510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

RayStation

K200569510(k) TraditionalCleared Sep 24, 2020204-day reviewClass II21 CFR 892.5050Recalls 14

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

RaySearch Laboratories AB (PUBL) · MUJ System, Planning, Radiation Therapy Treatment

Cleared on Sep 24, 2020 as substantially equivalent to K190387 (RaySearch Laboratories AB (PUBL), 2019) after a 204-day review. Its family tree reaches back 15 generations to 2001. 6 later devices descend from it. It has been recalled 14 times.

Decided
Sep 24, 2020
Review
204 days
Class
Class II
Product code
MUJ
Recalls
14
Predicates
1

Its longest line of descent runs back 15 generations to K010975 (2001). 70 ancestors in all, the oldest K802809 (1981). 6 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2012: 1 other ancestor2009: 2 other ancestors2002: 3 other ancestors2000: 5 other ancestors2011: 1 other ancestor2010: 1 other ancestor2008: 3 other ancestors2007: 1 other ancestor2003: 2 other ancestors2001: 5 other ancestors1999: 4 other ancestors1997: 2 other ancestors1996: 5 other ancestors2006: 2 other ancestors2005: 1 other ancestor2004: 2 other ancestors1992: 4 other ancestors1991: 1 other ancestor1994: 2 other ancestors1990: 2 other ancestors1989: 2 other ancestors1987: 1 other ancestor1986: 1 other ancestor1981: 1 other ancestor2021: 2 descendants cleared2022: 1 descendant cleared2023: 1 descendant cleared2025: 2 descendants clearedK010975 · Eclipse 6.5 · Varian Medical Systems, Inc. · 2001 · recalledK030981 · Eclipse · Varian Medical Systems, Inc. · 2003K041403 · Eclipse with Aaa · Varian Medical Systems · 2004K050296 · Eclipse Treatment Planning System · Varian Medical Systems · 2005×2: 2 generations cleared close together, K071873 (2007), K073020 (2007) K091492 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2009 · recalledK102216 · Xio RTP System · Computerized Medical Systems, Inc. · 2010 · recalled×2: 2 generations cleared close together, K140187 (2014), K141860 (2014) K160093 · RayStation · RaySearch Laboratories AB (PUBL) · 2016 · recalled×2: 2 generations cleared close together, K170355 (2017), K171536 (2017) K180379 · RayStation 7 · RaySearch Laboratories AB (PUBL) · 2018 · recalledK190387 · RayStation · RaySearch Laboratories AB (PUBL) · 2019 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 8 generations back: 1 other ancestor9 generations back: 4 other ancestors10 generations back: 7 other ancestors11 generations back: 11 other ancestors12 generations back: 14 other ancestors13 generations back: 14 other ancestors14 generations back: 4 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 1 descendant K190387 · RayStation · RaySearch Laboratories AB (PUBL) · 2019 · recalledK180379 · RayStation 7 · RaySearch Laboratories AB (PUBL) · 2018 · recalledK171536 · RayStation · RaySearch Laboratories AB (PUBL) · 2017 · recalledK170355 · RayStation 6 · RaySearch Laboratories AB (PUBL) · 2017 · recalledK160093 · RayStation · RaySearch Laboratories AB (PUBL) · 2016 · recalledK141860 · Raystation · Raysearch Laboratories AB · 2014 · recalledK140187 · Ray Station · Raysearch Laboratories AB · 2014 · recalledK102216 · Xio RTP System · Computerized Medical Systems, Inc. · 2010 · recalledK091492 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2009 · recalledK073020 · Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2007 · recalledK071873 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2007 · recalledK050296 · Eclipse Treatment Planning System · Varian Medical Systems · 2005K041403 · Eclipse with Aaa · Varian Medical Systems · 2004K030981 · Eclipse · Varian Medical Systems, Inc. · 2003K010975 · Eclipse 6.5 · Varian Medical Systems, Inc. · 2001 · recalled−15−10−50+5
  1. Root · 15 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
  2. 11 generations between
  3. 3 generations backK171536 RayStationRaySearch Laboratories AB (PUBL) · 2017 · Recalled
  4. 2 generations backK180379 RayStation 7RaySearch Laboratories AB (PUBL) · 2018 · Recalled
  5. 1 generation backK190387 RayStationRaySearch Laboratories AB (PUBL) · 2019 · Recalled
  6. This deviceK200569 RayStationRaySearch Laboratories AB (PUBL) · 2020
  7. Descendant · cites it directly · 5 of its ownK210645 RayStation 10.1RaySearch Laboratories AB (PUBL) · 2021 · Recalled
  8. Descendant · 2 generations on · 4 of its ownK211867 RayStation 11.0RaySearch Laboratories AB (PUBL) · 2021 · Recalled
  9. Descendant · 3 generations on · 3 of its ownK220141 RayStation 11BRaySearch Laboratories AB (PUBL) · 2022 · Recalled
  10. Descendant · 4 generations on · 2 of its ownK222312 RayStation 12ARaySearch Laboratories AB (PUBL) · 2023 · Recalled
  11. Descendant · 5 generations on · 1 of its ownK240398 RayStation 2023B, RayPlan 2023B, RayStation 2024A, RayPlan 2024A…RaySearch Laboratories AB (PUBL) · 2025 · Recalled
Show all 15 generations
  1. Root · 15 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
  2. 14 generations backK030981 EclipseVarian Medical Systems, Inc. · 2003
  3. 13 generations backK041403 Eclipse with AaaVarian Medical Systems · 2004
  4. 12 generations backK050296 Eclipse Treatment Planning SystemVarian Medical Systems · 2005
  5. 11 generations backK071873 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
  6. 10 generations backK073020 Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
  7. 9 generations backK091492 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2009 · Recalled
  8. 8 generations backK102216 Xio RTP SystemComputerized Medical Systems, Inc. · 2010 · Recalled
  9. 7 generations backK140187 Ray StationRaysearch Laboratories AB · 2014 · Recalled
  10. 6 generations backK141860 RaystationRaysearch Laboratories AB · 2014 · Recalled
  11. 5 generations backK160093 RayStationRaySearch Laboratories AB (PUBL) · 2016 · Recalled
  12. 4 generations backK170355 RayStation 6RaySearch Laboratories AB (PUBL) · 2017 · Recalled
  13. 3 generations backK171536 RayStationRaySearch Laboratories AB (PUBL) · 2017 · Recalled
  14. 2 generations backK180379 RayStation 7RaySearch Laboratories AB (PUBL) · 2018 · Recalled
  15. 1 generation backK190387 RayStationRaySearch Laboratories AB (PUBL) · 2019 · Recalled
  16. This deviceK200569 RayStationRaySearch Laboratories AB (PUBL) · 2020

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2021: 22022: 12023: 12025: 2
2021high 2 (2021) · low 0 (2024)2026

Most influential descendants

Showing 5 of 6 · See all
  • K210645RayStation 10.1RaySearch Laboratories AB (PUBL) · 2021 · cites it directly5descendants
  • K211867RayStation 11.0RaySearch Laboratories AB (PUBL) · 2021 · 2 generations on4descendants
  • K220141RayStation 11BRaySearch Laboratories AB (PUBL) · 2022 · 3 generations on3descendants
  • K222312RayStation 12ARaySearch Laboratories AB (PUBL) · 2023 · 4 generations on2descendants
  • K240398RayStation 2023B, RayPlan 2023B, RayStation 2024A, RayPlan 2024A, RayStation 2024A SP3…RaySearch Laboratories AB (PUBL) · 2025 · 5 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−15K010975Eclipse 6.5Varian Medical Systems, Inc.20013
−14K030981EclipseVarian Medical Systems, Inc.2003—
−13K041403Eclipse with AaaVarian Medical Systems2004—
−12K050296Eclipse Treatment Planning SystemVarian Medical Systems2005—
−11K071873Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20075
−10K073020Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20076
−9K091492Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20092
−8K102216Xio RTP SystemComputerized Medical Systems, Inc.20107
−7K140187Ray StationRaysearch Laboratories AB20141
−6K141860RaystationRaysearch Laboratories AB20145
−5K160093RayStationRaySearch Laboratories AB (PUBL)20167
−4K170355RayStation 6RaySearch Laboratories AB (PUBL)20171
−3K171536RayStationRaySearch Laboratories AB (PUBL)20176
−2K180379RayStation 7RaySearch Laboratories AB (PUBL)20187
−1K190387RayStationRaySearch Laboratories AB (PUBL)201932
0 (this device)K200569RayStationRaySearch Laboratories AB (PUBL)2020—
+1K210645RayStation 10.1RaySearch Laboratories AB (PUBL)20216
+2K211867RayStation 11.0RaySearch Laboratories AB (PUBL)20218
+3K220141RayStation 11BRaySearch Laboratories AB (PUBL)20226
+4K222312RayStation 12ARaySearch Laboratories AB (PUBL)20236
+5K240398RayStation 2023B, RayPlan 2023B, RayStation 2024A, RayPlan 2024A, RayStation…RaySearch Laboratories AB (PUBL)20252

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198710
198920
199020
199110
199240
199420
199650
199720
199940
200050
200150
200230
200320
200420
200510
200620
200710
200830
200920
201010
201110
201210
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202120
202210
202310
202520
FIG. 1 — Lineage of K200569, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUJ device

  1. #1 K232799 syngo.via RT Image Suitescore 18.7 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K190387RayStation
Summary page 4
…Classification name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ) 5.7 Predicate device RayStation 8.1 K190387 5.8 Device description RayStation 9.1/RayStation 9.2 are radiation therapy treatment planning systems, i.e. a software pr…
RaySearch Laboratories AB (PUBL)201932

Also cleared against K190387: 1 other device →

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K210645RayStation 10.1RaySearch Laboratories AB (PUBL)20215

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 8, 2025Class IIZ-2489-2025Under Investigation by firmDICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
Jan 16, 2025Class IIZ-1187-2025Under Investigation by firmInconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in…
Mar 28, 2024Class IIZ-1746-2024Software designPotential for reported SSD to be too high.
Mar 28, 2024Class IIZ-1745-2024Software designPotential for reported SSD to be too high.
Feb 23, 2024Class IIZ-1482-2024Under Investigation by firmA use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
Jun 1, 2022Class IIZ-1403-2022Software change controlAn issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.
Jun 8, 2021Class IIZ-2138-2021Under Investigation by firmInitial delivery positions will be set incorrectly when having setup beam(s) at the localization point.
Apr 21, 2021Class IIZ-1838-2021Under Investigation by firmEvaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present in…
Apr 7, 2021Class IIZ-1598-2021Software designWhen using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the…
Mar 25, 2021Class IIZ-1427-2021Under Investigation by firmFlags edited in RayCare after a patient has been checked in may not always be updated in RayTreat.
Feb 22, 2021Class IIZ-1597-2021Software designFor some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning…
Jan 22, 2021Class IIZ-2039-2023Software designA software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose statistics for an…
Nov 11, 2020Class IIZ-0589-2021Software designTwo issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch…
Oct 8, 2020Class IIZ-0478-2021Software designIf a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 5 companies; a new generation was cleared every 1.3 years on median.
  • 25 of the 70 ancestors have a recall of any class.
  • Of its descendants, K210645 (RaySearch Laboratories AB (PUBL), 2021) has gone furthest, with 5 descendants of its own.
  • 1 other device also cites its predicate K190387.

How does its category compare?

All MUJ devices

MUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · MUJ

10in 2025

2007: 42008: 92009: 162010: 112011: 92012: 82013: 52014: 92015: 132016: 52017: 72018: 72019: 92020: 92021: 112022: 82023: 102024: 92025: 102026 (partial year): 9
2007high 16 (2009) · low 4 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUJ

186in 2025

2007: 67 days median2008: 63 days median2009: 84 days median2010: 56 days median2011: 103 days median2012: 59 days median2013: 64 days median2014: 108 days median2015: 78 days median2016: 80 days median2017: 30 days median2018: 53 days median2019: 56 days median2020: 104 days median2021: 87 days median2022: 168 days median2023: 140 days median2024: 122 days median2025: 186 days median2026 (partial year): 233 days median
2007high 233 (2026) · low 30 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUJ

2in 2023

2007: 42008: 102009: 112010: 232011: 172012: 42013: 62014: 22015: 72016: 132017: 42018: 32019: 42020: 22021: 42022: 32023: 22026 (partial year): 2
2007high 23 (2010) · low 0 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUJ

21in 2025

2007: 242008: 512009: 242010: 782011: 732012: 392013: 332014: 352015: 302016: 272017: 162018: 252019: 252020: 262021: 152022: 222023: 762024: 92025: 21
2007high 78 (2010) · low 9 (2024)2025

MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007467424
20089631051
200916841124
201011562378
201191031773
2012859439
2013564633
20149108235
20151378730
20165801327
2017730416
2018753325
2019956425
20209104226
20211187415
20228168322
202310140276
2024912209
202510186021
202692332—

What is it indicated for?

RayStation is a software system designed for treatment planning and analysis of radiation therapy.

The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

The system functionality can be configured based on user needs.

The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K200569: RayStation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K200569

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026