510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy Analysis

K230162510(k) TraditionalCleared Aug 1, 2023193-day reviewClass II21 CFR 892.1750Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shanghai United Imaging Healthcare Co., Ltd. · JAK System, X-Ray, Tomography, Computed

Cleared on Aug 1, 2023 as substantially equivalent to K172135 (Shanghai United Imaging Healthcare Co., Ltd., 2018) and 2 other predicates after a 193-day review. Its family tree reaches back 15 generations to 2001. 2 later devices descend from it. It has been recalled 4 times.

Decided
Aug 1, 2023
Review
193 days
Class
Class II
Product code
JAK
Recalls
4
Predicates
3

Its longest line of descent runs back 15 generations to K013561 (2001). 58 ancestors in all, the oldest K930090 (1993). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2021: 1 other ancestor2018: 3 other ancestors2003: 1 other ancestor2019: 2 other ancestors2001: 3 other ancestors1998: 4 other ancestors2017: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2016: 1 other ancestor2015: 5 other ancestors1993: 1 other ancestor2014: 4 other ancestors2012: 3 other ancestors2008: 1 other ancestor2011: 1 other ancestor2006: 1 other ancestor2005: 1 other ancestor2004: 2 other ancestors2002: 1 other ancestor2000: 1 other ancestor1999: 1 other ancestor2025: 1 descendant cleared2026: 1 descendant clearedK013561 · Lightspeed 4.0 CT Scanner System · GE Medical Systems · 2001 · recalledK030420 · Lightspeed 5.0 Computed Tomography System · GE Medical Systems · 2003 · recalledK040372 · Lightspeed 7.0 CT Scanner System · Ge Medical Systems, LLC · 2004 · recalledK052216 · Somatom Definition · Siemens Medical Systems, Inc. · 2005K082220 · Somatom Flash DS · Siemens Medical Solutions USA, Inc. · 2008×2: 2 generations cleared close together, K113342 (2011), K121072 (2012) K133589 · Somatom Force · Siemens Medical Solutions USA, Inc. · 2014 · recalledK142955 · SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open, SOMATON Emotion 6/16, SOMATOM Sensation 64/Sensation Cardiac , SOMATOM Perspective SOMATOM Scope/ Scope Power · Siemens Medical Solutions USA, Inc. · 2015 · recalledK161196 · SOMATOM Drive · Siemens Medical Solutions USA, Inc. · 2016 · recalledK163296 · SOMATOM go.Up, SOMATOM go.Now · Siemens Medical Solutions USA, Inc. · 2017 · recalledK173632 · SOMATOM go.Up, SOMATOM go.Now, SOMATOM go.Top, SOMATOM go.All, Scan&GO · Siemens Medical Solutions USA, Inc. · 2018 · recalledK191891 · SOMATOM X.cite · Siemens Medical Solutions USA, Inc. · 2019K200524 · SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All, go. Sim, go. Open Pro, Scan&GO · Siemens Medical Solutions USA, Inc. · 2020 · recalledK211373 · SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform Scanners - SOMATOM X.cite and SOMATOM X.ceed, Scan&GO Software · Siemens Medical Solutions USA, Inc. · 2021 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 1 other ancestor3 generations back: 3 other ancestors4 generations back: 2 other ancestors5 generations back: 7 other ancestors6 generations back: 8 other ancestors7 generations back: 9 other ancestors8 generations back: 5 other ancestors9 generations back: 4 other ancestors10 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K211373 · SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform Scanners - SOMATOM X.cite and SOMATOM X.ceed, Scan&GO Software · Siemens Medical Solutions USA, Inc. · 2021 · recalledK200524 · SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All, go. Sim, go. Open Pro, Scan&GO · Siemens Medical Solutions USA, Inc. · 2020 · recalledK191891 · SOMATOM X.cite · Siemens Medical Solutions USA, Inc. · 2019K173632 · SOMATOM go.Up, SOMATOM go.Now, SOMATOM go.Top, SOMATOM go.All, Scan&GO · Siemens Medical Solutions USA, Inc. · 2018 · recalledK163296 · SOMATOM go.Up, SOMATOM go.Now · Siemens Medical Solutions USA, Inc. · 2017 · recalledK161196 · SOMATOM Drive · Siemens Medical Solutions USA, Inc. · 2016 · recalledK142955 · SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open, SOMATON Emotion 6/16, SOMATOM Sensation 64/Sensation Cardiac , SOMATOM Perspective SOMATOM Scope/ Scope Power · Siemens Medical Solutions USA, Inc. · 2015 · recalledK133589 · Somatom Force · Siemens Medical Solutions USA, Inc. · 2014 · recalledK121072 · Somatom Definition Flash · Siemens Medical Systems, Inc. · 2012 · recalledK113342 · Somatom Definition Flash · Siemens Medical Systems, Inc. · 2011 · recalledK082220 · Somatom Flash DS · Siemens Medical Solutions USA, Inc. · 2008K052216 · Somatom Definition · Siemens Medical Systems, Inc. · 2005K040372 · Lightspeed 7.0 CT Scanner System · Ge Medical Systems, LLC · 2004 · recalledK030420 · Lightspeed 5.0 Computed Tomography System · GE Medical Systems · 2003 · recalledK013561 · Lightspeed 4.0 CT Scanner System · GE Medical Systems · 2001 · recalled−15−10−50
  1. Root · 15 generations backK013561 Lightspeed 4.0 CT Scanner SystemGE Medical Systems · 2001 · Recalled
  2. 11 generations between
  3. 3 generations backK191891 SOMATOM X.citeSiemens Medical Solutions USA, Inc. · 2019
  4. 2 generations backK200524 SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now…Siemens Medical Solutions USA, Inc. · 2020 · Recalled
  5. 1 generation backK211373 SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All…Siemens Medical Solutions USA, Inc. · 2021 · Recalled
  6. This deviceK230162 uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual…Shanghai United Imaging Healthcare Co., Ltd. · 2023
  7. Descendant · cites it directly · 1 of its ownK241079 uCT 780Shanghai United Imaging Healthcare Co., Ltd. · 2025 · Recalled
  8. Descendant · 2 generations on · 0 of its ownK253173 uCT 780 with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2026 · Recalled
Show all 15 generations
  1. Root · 15 generations backK013561 Lightspeed 4.0 CT Scanner SystemGE Medical Systems · 2001 · Recalled
  2. 14 generations backK030420 Lightspeed 5.0 Computed Tomography SystemGE Medical Systems · 2003 · Recalled
  3. 13 generations backK040372 Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC · 2004 · Recalled
  4. 12 generations backK052216 Somatom DefinitionSiemens Medical Systems, Inc. · 2005
  5. 11 generations backK082220 Somatom Flash DSSiemens Medical Solutions USA, Inc. · 2008
  6. 10 generations backK113342 Somatom Definition FlashSiemens Medical Systems, Inc. · 2011 · Recalled
  7. 9 generations backK121072 Somatom Definition FlashSiemens Medical Systems, Inc. · 2012 · Recalled
  8. 8 generations backK133589 Somatom ForceSiemens Medical Solutions USA, Inc. · 2014 · Recalled
  9. 7 generations backK142955 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge…Siemens Medical Solutions USA, Inc. · 2015 · Recalled
  10. 6 generations backK161196 SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · Recalled
  11. 5 generations backK163296 SOMATOM go.Up, SOMATOM go.NowSiemens Medical Solutions USA, Inc. · 2017 · Recalled
  12. 4 generations backK173632 SOMATOM go.Up, SOMATOM go.Now, SOMATOM go.Top, SOMATOM go.All, Scan&GOSiemens Medical Solutions USA, Inc. · 2018 · Recalled
  13. 3 generations backK191891 SOMATOM X.citeSiemens Medical Solutions USA, Inc. · 2019
  14. 2 generations backK200524 SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now…Siemens Medical Solutions USA, Inc. · 2020 · Recalled
  15. 1 generation backK211373 SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All…Siemens Medical Solutions USA, Inc. · 2021 · Recalled
  16. This deviceK230162 uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual…Shanghai United Imaging Healthcare Co., Ltd. · 2023

Most influential descendants

Showing 2 of 2 · See all
  • K241079uCT 780Shanghai United Imaging Healthcare Co., Ltd. · 2025 · cites it directly1descendant
  • K253173uCT 780 with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−15K013561Lightspeed 4.0 CT Scanner SystemGE Medical Systems20015
−14K030420Lightspeed 5.0 Computed Tomography SystemGE Medical Systems20034
−13K040372Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC20045
−12K052216Somatom DefinitionSiemens Medical Systems, Inc.2005—
−11K082220Somatom Flash DSSiemens Medical Solutions USA, Inc.2008—
−10K113342Somatom Definition FlashSiemens Medical Systems, Inc.20111
−9K121072Somatom Definition FlashSiemens Medical Systems, Inc.20125
−8K133589Somatom ForceSiemens Medical Solutions USA, Inc.20149
−7K142955SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM…Siemens Medical Solutions USA, Inc.201516
−6K161196SOMATOM DriveSiemens Medical Solutions USA, Inc.20164
−5K163296SOMATOM go.Up, SOMATOM go.NowSiemens Medical Solutions USA, Inc.20176
−4K173632SOMATOM go.Up, SOMATOM go.Now, SOMATOM go.Top, SOMATOM go.All, Scan&GOSiemens Medical Solutions USA, Inc.201818
−3K191891SOMATOM X.citeSiemens Medical Solutions USA, Inc.2019—
−2K200524SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All…Siemens Medical Solutions USA, Inc.20207
−1K211373SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM…Siemens Medical Solutions USA, Inc.20213
0 (this device)K230162uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy…Shanghai United Imaging Healthcare Co., Ltd.2023—
+1K241079uCT 780Shanghai United Imaging Healthcare Co., Ltd.20252
+2K253173uCT 780 with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd.20261

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199420
199510
199610
199710
199840
199910
200010
200130
200210
200310
200420
200510
200610
200810
201110
201230
201440
201550
201610
201710
201830
201920
202110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
202610
FIG. 1 — Lineage of K230162, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K172135uCT Computed Tomography X-ray System
Summary page 4
…ode: JAK Review Panel: Radiology 4. Identification of Predicate Device(s) Primary Predicate Device 510(k) Number: K172135 Device Name: uCT 760, uCT 780 Regulation Name: Computed Tomography X-ray System…
Shanghai United Imaging Healthcare Co., Ltd.20187
*K203448uCT ATLAS with uWS-CT-Dual Energy Analysis
Summary page 5
…Regulatory Class: II Product Code: JAK Review Panel: Radiology Secondary Predicate Device 1 510(k) Number: K203448 Device Name: uCT ATLAS with uWS-CT-Dual Energy Analysis Regulation Name: Computed Tomography X-ray Sy…
Shanghai United Imaging Healthcare Co., Ltd.20213
*K211373SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X.…
Summary page 5
…Regulatory Class: II Product Code: JAK Review Panel: Radiology Secondary Predicate Device 2 510(k) Number: K211373 Device Name: SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM g…
Siemens Medical Solutions USA, Inc.20213

Also cleared against K172135: 4 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K241079uCT 780Shanghai United Imaging Healthcare Co., Ltd.20251

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 11, 2026Class IIZ-3279-2026Software designPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on…
Aug 11, 2026Class IIZ-3280-2026Software designPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on…
Sep 12, 2025Class IIZ-0219-2026OtherThird-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Sep 12, 2025Class IIZ-0220-2026OtherThird-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 5 companies; a new generation was cleared every 1.4 years on median.
  • 44 of the 58 ancestors have a recall of any class.
  • Of its descendants, K241079 (Shanghai United Imaging Healthcare Co., Ltd., 2025) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K172135.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

uCT 760/780 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac and vascular). uCT 760/780 is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the…

Read the full statement (210 words)

uCT 760/780 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac and vascular).

uCT 760/780 is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K230162: uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy Analysis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K230162

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026