510(k) · Cleared
Recalled · Class II · 2026uCT ATLAS with uWS-CT-Dual Energy Analysis
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Shanghai United Imaging Healthcare Co., Ltd. · JAK System, X-Ray, Tomography, Computed
Cleared on Aug 11, 2021 as substantially equivalent to K172135 (Shanghai United Imaging Healthcare Co., Ltd., 2018) after a 261-day review. Its family tree reaches back 6 generations to 1993. 6 later devices descend from it. It has been recalled 3 times.
- Decided
- Aug 11, 2021
- Review
- 261 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 3
- Predicates
- 1
Its longest line of descent runs back 6 generations to K930090 (1993). 11 ancestors in all. 6 later devices descend from it, 3 citing it directly.
- Root · 6 generations backK930090 HelicatElscint, Inc. · 1993
- 5 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
- 4 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
- 3 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
- 2 generations backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
- 1 generation backK172135 uCT Computed Tomography X-ray SystemShanghai United Imaging Healthcare Co., Ltd. · 2018 · Recalled
- This deviceK203448 uCT ATLAS with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2021
- Descendant · cites it directly · 2 of its ownK230162 uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual…Shanghai United Imaging Healthcare Co., Ltd. · 2023 · Recalled
- Descendant · cites it directly · 2 of its ownK223028 uCT ATLAS Astound with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2023
- Descendant · 2 generations on · 1 of its ownK241079 uCT 780Shanghai United Imaging Healthcare Co., Ltd. · 2025 · Recalled
- Descendant · cites it directly · 1 of its ownK231482 uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with…Shanghai United Imaging Healthcare Co., Ltd. · 2023 · Recalled
- Descendant · 3 generations on · 0 of its ownK253173 uCT 780 with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2026 · Recalled
Showing 5 of 6 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 6 · See all- K230162uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2023 · cites it directly2descendants
- K223028uCT ATLAS Astound with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2023 · cites it directly2descendants
- K241079uCT 780Shanghai United Imaging Healthcare Co., Ltd. · 2025 · 2 generations on1descendant
- K231482uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy…Shanghai United Imaging Healthcare Co., Ltd. · 2023 · cites it directly1descendant
- K253173uCT 780 with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2026 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K930090 | Helicat | Elscint, Inc. | 1993 | — |
| −5 | K961464 | Select/SP | Elscint, Inc. | 1996 | 3 |
| −4 | K982060 | Volumax CT Imaging System | Elscint, Inc. | 1998 | — |
| −3 | K012009 | MX8000 Idt CT System | Marconi Medical Systems, Inc. | 2001 | 61 |
| −2 | K033326 | Philips Plus Computed Tomography System | Philips Medical Systems, Inc. | 2003 | 131 |
| −1 | K172135 | uCT Computed Tomography X-ray System | Shanghai United Imaging Healthcare Co., Ltd. | 2018 | 7 |
| 0 (this device) | K203448 | uCT ATLAS with uWS-CT-Dual Energy Analysis | Shanghai United Imaging Healthcare Co., Ltd. | 2021 | — |
| +1 | K230162 | uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy… | Shanghai United Imaging Healthcare Co., Ltd. | 2023 | 4 |
| +1 | K223028 | uCT ATLAS Astound with uWS-CT-Dual Energy Analysis | Shanghai United Imaging Healthcare Co., Ltd. | 2023 | — |
| +2 | K241079 | uCT 780 | Shanghai United Imaging Healthcare Co., Ltd. | 2025 | 2 |
| +1 | K231482 | uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual… | Shanghai United Imaging Healthcare Co., Ltd. | 2023 | 3 |
| +3 | K253173 | uCT 780 with uWS-CT-Dual Energy Analysis | Shanghai United Imaging Healthcare Co., Ltd. | 2026 | 1 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1994 | 2 | 0 |
| 1995 | 1 | 0 |
| 1997 | 1 | 0 |
| 2001 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2023 | 3 | 0 |
| 2025 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K172135 | uCT Computed Tomography X-ray SystemSummary page 4…egulatory Class: II Product Code: JAK Review Panel: Radiology 4. Identification of Predicate/Reference Device(s) Predicate Device 510(k) Number: K172135 Device Name: uCT 760, uCT 780 Regulation Name: Computed Tomography X-ray System Regulatory Class: II Product… | 2018 |
| † | K161009 | Aquilion ONE Vision with FIRST 2.0 (CCRS-001B) V7.4 | 2016 |
| † | K193073 | Deep Recon | 2020 |
| † | K132813 | Dual Energy System Package | 2014 |
| † | K120833 | Discovery CT750 HD | 2012 |
What descends from it? (6)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 11, 2026 | Class II | Z-3282-2026 | Software design | Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on… |
| Aug 11, 2026 | Class II | Z-3281-2026 | Software design | Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on… |
| Sep 12, 2025 | Class II | Z-0221-2026 | Other | Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 28 years and passes through 4 companies; a new generation was cleared every 3.3 years on median.
- 5 of the 11 ancestors have a recall of any class.
- Of its descendants, K230162 (Shanghai United Imaging Healthcare Co., Ltd., 2023) has gone furthest, with 2 descendants of its own.
- 4 other devices also cite its predicate K172135.
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
uCT ATLAS is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography. uCT ATLAS has the capability to image a whole organ in a single rotation. Organs include, but not limited to head, heart, liver, kidney, pancreas…
Read the full statement (237 words)
uCT ATLAS is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.
uCT ATLAS has the capability to image a whole organ in a single rotation. Organs include, but not limited to head, heart, liver, kidney, pancreas, joints, etc.
uCT ATLAS is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
uWS-CT-Dual Energy Analysis software uses UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K203448 · printed Nov 6, 2026
Cite this page
510(k) Report. “K203448: uCT ATLAS with uWS-CT-Dual Energy Analysis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K203448
Report an extraction error