Stimuplex Onvision System

K192914510(k) TraditionalCleared May 27, 2020225-day reviewClass II21 CFR 868.5150

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Nederland B.V. · BSP Needle, Conduction, Anesthetic (W/Wo Introducer)

Cleared on May 27, 2020 as substantially equivalent to K121812 (Ultrasonix Medical Corporation, 2013) and 2 other predicates after a 225-day review. Its family tree reaches back 8 generations to 1994. 7 of the 68 devices in its family, counting itself, have a Class I recall.

Decided
May 27, 2020
Review
225 days
Class
Class II
Product code
BSP
Recalls
None
Predicates
3

Its longest line of descent runs back 8 generations to K911783 (1994). 67 ancestors in all, the oldest K801912 (1980). No later device cites it as a predicate yet. 7 of the 68 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2013: 1 other ancestor2010: 1 other ancestor2009: 1 other ancestor2007: 2 other ancestors2002: 9 other ancestors2000: 7 other ancestors2004: 2 other ancestors, 1 with a Class I recall12003: 5 other ancestors1984: 1 other ancestor2005: 2 other ancestors2001: 7 other ancestors1999: 4 other ancestors1998: 7 other ancestors, 1 with a Class I recall11997: 2 other ancestors1996: 3 other ancestors, 1 with a Class I recall11995: 1 other ancestor1993: 1 other ancestor1988: 1 other ancestor1983: 1 other ancestor1980: 1 other ancestor, 1 with a Class I recall1K911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994×2: 2 generations cleared close together, K981998 (1998), K983529 (1998) K994270 · Instatrak 3000 System with Fluorotrak Module, Model IT3000 · Visualization Technology, Inc. · 2000 · Class I recallK053610 · Abaris · Traxtal Technologies, Inc. · 2006 · recalledK121498 · Percunav · Philips Health Care · 2012K132087 · Percunav System · Philips Health Care · 2013K170716 · PercuNav Image Fusion and Interventional Navigation · Philips Ultrasound, Inc. · 201719801990200020102020
  1. Root · 8 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 7 generations backK981998 Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
  3. 6 generations backK983529 Instatrak 3000Visualization Technology, Inc. · 1998 · Recalled (Class I)
  4. 5 generations backK994270 Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Recalled (Class I)
  5. 4 generations backK053610 AbarisTraxtal Technologies, Inc. · 2006 · Recalled
  6. 3 generations backK121498 PercunavPhilips Health Care · 2012
  7. 2 generations backK132087 Percunav SystemPhilips Health Care · 2013
  8. 1 generation backK170716 PercuNav Image Fusion and Interventional NavigationPhilips Ultrasound, Inc. · 2017
  9. This deviceK192914 Stimuplex Onvision SystemPhilips Medical Systems Nederland B.V. · 2020

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K911783Allegro Viewing WandI.S.G. Technologies, Inc.1994—
−7K981998Pediatric Instatrak SystemVisualization Technology, Inc.1998Class I
−6K983529Instatrak 3000Visualization Technology, Inc.1998Class I
−5K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc.2000Class I
−4K053610AbarisTraxtal Technologies, Inc.20063
−3K121498PercunavPhilips Health Care2012—
−2K132087Percunav SystemPhilips Health Care2013—
−1K170716PercuNav Image Fusion and Interventional NavigationPhilips Ultrasound, Inc.2017—
0 (this device)K192914Stimuplex Onvision SystemPhilips Medical Systems Nederland B.V.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198011
198310
198410
198810
199310
199510
199631
199720
199871
199940
200070
200170
200290
200350
200421
200520
200720
200910
201010
201310
FIG. 1 — Lineage of K192914, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSP device

  1. #1 K241954 SonoBlock; SonoBlock IIscore 11.0 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K121812Sonixgps
Summary page 4
…Product code IYO PREDICATE DEVICES: • Primary predicate device: SonixGPS Nerve Block Needle Kit, Ultrasonix Medical Corporation (K121812) • Additional predicate devices: o PercuNav Image Fusion and Interventional Navigation (K170716) o B.Bra…
Ultrasonix Medical Corporation2013—
*K170716PercuNav Image Fusion and Interventional Navigation
Summary page 4
…Block Needle Kit, Ultrasonix Medical Corporation (K121812) • Additional predicate devices: o PercuNav Image Fusion and Interventional Navigation (K170716) o B.Braun Stimuplex® D Plus Insulated Echogenic Needle(K100241) DEVICE DESCRIPTION: The proposed Stimuplex® On…
Philips Ultrasound, Inc.2017—
*K100241Contiplex D Continuous Nerve Block Set, Model 331687, 331686, 331685, 331680
Summary page 4
…Additional predicate devices: o PercuNav Image Fusion and Interventional Navigation (K170716) o B.Braun Stimuplex® D Plus Insulated Echogenic Needle(K100241) DEVICE DESCRIPTION: The proposed Stimuplex® Onvision® system is an add-on device to the Philips Xperius ultrasoun…
B.Braun Medical, Inc.2010—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (7 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K040050Instatrak System with Multiple Dataset Navigation, Model IT3500Ge Medical Systems Navigation and Visualization IN · 2004 · 5 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 5 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · 5 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 5 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall
  • K801912Arrow Positive Placement Con/Epidural anArrow Intl., Inc. · 1980 · 6 generations back · ◆ in the 1980 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 6 companies; a new generation was cleared every 3.7 years on median.
  • 22 of the 67 ancestors have a recall of any class; 7 have a Class I recall.

How does its category compare?

All BSP devices

BSP Needle, Conduction, Anesthetic (W/Wo Introducer): FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · BSP

4in 2025

2007: 32008: 82009: 52010: 32011: 32012: 32013: 22014: 22015: 22016: 62017: 42018: 82019: 22020: 32021: 32022: 22024: 32025: 4
2007high 8 (2008) · low 0 (2023)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSP

170in 2025

2007: 189 days median2008: 127 days median2009: 97 days median2010: 79 days median2011: 114 days median2012: 148 days median2016: 268 days median2017: 186 days median2018: 241 days median2020: 269 days median2021: 265 days median2024: 70 days median2025: 170 days median
2007high 269 (2020) · low 70 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSP

1in 2020

2016: 12019: 12020: 12026 (partial year): 1, 1 Class I
2007high 1 (2016) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · BSP

86in 2025

2007: 262008: 242009: 302010: 492011: 332012: 202013: 322014: 412015: 382016: 1522017: 342018: 692019: 862020: 782021: 522022: 842023: 752024: 662025: 86
2007high 152 (2016) · low 20 (2012)2025

MAUDE reports naming product code BSP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073189026
20088127024
2009597030
2010379049
20113114033
20123148020
20132—032
20142—041
20152—038
201662681152
20174186034
20188241069
20192—186
20203269178
20213265052
20222—084
20230—075
2024370066
20254170086
20260—1 (1)—

What is it indicated for?

The Stimuplex® Onvision® system is intended for clinically trained physicians to perform regional anesthesia and pain therapy to target peripheral nerves by tracking the B.Braun Stimuplex® Onvision® needle tip position by overlaying a circle at the given location using the Philips Xperius® Ultrasound and Onvision® System. The B.Braun Stimuplex® Onvision® needle is used to inject local anesthesia or analgesic to the targeted nerve bundle to induce regional anesthesia for surgical procedures. The Stimuplex® HNS 12 can also be used for nerve stimulation when connected to the system cable.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K192914: Stimuplex Onvision System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K192914

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.