Breast Volume Navigator (BVN) Model G-2000

K211098510(k) TraditionalCleared May 26, 202143-day reviewClass II21 CFR 892.1560

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Metritrack, Inc. · IYO System, Imaging, Pulsed Echo, Ultrasonic

Cleared on May 26, 2021 as substantially equivalent to K141870 (Metritrack, LLC, 2014) and 2 other predicates after a 43-day review. Its family tree reaches back 9 generations to 1994. 6 of the 104 devices in its family, counting itself, have a Class I recall.

Decided
May 26, 2021
Review
43 days
Class
Class II
Product code
IYO
Recalls
None
Predicates
3

Its longest line of descent runs back 9 generations to K911783 (1994). 103 ancestors in all, the oldest K844472 (1985). No later device cites it as a predicate yet. 6 of the 104 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2017: 1 other ancestor2014: 1 other ancestor2013: 1 other ancestor2005: 7 other ancestors2000: 10 other ancestors1996: 7 other ancestors, 1 with a Class I recall12009: 1 other ancestor2003: 11 other ancestors2002: 10 other ancestors1995: 3 other ancestors2008: 2 other ancestors2007: 3 other ancestors2004: 3 other ancestors, 1 with a Class I recall11998: 8 other ancestors, 1 with a Class I recall11997: 3 other ancestors1988: 1 other ancestor2006: 1 other ancestor2001: 8 other ancestors1999: 6 other ancestors1993: 1 other ancestor1992: 1 other ancestor1985: 2 other ancestors1986: 1 other ancestor1994: 1 other ancestor1991: 1 other ancestorK911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994×2: 2 generations cleared close together, K981998 (1998), K983529 (1998) K994270 · Instatrak 3000 System with Fluorotrak Module, Model IT3000 · Visualization Technology, Inc. · 2000 · Class I recallK053610 · Abaris · Traxtal Technologies, Inc. · 2006 · recalledK080624 · Navstationirad Emt, Version 1.1 · Cas Innovations AG · 2008K092619 · Etrax Needle System, Etrax Variable Angle Needle Guidance System, Virtutrax Universal Tracker · Civco · 2009 · recalledK131489 · Tractus Tissuemapper Image Recording System Application · Tractus Corporation · 2013K171309 · BR-FHUS Navigation 1.0 · Taihao Medical, Inc. · 20171990200020102020
  1. Root · 9 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 5 generations between
  3. 3 generations backK092619 Etrax Needle System, Etrax Variable Angle Needle Guidance System…Civco · 2009 · Recalled
  4. 2 generations backK131489 Tractus Tissuemapper Image Recording System ApplicationTractus Corporation · 2013
  5. 1 generation backK171309 BR-FHUS Navigation 1.0Taihao Medical, Inc. · 2017
  6. This deviceK211098 Breast Volume Navigator (BVN) Model G-2000Metritrack, Inc. · 2021
Show all 9 generations
  1. Root · 9 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 8 generations backK981998 Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
  3. 7 generations backK983529 Instatrak 3000Visualization Technology, Inc. · 1998 · Recalled (Class I)
  4. 6 generations backK994270 Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Recalled (Class I)
  5. 5 generations backK053610 AbarisTraxtal Technologies, Inc. · 2006 · Recalled
  6. 4 generations backK080624 Navstationirad Emt, Version 1.1Cas Innovations AG · 2008
  7. 3 generations backK092619 Etrax Needle System, Etrax Variable Angle Needle Guidance System…Civco · 2009 · Recalled
  8. 2 generations backK131489 Tractus Tissuemapper Image Recording System ApplicationTractus Corporation · 2013
  9. 1 generation backK171309 BR-FHUS Navigation 1.0Taihao Medical, Inc. · 2017
  10. This deviceK211098 Breast Volume Navigator (BVN) Model G-2000Metritrack, Inc. · 2021

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K911783Allegro Viewing WandI.S.G. Technologies, Inc.1994—
−8K981998Pediatric Instatrak SystemVisualization Technology, Inc.1998Class I
−7K983529Instatrak 3000Visualization Technology, Inc.1998Class I
−6K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc.2000Class I
−5K053610AbarisTraxtal Technologies, Inc.20063
−4K080624Navstationirad Emt, Version 1.1Cas Innovations AG2008—
−3K092619Etrax Needle System, Etrax Variable Angle Needle Guidance System, Virtutrax…Civco200917
−2K131489Tractus Tissuemapper Image Recording System ApplicationTractus Corporation2013—
−1K171309BR-FHUS Navigation 1.0Taihao Medical, Inc.2017—
0 (this device)K211098Breast Volume Navigator (BVN) Model G-2000Metritrack, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198520
198610
198810
199110
199210
199310
199410
199530
199671
199730
199881
199960
2000100
200180
2002100
2003110
200431
200570
200610
200730
200820
200910
201310
201410
201710
FIG. 1 — Lineage of K211098, 1994–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYO device

  1. #1 K243937 Accuro 3Sscore 20.5 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141870Breast Volume Navigator
Summary page 4
…evice Class: II Primary Product Code IYO, ITX, LLZ Predicate Devices: Primary predicate: MetriTrack, LLC Breast Volume Navigator (BVN™) (K141870) Reference device: TaiHao Medical Inc. BR-FHUS Navigation 1.0 (K171309) Reference device: TaiHao Medical Inc. BR-FHUS Viewer (K17…
Metritrack, LLC2014—
*K171309BR-FHUS Navigation 1.0
Summary page 4
…O, ITX, LLZ Predicate Devices: Primary predicate: MetriTrack, LLC Breast Volume Navigator (BVN™) (K141870) Reference device: TaiHao Medical Inc. BR-FHUS Navigation 1.0 (K171309) Reference device: TaiHao Medical Inc. BR-FHUS Viewer (K171709) Indications for Use / Intended Use: The Breast Volume Navigator…
Taihao Medical, Inc.2017—
*K171709BR-FHUS Viewer 1.0
Summary page 4
…LLC Breast Volume Navigator (BVN™) (K141870) Reference device: TaiHao Medical Inc. BR-FHUS Navigation 1.0 (K171309) Reference device: TaiHao Medical Inc. BR-FHUS Viewer (K171709) Indications for Use / Intended Use: The Breast Volume Navigator (BVN™) is an add-on accessory for existing ultrasound imaging s…
Taihao Medical, Inc.2017—

Also cleared against K141870: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (6 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K040050Instatrak System with Multiple Dataset Navigation, Model IT3500Ge Medical Systems Navigation and Visualization IN · 2004 · 4 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 5 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · 5 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 5 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 27 years and passes through 8 companies; a new generation was cleared every 3.6 years on median.
  • 28 of the 103 ancestors have a recall of any class; 6 have a Class I recall.
  • 1 other device also cites its predicate K141870.

How does its category compare?

All IYO devices

IYO System, Imaging, Pulsed Echo, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · IYO

7in 2025

2007: 202008: 142009: 132010: 152011: 112012: 152013: 172014: 142015: 192016: 102017: 192018: 142019: 132020: 112021: 152022: 52023: 112024: 102025: 72026 (partial year): 10
2007high 20 (2007) · low 5 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYO

171in 2025

2007: 17 days median2008: 36 days median2009: 79 days median2010: 113 days median2011: 69 days median2012: 68 days median2013: 50 days median2014: 63 days median2015: 46 days median2016: 36 days median2017: 72 days median2018: 56 days median2019: 59 days median2020: 57 days median2021: 64 days median2022: 190 days median2023: 177 days median2024: 220 days median2025: 171 days median2026 (partial year): 139 days median
2007high 220 (2024) · low 17 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYO

1in 2022

2007: 42008: 52009: 92010: 82012: 22013: 32014: 22015: 12018: 12019: 12021: 112022: 12026 (partial year): 1
2007high 11 (2021) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYO

485in 2025

2007: 392008: 202009: 272010: 202011: 582012: 782013: 4362014: 3042015: 2682016: 2542017: 2382018: 3682019: 5372020: 4982021: 6172022: 5112023: 5272024: 6402025: 485
2007high 640 (2024) · low 20 (2008)2025

MAUDE reports naming product code IYO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072017439
20081436520
20091379927
201015113820
20111169058
20121568278
201317503436
201414632304
201519461268
201610360254
201719720238
201814561368
201913591537
202011570498
2021156411617
202251901511
2023111770527
2024102200640
202571710485
2026101391—

What is it indicated for?

The Breast Volume Navigator (BVN) is an add-on accessory for existing ultrasound imaging systems and is intended to control position and movement of ultrasound transducers for the systematic acquisition of 2-dimensional image slices throughout a volume of interest. The BVN is intended to acquire, analyze, store and retrieve digital ultrasound images for computerized 3-dimensional image processing. The BVN is intended to assist health care providers…

Read the full statement (318 words)

The Breast Volume Navigator (BVN) is an add-on accessory for existing ultrasound imaging systems and is intended to control position and movement of ultrasound transducers for the systematic acquisition of 2-dimensional image slices throughout a volume of interest. The BVN is intended to acquire, analyze, store and retrieve digital ultrasound images for computerized 3-dimensional image processing.

The BVN is intended to assist health care providers trained in breast ultrasound with tools for electromagnetic tracking of instruments with respect of breast ultrasound images generated from FDA cleared hand-held ultrasound devices.

The BVN is indicated for use as an adjunct to mammography for B-mode ultrasonic imaging of a patient's breast when used with a linear array transducer.

The BVN is indicated for use as an adjunct to hand-held breast ultrasound to assist the health care providers trained in breast ultrasound in their scanning process. The scanning paths are displayed on a route map and provide quality control of scanning to provide an overall observation of scanning process.

The BVN Software Application is intended as a standalone software device installed on a windows-based computer to assist physicians with manipulation and analysis tools in reviewing breast ultrasound images. Images and data are previously recorded from various imaging systems and other sources such as calibrated spatial positioning devices. The BVN Software Application provides the capability to visualize ultrasound images along with the scanning paths and position information of the probe that is stored in the DICOM file in advance.

The BVN will allow exporting to any third-party application that has the appropriate level of DICOM compliance. The BVN is intended as a general-purpose digital 3D breast ultrasound image processing tool for radiology and surgery.

The device is not intended to be used in the environment of strong magnetic or electromagnetic fields, such as in Magnetic Resonance Imaging (MRI) room.

The device is not intended to be used as a replacement for screening mammography.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K211098: Breast Volume Navigator (BVN) Model G-2000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K211098

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026