510(k) · Cleared
ClearedHamilton-C2
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Hamilton Medical AG · CBK Ventilator, Continuous, Facility Use
Cleared on Jan 21, 2011 as substantially equivalent to K081521 (Hamilton Medical AG, 2008) and 5 other predicates after a 119-day review. Its family tree reaches back 10 generations to 1991. 13 of the 82 devices in its family, counting itself, have a Class I recall. 14 later devices descend from it.
- Decided
- Jan 21, 2011
- Review
- 119 days
- Class
- Class II
- Product code
- CBK
- Recalls
- None
- Predicates
- 6
Its longest line of descent runs back 10 generations to K902859 (1991). 67 ancestors in all, the oldest K833982 (1984). 14 later devices descend from it, 2 citing it directly. 13 of the 82 devices in this family, counting this one, have a Class I recall.
- Root · 10 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
- 6 generations between
- 3 generations backK061090 Galileo Gold AsvHamilton Medical AG · 2006
- 2 generations backK070513 Hamilton-G5Hamilton Medical AG · 2007 · Recalled
- 1 generation backK081521 Modification to Hamilton-G5Hamilton Medical AG · 2008 · Recalled (Class I)
- This deviceK102775 Hamilton-C2Hamilton Medical AG · 2011
- Descendant · cites it directly · 13 of its ownK112006 Hamilton-TC1Hamilton Medical AG · 2012
- Descendant · 2 generations on · 11 of its ownK120670 Hamilton-T1Hamilton Medical AG · 2012 · Recalled (Class I)
- Descendant · cites it directly · 8 of its ownK121225 Hamilton-C2Hamilton Medical AG · 2012 · Recalled (Class I)
- Descendant · 2 generations on · 4 of its ownK123637 Hamilton-C3Hamilton Medical AG · 2012
- Descendant · 3 generations on · 3 of its ownK161450 Hamilton-C3Hamilton Medical AG · 2017 · Recalled
Show all 10 generations
- Root · 10 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
- 9 generations backK961687 EVITA 4Drager, Inc. · 1996
- 8 generations backK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · Recalled (Class I)
- 7 generations backK984535 Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · Recalled (Class I)
- 6 generations backK022679 RaphaelHamilton Medical AG · 2003 · Recalled
- 5 generations backK040574 Galileo GoldHamilton Medical, Inc. · 2004
- 4 generations backK051263 Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa · 2005 · Recalled
- 3 generations backK061090 Galileo Gold AsvHamilton Medical AG · 2006
- 2 generations backK070513 Hamilton-G5Hamilton Medical AG · 2007 · Recalled
- 1 generation backK081521 Modification to Hamilton-G5Hamilton Medical AG · 2008 · Recalled (Class I)
- This deviceK102775 Hamilton-C2Hamilton Medical AG · 2011
Showing 5 of 14 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 14 · See all- K112006Hamilton-TC1Hamilton Medical AG · 2012 · cites it directly13descendants
- K120670Hamilton-T1Hamilton Medical AG · 2012 · 2 generations on11descendants
- K121225Hamilton-C2Hamilton Medical AG · 2012 · cites it directly8descendants
- K123637Hamilton-C3Hamilton Medical AG · 2012 · 2 generations on4descendants
- K161450Hamilton-C3Hamilton Medical AG · 2017 · 3 generations on3descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −10 | K902859 | Servo Ventilator 300 | Siemens Life Support Systems | 1991 | 3 |
| −9 | K961687 | EVITA 4 | Drager, Inc. | 1996 | — |
| −8 | K970460 | Nellcor Puritan Bennett, 840 Ventilator System (Model 840) | Puritan Bennett Corp. | 1998 | Class I |
| −7 | K984535 | Puritan-Bennett 840 Ventilatory System with Bilevel Option | Puritan Bennett Corp. | 1998 | Class I |
| −6 | K022679 | Raphael | Hamilton Medical AG | 2003 | 3 |
| −5 | K040574 | Galileo Gold | Hamilton Medical, Inc. | 2004 | — |
| −4 | K051263 | Evitaxl with Option Smartcare | Drager Medical AG & Co. Kgaa | 2005 | 1 |
| −3 | K061090 | Galileo Gold Asv | Hamilton Medical AG | 2006 | — |
| −2 | K070513 | Hamilton-G5 | Hamilton Medical AG | 2007 | 1 |
| −1 | K081521 | Modification to Hamilton-G5 | Hamilton Medical AG | 2008 | Class I |
| 0 (this device) | K102775 | Hamilton-C2 | Hamilton Medical AG | 2011 | — |
| +1 | K112006 | Hamilton-TC1 | Hamilton Medical AG | 2012 | — |
| +2 | K120670 | Hamilton-T1 | Hamilton Medical AG | 2012 | Class I |
| +1 | K121225 | Hamilton-C2 | Hamilton Medical AG | 2012 | Class I |
| +2 | K123637 | Hamilton-C3 | Hamilton Medical AG | 2012 | — |
| +3 | K161450 | Hamilton-C3 | Hamilton Medical AG | 2017 | 2 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1984 | 2 | 0 |
| 1988 | 1 | 0 |
| 1989 | 1 | 0 |
| 1991 | 4 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1994 | 4 | 0 |
| 1995 | 2 | 0 |
| 1996 | 3 | 0 |
| 1997 | 3 | 0 |
| 1998 | 6 | 0 |
| 1999 | 4 | 0 |
| 2000 | 2 | 1 |
| 2001 | 5 | 0 |
| 2002 | 3 | 0 |
| 2003 | 3 | 1 |
| 2004 | 6 | 1 |
| 2005 | 1 | 0 |
| 2006 | 3 | 0 |
| 2008 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2012 | 5 | 2 |
| 2015 | 1 | 0 |
| 2016 | 1 | 1 |
| 2017 | 2 | 1 |
| 2019 | 1 | 1 |
| 2020 | 1 | 1 |
| 2021 | 1 | 1 |
| 2024 | 1 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
13 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new CBK device
- #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (6)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K081521 | Modification to Hamilton-G5Summary page 1…SSIFICATION Ventilator, Continuous, Facility Use NAME: (21 CFR 868.5895, Product Code: 08K) LEGALLY MARKETED HAMILTON-GS (K081521) DEVICES TO WHICH HAMILTON GALILEO (K080181) EQUIVALNCE ISDrtlger EvitaXL with Neolow option (K072412, K98… | 2008 |
| * | K080181 | Modification to Galileo GoldSummary page 1…lity Use NAME: (21 CFR 868.5895, Product Code: 08K) LEGALLY MARKETED HAMILTON-GS (K081521) DEVICES TO WHICH HAMILTON GALILEO (K080181) EQUIVALNCE ISDrtlger EvitaXL with Neolow option (K072412, K983219) EIVGCAE :I Dr~ger SAVINA (K04064… | 2008 |
| * | K072412 | Option Smartcare for Evitaxl,Smartcare Kit Capno Package, CO2 Sensor Capnosmart, Models…Summary page 1…08K) LEGALLY MARKETED HAMILTON-GS (K081521) DEVICES TO WHICH HAMILTON GALILEO (K080181) EQUIVALNCE ISDrtlger EvitaXL with Neolow option (K072412, K983219) EIVGCAE :I Dr~ger SAVINA (K040642) BEIN CLAMED:MAQUET Servo-i (K041 223) DEVICE DESCRIPT… | 2008 |
| * | K983219 | Neoflow Option Model Numbers 8413563, 8411900Summary page 1…ALLY MARKETED HAMILTON-GS (K081521) DEVICES TO WHICH HAMILTON GALILEO (K080181) EQUIVALNCE ISDrtlger EvitaXL with Neolow option (K072412, K983219) EIVGCAE :I Dr~ger SAVINA (K040642) BEIN CLAMED:MAQUET Servo-i (K041 223) DEVICE DESCRIPTION The H… | 1998 |
| * | K040642 | Lpo Option for Drager Intensive Care Ventilator SavinaSummary page 1…ES TO WHICH HAMILTON GALILEO (K080181) EQUIVALNCE ISDrtlger EvitaXL with Neolow option (K072412, K983219) EIVGCAE :I Dr~ger SAVINA (K040642) BEIN CLAMED:MAQUET Servo-i (K041 223) DEVICE DESCRIPTION The HAMILTON-C2 is an electronically controlled pneumati… | 2004 |
| * | K041223 | Modification to Servo-I Ventilator System, Model 64 87 800 E407ESummary page 1…EQUIVALNCE ISDrtlger EvitaXL with Neolow option (K072412, K983219) EIVGCAE :I Dr~ger SAVINA (K040642) BEIN CLAMED:MAQUET Servo-i (K041 223) DEVICE DESCRIPTION The HAMILTON-C2 is an electronically controlled pneumatic intensive care ventilator. Due to its compact des… | 2004 |
What descends from it? (14)
See all 142 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (13 Class I)
No FDA recall names this device.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
- K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Direct predicate · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
- K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 4 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
- K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 5 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
- K121225Hamilton-C2Hamilton Medical AG · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
- K120670Hamilton-T1Hamilton Medical AG · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
- K162832731 Series VentilatorsZOLL Medical Corporation · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
- K181216Hamilton-T1, Hamilton-C1Hamilton Medical AG · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 15 Class I recalls
- K153046Hamilton-MR1Hamilton Medical AG · 2016 · 3 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
- K201306Hamilton-C3Hamilton Medical AG · 2021 · 4 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
Nearest 12 of 13 shown.
Nearest 5 of 13; all are on the web page.
What does this lineage say?
- The line of descent spans 20 years and passes through 6 companies; a new generation was cleared every 1.4 years on median.
- 24 of the 67 ancestors have a recall of any class; 6 have a Class I recall.
- Of its descendants, K112006 (Hamilton Medical AG, 2012) has gone furthest, with 13 descendants of its own.
- 1 other device also cites its predicate K081521.
How does its category compare?
All CBK devicesCBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · CBK
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · CBK
163in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · CBK
14in 2024
Class I (81 since 2007) Class II/III not classified
Adverse event reports per year · CBK
13,991in 2025
MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 11 | 142 | 17 | 1,531 |
| 2008 | 18 | 115 | 23 | 1,933 |
| 2009 | 12 | 122 | 8 | 1,766 |
| 2010 | 14 | 150 | 9 | 1,799 |
| 2011 | 13 | 192 | 12 | 1,877 |
| 2012 | 17 | 207 | 2 (1) | 5,509 |
| 2013 | 5 | 277 | 5 (2) | 7,921 |
| 2014 | 5 | 439 | 5 (4) | 7,600 |
| 2015 | 7 | 266 | 6 (4) | 11,876 |
| 2016 | 4 | 323 | 4 (1) | 16,056 |
| 2017 | 6 | 280 | 1 | 14,451 |
| 2018 | 4 | 183 | 3 (3) | 9,849 |
| 2019 | 8 | 435 | 14 (13) | 9,100 |
| 2020 | 5 | 176 | 4 | 11,096 |
| 2021 | 6 | 267 | 21 (19) | 13,322 |
| 2022 | 0 | — | 8 (4) | 15,974 |
| 2023 | 5 | 309 | 4 (3) | 24,928 |
| 2024 | 9 | 203 | 14 (14) | 20,336 |
| 2025 | 4 | 163 | 0 | 13,991 |
| 2026 | 3 | 235 | 15 (13) | — |
What is it indicated for?
The HAMILTON-C2 ventilator is intended to provide positive pressure ventilatory support to adults, pediatrics, infants, and neonates.
Intended areas of use:
* In the intensive care ward or in the recovery room.
* During secondary transport from one hospital to another.
* During transfer of ventilated patients within the hospital.
The HAMILTON-C2 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K102775 · printed Nov 3, 2026
Cite this page
510(k) Report. “K102775: Hamilton-C2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K102775
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026