510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Infx-8000C, V6.20

K152697510(k) TraditionalCleared Jan 15, 2016116-day reviewClass II21 CFR 892.1650Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshibamedical Systems Corporation · OWB Interventional Fluoroscopic X-Ray System

Cleared on Jan 15, 2016 as substantially equivalent to K113052 (Toshiba America Medical Systems, In.C, 2011) after a 116-day review. Its family tree reaches back 8 generations to 1989. 3 later devices descend from it. It has been recalled 5 times.

Decided
Jan 15, 2016
Review
116 days
Class
Class II
Product code
OWB
Recalls
5
Predicates
1

Its longest line of descent runs back 8 generations to K890348 (1989). 20 ancestors in all. 3 later devices descend from it, 1 citing it directly.

  1. Root · 8 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 7 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. 6 generations backK022322 Innova and Innova 2000 SGe Medical Systems, Inc. · 2002 · Recalled
  4. 5 generations backK031637 Innova 3100GE Medical Systems · 2003 · Recalled
  5. 4 generations backK050489 Innova 2100Ge Medical Systems, LLC · 2005 · Recalled
  6. 3 generations backK052412 Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova…GE Healthcare · 2005 · Recalled
  7. 2 generations backK081621 Infx-8000C (Infinix CC-I and VC-I)Toshiba America Medical Systems, In.C · 2008 · Recalled
  8. 1 generation backK113052 Infinix-CciToshiba America Medical Systems, In.C · 2011 · Recalled
  9. This deviceK152697 Infx-8000C, V6.20Toshibamedical Systems Corporation · 2016
  10. Descendant · cites it directly · 2 of its ownK181804 Alphenix, INFX-8000C/B, V8.0Canon Medical Systems Corporation · 2018 · Recalled
  11. Descendant · 2 generations on · 1 of its ownK210900 Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2Canon Medical Systems Corporation · 2021 · Recalled
  12. Descendant · 3 generations on · 0 of its ownK251645 Self-Propelled CT Scan Base Kit, CGBA-035A; Alphenix, INFX-8000C/B…Canon Medical Systems Corporation · 2025

Most influential descendants

Showing 3 of 3 · See all
  • K181804Alphenix, INFX-8000C/B, V8.0Canon Medical Systems Corporation · 2018 · cites it directly2descendants
  • K210900Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2Canon Medical Systems Corporation · 2021 · 2 generations on1descendant
  • K251645Self-Propelled CT Scan Base Kit, CGBA-035A; Alphenix, INFX-8000C/B, INFX-8000C/S, V9.6…Canon Medical Systems Corporation · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.19895
−7K993037LCV+ Version 2Ge Medical Systems, Inc.20006
−6K022322Innova and Innova 2000 SGe Medical Systems, Inc.20027
−5K031637Innova 3100GE Medical Systems200325
−4K050489Innova 2100Ge Medical Systems, LLC200527
−3K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare200550
−2K081621Infx-8000C (Infinix CC-I and VC-I)Toshiba America Medical Systems, In.C20081
−1K113052Infinix-CciToshiba America Medical Systems, In.C20111
0 (this device)K152697Infx-8000C, V6.20Toshibamedical Systems Corporation2016—
+1K181804Alphenix, INFX-8000C/B, V8.0Canon Medical Systems Corporation20183
+2K210900Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2Canon Medical Systems Corporation20217
+3K251645Self-Propelled CT Scan Base Kit, CGBA-035A; Alphenix, INFX-8000C/B…Canon Medical Systems Corporation2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199410
199530
199810
200120
200220
200310
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202110
202510
FIG. 1 — Lineage of K152697, 1989–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K113052Infinix-Cci
Summary page 5
…ing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard]. 11. PREDICATE DEVICE: INFX-8000C, w/Spot Fluoroscopy Option (K113052) TABLE 1: Predicate Device Product Marketed by 510(k) Number Clearance Date…
Toshiba America Medical Systems, In.C20111

Also cleared against K113052: 2 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181804Alphenix, INFX-8000C/B, V8.0Canon Medical Systems Corporation20182

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 5, 2021Class IIZ-1175-2021Software designThe x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.
Dec 8, 2020Class IIZ-1270-2021Component design/selectionSystem table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
Oct 8, 2020Class IIZ-0532-2021Component design/selectionThere is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage…
Mar 3, 2020Class IIZ-2326-2020Process controlBolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
Dec 29, 2017Class IIZ-0843-2018Process controlDuring an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 7 companies; a new generation was cleared every 2.8 years on median.
  • 16 of the 20 ancestors have a recall of any class.
  • Of its descendants, K181804 (Canon Medical Systems Corporation, 2018) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K113052.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration.

The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen

and lower extremities.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152697: Infx-8000C, V6.20.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152697

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026