AS/3 Anaesthesia Delivery Unit (as/3 Adu)

K973985510(k) TraditionalCleared Oct 30, 1998375-day reviewClass II21 CFR 868.5160Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Engstrom, Inc. · BSZ Gas-Machine, Anesthesia

Cleared on Oct 30, 1998 after a 375-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 46 of the 401 devices in its family, counting itself, have a Class I recall. 400 later devices descend from it. It has been recalled 8 times.

Decided
Oct 30, 1998
Review
375 days
Class
Class II
Product code
BSZ
Recalls
8
Predicates
0

400 later devices descend from it, 6 citing it directly. 46 of the 401 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2009: 13 descendants cleared, 2 with a Class I recall22003: 1 descendant cleared, 1 with a Class I recall12004: 9 descendants cleared, 1 with a Class I recall12005: 13 descendants cleared2017: 19 descendants cleared, 4 with a Class I recall42006: 15 descendants cleared2007: 8 descendants cleared, 2 with a Class I recall22008: 13 descendants cleared, 2 with a Class I recall22015: 15 descendants cleared, 1 with a Class I recall12014: 17 descendants cleared, 2 with a Class I recall22010: 16 descendants cleared, 1 with a Class I recall12012: 25 descendants cleared, 6 with a Class I recall62013: 22 descendants cleared, 4 with a Class I recall42011: 23 descendants cleared, 7 with a Class I recall72016: 19 descendants cleared, 3 with a Class I recall32018: 18 descendants cleared, 1 with a Class I recall12026: 18 descendants cleared2023: 23 descendants cleared, 2 with a Class I recall22021: 19 descendants cleared, 2 with a Class I recall22019: 15 descendants cleared, 3 with a Class I recall32020: 13 descendants cleared, 2 with a Class I recall22022: 20 descendants cleared2024: 23 descendants cleared2025: 23 descendants cleared1995200020052010201520202025
  1. This deviceK973985 AS/3 Anaesthesia Delivery Unit (as/3 Adu)Datex-Engstrom, Inc. · 1998
  2. Descendant · cites it directly · 270 of its ownK040847 Modification to Vamos Anesthetic Gas MonitorDraeger Medical, Inc. · 2004
  3. Descendant · 2 generations on · 251 of its ownK041956 Emg - Essential Gas Module, Model M1013ADraeget Medical AG & Co. Kgaa · 2004
  4. Descendant · 3 generations on · 249 of its ownK050762 Philips Intellivue Patient Monitor, Models MP20, MP30, MP40, MP50…Philips Medizin Systeme Boeblingen GmbH · 2005
  5. Descendant · 4 generations on · 247 of its ownK051106 The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90…Philips Medizin Systeme Boeblingen GmbH · 2005
  6. Descendant · 4 generations on · 234 of its ownK052801 Intellivue Patient Monitors with Release C.02, Models MP20, MP30…Philips Medizin Systeme Boeblingen GmbH · 2005

Showing 5 of 400 descendants. See all descendants

Descendants cleared per year

2003: 12004: 92005: 132006: 152007: 82008: 132009: 132010: 162011: 232012: 252013: 222014: 172015: 152016: 192017: 192018: 182019: 152020: 132021: 192022: 202023: 232024: 232025: 232026: 18
2003high 25 (2012) · low 1 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 400 · See all
  • K040847Modification to Vamos Anesthetic Gas MonitorDraeger Medical, Inc. · 2004 · cites it directly270descendants
  • K041956Emg - Essential Gas Module, Model M1013ADraeget Medical AG & Co. Kgaa · 2004 · 2 generations on251descendants
  • K050762Philips Intellivue Patient Monitor, Models MP20, MP30, MP40, MP50, MP60, MP70 and MP90Philips Medizin Systeme Boeblingen GmbH · 2005 · 3 generations on249descendants
  • K051106The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue Patient MonitorsPhilips Medizin Systeme Boeblingen GmbH · 2005 · 4 generations on247descendants
  • K052801Intellivue Patient Monitors with Release C.02, Models MP20, MP30, MP40, MP50, MP60…Philips Medizin Systeme Boeblingen GmbH · 2005 · 4 generations on234descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K973985AS/3 Anaesthesia Delivery Unit (as/3 Adu)Datex-Engstrom, Inc.1998—
+1K040847Modification to Vamos Anesthetic Gas MonitorDraeger Medical, Inc.2004—
+2K041956Emg - Essential Gas Module, Model M1013ADraeget Medical AG & Co. Kgaa2004—
+3K050762Philips Intellivue Patient Monitor, Models MP20, MP30, MP40, MP50, MP60, MP70…Philips Medizin Systeme Boeblingen GmbH2005—
+4K051106The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 Intellivue…Philips Medizin Systeme Boeblingen GmbH2005—
+4K052801Intellivue Patient Monitors with Release C.02, Models MP20, MP30, MP40, MP50…Philips Medizin Systeme Boeblingen GmbH2005—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200311
200491
2005130
2006150
200782
2008132
2009132
2010161
2011237
2012256
2013224
2014172
2015151
2016193
2017194
2018181
2019153
2020132
2021192
2022200
2023232
2024230
2025230
2026180
FIG. 1 — Lineage of K973985, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

46 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (400)

See all 400

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K040847Modification to Vamos Anesthetic Gas MonitorDraeger Medical, Inc.2004270
K023366Datex-Ohmeda 7900 Ventilator Enhancements Class IDatex-Ohmeda, Inc.2003134
K032803Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda200439
K042154GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc.200436
K050676GE Detax-Ohmeda Anesthesia Delivery Unit (Adu)Datex-Ohmeda, Inc.20051
K090892GE Datex-Ohmeda S/5 Adu CarestationGE Healthcare Finland Oy20090

Has anything in its family been recalled? (46 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 30, 2008Not classifiedZ-1633-2009OtherGE Healthcare has received reports of breathing circuit or tubing misconnections of the Anesthesia Delivery Unit (ADU), that may impact patient safety.
Oct 30, 2008Not classifiedZ-1635-2009OtherGE Healthcare has received reports of breathing circuit or tubing misconnections of the Anesthesia Delivery Unit (ADU), that may impact patient safety.
Oct 30, 2008Not classifiedZ-1634-2009OtherGE Healthcare has received reports of breathing circuit or tubing misconnections of the Anesthesia Delivery Unit (ADU), that may impact patient safety.
Sep 11, 2006Not classifiedZ-0291-2007OtherCertain Compact Absorbers may have an increased resistance to gas flow due to an improperly manufactured foam filter. The increased resistance can cause an elevated pressure at the ventilator end of the inspiratory circuit, but the…
Mar 22, 2006Not classifiedZ-1196-06OtherThere exists the possibility that the ADU Fresh Gas Control unit (A-FGC1) can shut off during normal handling of the anesthesia machine and result in no anesthetic agent output.
Nov 16, 2005Not classifiedZ-0857-06OtherAlarm sound-It is possible that the audible alarms from the speaker from the Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU) machine may dampen or disappear as a result of normal handling of the anesthesia machine. Visual alarms on the…
Feb 3, 2004Not classifiedZ-0545-04OtherAnesthetic agents could be delivered at concentrations above the concentration set on the S/5 and AS/3 Anesthesia Delivery Units (ADU) having serial numbers within a specified range.
Jul 15, 2003Not classifiedZ-0937-03OtherPossible disruption of mechanical ventilation or disruption of anesthetic agent delivery.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · Cites it directly · ◆ in the 2003 descendants bar of FIG. 11 Class I recall
  • K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900…Datex-Ohmeda, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 110 Class I recalls
  • K042419Fabius GS/Fabius Tiro Anesthesia SystemDraeger Medical, Inc. · 2004 · 2 generations on · ◆ in the 2004 descendants bar of FIG. 14 Class I recalls
  • K072884Fabius MriDraeger Medical AG & Co. KG · 2008 · 3 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K133886Perseus A500Draeger Medical GmbH · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · 4 generations on · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K083526Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group · 2009 · 4 generations on · ◆ in the 2009 descendants bar of FIG. 15 Class I recalls
  • K070899Puritan Bennett Legendair XL2Nellcor Puritan Bennett · 2007 · 4 generations on · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K122445Aespire 7900, Aespire ViewDatex-Ohmeda · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K230931AtlanDrägerwerk AG Co. Kgaa · 2023 · 4 generations on · ◆ in the 2023 descendants bar of FIG. 18 Class I recalls
  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · 5 generations on · ◆ in the 2007 descendants bar of FIG. 12 Class I recalls
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls

Nearest 12 of 46 shown.

What does this lineage say?

  • Of its descendants, K040847 (Draeger Medical, Inc., 2004) has gone furthest, with 270 descendants of its own.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The AS/3™ Anaesthesia Delivery Unit (AS/3™ ADU) is an Anesthesia Gas Machine for composition of anesthetic breathing gas and manual or mechanical ventilation for patients undergoing anesthesia.

The AS/3™ ADU is indicated for all persons that require inhalation anesthesia.

The AS/3™ ADU shall be operated by qualified anesthesia personnel.

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K973985: AS/3 Anaesthesia Delivery Unit (as/3 Adu).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K973985

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026