent to K122809 (Advin Biotech, 2013) after a 73-day review. Its family tree reaches back 6 generations to 1995. 13 later devices descend from it.

Decided
Dec 11, 2014
Review
73 days
Class
Class II
Product code
DKZ
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K945610 (1995). 27 ancestors in all. 13 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2005: 1 other ancestor2004: 3 other ancestors2003: 2 other ancestors2002: 2 other ancestors2001: 7 other ancestors2000: 1 other ancestor1999: 1 other ancestor1998: 1 other ancestor1996: 2 other ancestors2016: 2 descendants cleared2024: 2 descendants cleared2019: 4 descendants cleared2018: 2 descendants cleared2023: 2 descendants cleared2020: 1 descendant clearedK945610 · Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-Amp · Princeton BioMeditech Corp. · 1995K011673 · Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test Device · ACON Laboratories, Inc. · 2001K020313 · Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug Multi-Line Screen Test Device · Acon Laboratories Co. · 2002K031759 · Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card · ACON Laboratories, Inc. · 2003 · recalledK061718 · Innovacon Spectrum II Test Card · Innovacon, Inc. · 2006 · recalledK122809 · Advin Multi-Drug Screen Test Cassette, Dip Card and Cup · Advin Biotech · 20131995200020052010201520202025
  1. Root · 6 generations backK945610 Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM)…Princeton BioMeditech Corp. · 1995
  2. 5 generations backK011673 Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step…ACON Laboratories, Inc. · 2001
  3. 4 generations backK020313 Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step…Acon Laboratories Co. · 2002
  4. 3 generations backK031759 Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon…ACON Laboratories, Inc. · 2003 · Recalled
  5. 2 generations backK061718 Innovacon Spectrum II Test CardInnovacon, Inc. · 2006 · Recalled
  6. 1 generation backK122809 Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech · 2013
  7. This deviceK142800 Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test…Co-Innovation Biotech Co., Ltd. · 2014
  8. Descendant · cites it directly · 11 of its ownK153050 Rapid Single/Multi-drug Test Cup, Rapid Single/Multi-drug Test DipcardCo-Innovation Biotech Co., Ltd. · 2016
  9. Descendant · 2 generations on · 5 of its ownK182530 BIOEASY Multi-Drug Test CupShenzhen Bioeasy Biotechnology Co., Ltd. · 2018
  10. Descendant · 3 generations on · 1 of its ownK192515 BIOEASY Marijuana Test Dip Card 40, BIOEASY Marijuana Test Dip Card…Shenzhen Bioeasy Biotechnology Co., Ltd. · 2019
  11. Descendant · 3 generations on · 1 of its ownK192301 BIOEASY Marijuana Test Dip Card, BIOEASY Marijuana Test StripShenzhen Bioeasy Biotechnology Co., Ltd. · 2019
  12. Descendant · 2 generations on · 1 of its ownK180349 AssureTech Panel Dip Tests, AssureTech Quick Cup TestsAssure Tech. (Hangzhou) Co, Ltd. · 2018

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2016: 22018: 22019: 42020: 12023: 22024: 2
2016high 4 (2019) · low 0 (2017)2026

Most influential descendants

Showing 5 of 13 · See all
  • K153050Rapid Single/Multi-drug Test Cup, Rapid Single/Multi-drug Test DipcardCo-Innovation Biotech Co., Ltd. · 2016 · cites it directly11descendants
  • K182530BIOEASY Multi-Drug Test CupShenzhen Bioeasy Biotechnology Co., Ltd. · 2018 · 2 generations on5descendants
  • K192515BIOEASY Marijuana Test Dip Card 40, BIOEASY Marijuana Test Dip Card 20, BIOEASY…Shenzhen Bioeasy Biotechnology Co., Ltd. · 2019 · 3 generations on1descendant
  • K192301BIOEASY Marijuana Test Dip Card, BIOEASY Marijuana Test StripShenzhen Bioeasy Biotechnology Co., Ltd. · 2019 · 3 generations on1descendant
  • K180349AssureTech Panel Dip Tests, AssureTech Quick Cup TestsAssure Tech. (Hangzhou) Co, Ltd. · 2018 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K945610Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-AmpPrinceton BioMeditech Corp.1995—
−5K011673Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test…ACON Laboratories, Inc.2001—
−4K020313Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug…Acon Laboratories Co.2002—
−3K031759Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum…ACON Laboratories, Inc.20031
−2K061718Innovacon Spectrum II Test CardInnovacon, Inc.20066
−1K122809Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech2013—
0 (this device)K142800Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test DipcardCo-Innovation Biotech Co., Ltd.2014—
+1K153050Rapid Single/Multi-drug Test Cup, Rapid Single/Multi-drug Test DipcardCo-Innovation Biotech Co., Ltd.2016—
+2K182530BIOEASY Multi-Drug Test CupShenzhen Bioeasy Biotechnology Co., Ltd.2018—
+3K192515BIOEASY Marijuana Test Dip Card 40, BIOEASY Marijuana Test Dip Card 20…Shenzhen Bioeasy Biotechnology Co., Ltd.2019—
+3K192301BIOEASY Marijuana Test Dip Card, BIOEASY Marijuana Test StripShenzhen Bioeasy Biotechnology Co., Ltd.2019—
+2K180349AssureTech Panel Dip Tests, AssureTech Quick Cup TestsAssure Tech. (Hangzhou) Co, Ltd.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199620
199810
199910
200010
200170
200220
200320
200430
200510
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201620
201820
201940
202010
202320
202420
FIG. 1 — Lineage of K142800, 1995–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DKZ device

  1. #1 K233019 AllTest Multi-Drug Rapid Test Cup; AllTest…score 14.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122809Advin Multi-Drug Screen Test Cassette, Dip Card and Cup
Summary page 6
…JXM,DJR,DJG,LCM 5-1 Co-Innovation Biotech Co.,Ltd. 4. The Predicate Devices K122809 Advin Multi-Drug Screen Test Dip Card Advin Multi-Drug Screen Test Cup Advin Mul…
Advin Biotech2013—

Also cleared against K122809: 11 other devices →

What descends from it? (13)

See all 13

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K153050Rapid Single/Multi-drug Test Cup, Rapid Single/Multi-drug Test DipcardCo-Innovation Biotech Co., Ltd.201611
K160793First Sign Drug of Abuse Dip Card Test (MDMA, EDDP, Nortriptyline), First Sign Drug of…W.H.P.M., Inc.20160

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 6 companies; a new generation was cleared every 3.3 years on median.
  • 2 of the 27 ancestors have a recall of any class.
  • Of its descendants, K153050 (Co-Innovation Biotech Co., Ltd., 2016) has gone furthest, with 11 descendants of its own.
  • 11 other devices also cite its predicate K122809.

How does its category compare?

All DKZ devices

DKZ Enzyme Immunoassay, Amphetamine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · DKZ

1in 2024

2008: 22009: 22010: 42011: 62012: 52013: 82014: 82015: 52016: 52018: 32019: 32020: 12023: 22024: 1
2007high 8 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DKZ

180in 2019

2010: 233 days median2011: 123 days median2012: 55 days median2013: 103 days median2014: 128 days median2015: 107 days median2016: 215 days median2018: 128 days median2019: 180 days median
2007high 233 (2010) · low 55 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DKZ

1in 2017

2007: 12008: 32009: 52011: 42013: 12014: 32017: 1
2007high 5 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DKZ

2in 2025

2008: 122009: 12010: 12011: 22013: 12014: 92015: 82016: 82021: 32022: 42023: 72024: 12025: 2
2007high 12 (2008) · low 0 (2007)2025

MAUDE reports naming product code DKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20082—312
20092—51
2010423301
2011612342
201255500
2013810311
2014812839
2015510708
2016521508
20170—10
2018312800
2019318000
20201—00
20210—03
20220—04
20232—07
20241—01
20250—02
20260—0—

What is it indicated for?

Rapid Single/Multi-drug test Cup and Rapid Single/Multi-drug test Dipcard are lateral flow chromatographic immunoassays designed to qualitatively detect the presence of drugs and drug metabolites in human urine at the following cut-off concentrations: Test Calibrator Cut-off level Marijuana (THC) Delta-9-THC-COOH 50ng/mL Cocaine (COC) Benzolecgonine 300ng/mL Amphetamine (AMP) D-Amphetamine 1000ng/mL Methamphetamine (MET) D-Methamphetamine 1000ng/mL…

Read the full statement (217 words)

Rapid Single/Multi-drug test Cup and Rapid Single/Multi-drug test Dipcard are lateral flow chromatographic immunoassays designed to qualitatively detect the presence of drugs and drug metabolites in human urine at the following cut-off concentrations:

Test Calibrator Cut-off level

Marijuana (THC) Delta-9-THC-COOH 50ng/mL

Cocaine (COC) Benzolecgonine 300ng/mL

Amphetamine (AMP) D-Amphetamine 1000ng/mL

Methamphetamine (MET) D-Methamphetamine 1000ng/mL

Morphine 2000 (MOP) Morphine 2000ng/mL

Barbiturates (BAR) Secobarbital 300ng/mL

Benzodiazepines (BZO) Oxazepam 300ng/mL

Methylenedioxymethamphetamine (MDMA) 3,4-Methylenedioxymethamphetamine 500ng/mL

Methadone (MTD) Methadone 300ng/mL

Oxycodone (OXY) Oxycodone 100ng/mL

Phencyclidine (PCP) Phencyclidine 25ng/mL

The tests contain two formats: 1) Test Cup and 2) Test Dipcard. The tests may be configured as single drug tests or multiple drug tests in any combination of the drug analytes listed in the table above. These tests are intended for in vitro diagnostics use. They are intended for prescription use including point of care sites and over-the-counter use.

The tests will yield preliminary positive results when prescription drugs Barbiturates, Benzodiazepine, and Methadone are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturates and Benzodiazepine in urine.

The assays provide only a preliminary analytical test result. Gas Chromatography/Mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142800: Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test Dipcard.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142800

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026