510(k) Report

510(k) · Cleared

Cleared

nView s1 with nav option

K211064510(k) TraditionalCleared Oct 29, 2021203-day reviewClass II21 CFR 892.1650

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nview Medical · OWB Interventional Fluoroscopic X-Ray System

Cleared on Oct 29, 2021 as substantially equivalent to K190064 (Nview Medical, Inc., 2019) after a 203-day review. Its family tree reaches back 8 generations to 2000. 3 of the 17 devices in its family, counting itself, have a Class I recall. 1 later device descends from it.

Decided
Oct 29, 2021
Review
203 days
Class
Class II
Product code
OWB
Recalls
None
Predicates
1

Its longest line of descent runs back 8 generations to K003266 (2000). 15 ancestors in all. 1 later device descends from it, 1 citing it directly. 3 of the 17 devices in this family, counting this one, have a Class I recall.

  1. Root · 8 generations backK003266 Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc. · 2000
  2. 7 generations backK032280 Siremobil Iso-C 3DSiemens AG · 2003
  3. 6 generations backK061310 Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc. · 2006
  4. 5 generations backK073346 Ziehm Vision 2Ziehm Imaging, Inc. · 2008 · Recalled
  5. 4 generations backK083545 Ziehm Vision RFDZiehm Imaging, Inc. · 2009 · Recalled
  6. 3 generations backK132904 Ziehm Vision RFDZiehm Imaging GmbH · 2013 · Recalled
  7. 2 generations backK142740 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2015 · Recalled
  8. 1 generation backK190064 nView system 1Nview Medical, Inc. · 2019
  9. This deviceK211064 nView s1 with nav optionNview Medical · 2021
  10. Descendant · cites it directly · 0 of its ownK221344 nView s1 with nav optionNview Medical · 2022

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K221344nView s1 with nav optionNview Medical · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K003266Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc.2000—
−7K032280Siremobil Iso-C 3DSiemens AG2003—
−6K061310Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc.2006—
−5K073346Ziehm Vision 2Ziehm Imaging, Inc.20085
−4K083545Ziehm Vision RFDZiehm Imaging, Inc.20093
−3K132904Ziehm Vision RFDZiehm Imaging GmbH20131
−2K142740Ziehm Vision RFD 3DZiehm Imaging GmbH20151
−1K190064nView system 1Nview Medical, Inc.2019—
0 (this device)K211064nView s1 with nav optionNview Medical2021—
+1K221344nView s1 with nav optionNview Medical2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200131
200211
200311
200620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
FIG. 1 — Lineage of K211064, 2000–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K190064nView system 1
Summary page 4
…OWB Additional Product Codes: OXO, JAA, JAK, OLO Primary Predicate Device Information Device Name (510(k) number): nView medical - nView system 1 (K190064) Common/Usual Name: Mobile Fluoroscopic C-Arm Regulation Number: CFR 892.1650 Regulation Name…
Nview Medical, Inc.2019—
†K162309StealthStation S8 System Platforms and StealthStation Cranial SoftwareMedtronic Navigation, Inc.2017Class I

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K221344nView s1 with nav optionNview Medical20220

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003 · 6 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · 6 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 6 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 21 years and passes through 6 companies; a new generation was cleared every 2.8 years on median.
  • 9 of the 15 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K221344 (Nview Medical, 2022) has gone furthest, with 0 descendants of its own.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The nView sl is intended as an imaging system to provide both 2D and 3D imaging of adult and pediatric populations over 6 years of age. The device is intended to provide fluoroscopic and tomographic imaging of patients during orthopedic surgical procedures where the clinician benefits from 3D visualization of complex anatomical structures, such as high contrast objects, bones, joints, cervical, thoracic, and lumbar regions of the spine, and joint…

Read the full statement (184 words)

The nView sl is intended as an imaging system to provide both 2D and 3D imaging of adult and pediatric populations over 6 years of age. The device is intended to provide fluoroscopic and tomographic imaging of patients during orthopedic surgical procedures where the clinician benefits from 3D visualization of complex anatomical structures, such as high contrast objects, bones, joints, cervical, thoracic, and lumbar regions of the spine, and joint fractures of the upper and lower extremities.

The nView sl is indicated to image human anatomy up to 30 cm thickness.

The nView sl is not indicated for mammographic or lung nodule applications.

The nView sl with navigation option is intended as a navigation system to aid in pilot hole verification for pedicle screw placement of 4.5 mm screws in the thoracic and lumbar spine regions. The navigation option is indicated for posterior approach open spinal procedures in pediatric populations over 6 years of age in which the use of stereotactic surgery may be appropriate, and where reference to a rigid spinous process can be identified relative to nView s1 images of the anatomy.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K211064: nView s1 with nav option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K211064

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026