510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Perfusor® Space Syringe Infusion Pump System

K172831510(k) TraditionalCleared Jun 12, 2018267-day reviewClass II21 CFR 880.5725Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

B.Braun Medical, Inc. · FRN Pump, Infusion

Cleared on Jun 12, 2018 as substantially equivalent to K092313 (B.Braun Medical, Inc., 2009) and 1 other predicate after a 267-day review. Its family tree reaches back 5 generations to 1990. 4 of the 19 devices in its family, counting itself, have a Class I recall. It has been recalled 4 times.

Decided
Jun 12, 2018
Review
267 days
Class
Class II
Product code
FRN
Recalls
4
Predicates
2

Its longest line of descent runs back 5 generations to K904518 (1990). 18 ancestors in all, the oldest K890120 (1989). No later device cites it as a predicate yet. 4 of the 19 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2003: 2 other ancestors2002: 1 other ancestor, 1 with a Class I recall12001: 1 other ancestor, 1 with a Class I recall11999: 1 other ancestor1998: 1 other ancestor1995: 2 other ancestors, 1 with a Class I recall11996: 1 other ancestor1994: 2 other ancestors1990: 1 other ancestor1989: 1 other ancestorReference device: K142596 · Infusomat Space Volumetric Infusion Pump · B.Braun Medical, Inc. · 2015 · Class I recallK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK022575 · Perfusor Compact with FM System · B.Braun Medical, Inc. · 2002K062699 · Perfusor Space Infusion Syringe Pump System · B.Braun Medical, Inc. · 2006K092313 · Perfusor Space Infusion Syringe Pump System · B.Braun Medical, Inc. · 2009 · recalled19851990199520002005201020152020
  1. Root · 5 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 4 generations backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. 3 generations backK022575 Perfusor Compact with FM SystemB.Braun Medical, Inc. · 2002
  4. 2 generations backK062699 Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc. · 2006
  5. 1 generation backK092313 Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc. · 2009 · Recalled
  6. This deviceK172831 Perfusor® Space Syringe Infusion Pump SystemB.Braun Medical, Inc. · 2018

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−4K994375Horizon LightB.Braun Medical, Inc.2000Class I
−3K022575Perfusor Compact with FM SystemB.Braun Medical, Inc.2002—
−2K062699Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc.2006—
−1K092313Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc.20094
0 (this device)K172831Perfusor® Space Syringe Infusion Pump SystemB.Braun Medical, Inc.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199010
199420
199521
199610
199810
199910
200111
200211
200320
FIG. 1 — Lineage of K172831, 1990–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092313Perfusor Space Infusion Syringe Pump System
Summary page 4
…FRN, pump, infusion LZH, pump, infusion, enteral Device Class: Class II Primary Predicate Device: K092313, Perfusor® Space Syringe Infusion Pump System Secondary Predicate Device: K062699, Perfusor® Space Syringe Infusion Pump S…
B.Braun Medical, Inc.20094
*K062699Perfusor Space Infusion Syringe Pump System
Summary page 4
…Device Class: Class II Primary Predicate Device: K092313, Perfusor® Space Syringe Infusion Pump System Secondary Predicate Device: K062699, Perfusor® Space Syringe Infusion Pump System Reference Device K142596, B. Braun Infusomat Space Volumetric I…
B.Braun Medical, Inc.2006—
†K142596Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc.2015Class I

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 14, 2024Class IIZ-2433-2024Device DesignPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility…
Jun 14, 2024Class IIZ-2434-2024Device DesignPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility…
Jun 14, 2024Class IIZ-2437-2024Device DesignPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility…
Feb 7, 2022Class IIZ-0819-2022Under Investigation by firmUnits not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion accuracy that can result in delays in therapy, overinfusion, or underinfusion.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · 2 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · 2 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall
  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K950419The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · 3 generations back · ◆ in the 1995 ancestors bar of FIG. 14 Class I recalls

What does this lineage say?

  • The line of descent spans 28 years and passes through 2 companies; a new generation was cleared every 4.2 years on median.
  • 10 of the 18 ancestors have a recall of any class; 4 have a Class I recall.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Perfusor® Space Syringe Infusion Pump System is intended for use on adults, pediatrics and neonates for the intermittent or continuous delivery of parenteral fluids, enteral fluids, medications, blood and blood products through clinically accepted routes of administration. These routes include intravenous, intra-arterial, subcutaneous, epidural and enteral.

The Perfusor® Space Syringe Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities and for medical ground and/or air transport situations.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172831: Perfusor® Space Syringe Infusion Pump System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172831

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026