Philips Integris Allura Flat Detector Release 1.0

K022899510(k) TraditionalCleared Nov 22, 200280-day reviewClass II21 CFR 892.1650Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America Co. · JAA System, X-Ray, Fluoroscopic, Image-Intensified

Cleared on Nov 22, 2002 as substantially equivalent to K020055 (Philips Medical Systems, 2002) after an 80-day review. Its family tree reaches back 3 generations to 1989. 10 of the 121 devices in its family, counting itself, have a Class I recall. 114 later devices descend from it. It has been recalled 8 times.

Decided
Nov 22, 2002
Review
80 days
Class
Class II
Product code
JAA
Recalls
8
Predicates
1

Its longest line of descent runs back 3 generations to K890348 (1989). 6 ancestors in all. 114 later devices descend from it, 1 citing it directly. 10 of the 121 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1999: 1 other ancestor1995: 1 other ancestor1993: 1 other ancestor2003: 2 descendants cleared2004: 1 descendant cleared2006: 1 descendant cleared2009: 3 descendants cleared2013: 6 descendants cleared, 1 with a Class I recall12010: 5 descendants cleared2008: 1 descendant cleared2017: 5 descendants cleared, 2 with a Class I recall22014: 9 descendants cleared2018: 7 descendants cleared2011: 2 descendants cleared2021: 4 descendants cleared2016: 7 descendants cleared, 1 with a Class I recall12025: 15 descendants cleared, 1 with a Class I recall12012: 6 descendants cleared2015: 2 descendants cleared2022: 6 descendants cleared2019: 2 descendants cleared, 1 with a Class I recall12020: 5 descendants cleared, 1 with a Class I recall12023: 6 descendants cleared, 1 with a Class I recall12024: 8 descendants cleared, 2 with a Class I recall22026: 11 descendants clearedK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK020055 · Philips Integris Allura 9 with FD Option · Philips Medical Systems · 2002 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 8 descendants, 1 with a Class I recall14 generations forward: 9 descendants5 generations forward: 13 descendants6 generations forward: 15 descendants7 generations forward: 12 descendants, 3 with a Class I recall38 generations forward: 10 descendants, 5 with a Class I recall59 generations forward: 10 descendants, 1 with a Class I recall110 generations forward: 8 descendants11 generations forward: 7 descendants12 generations forward: 2 descendants13 generations forward: 3 descendants14 generations forward: 6 descendants15 generations forward: 4 descendants16 generations forward: 3 descendants17 generations forward: 1 descendant K020055 · Philips Integris Allura 9 with FD Option · Philips Medical Systems · 2002 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalled0+5+10+15
  1. Root · 3 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 2 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. 1 generation backK020055 Philips Integris Allura 9 with FD OptionPhilips Medical Systems · 2002 · Recalled
  4. This deviceK022899 Philips Integris Allura Flat Detector Release 1.0Philips Medical Systems North America Co. · 2002
  5. Descendant · cites it directly · 113 of its ownK031333 Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc. · 2003 · Recalled
  6. Descendant · 2 generations on · 64 of its ownK041949 Allura Xper FD10Philips Medical Systems North America Co. · 2004 · Recalled
  7. Descendant · 2 generations on · 51 of its ownK031836 Philips Fresco Release 1Philips Medical Systems North America Co. · 2003
  8. Descendant · 3 generations on · 49 of its ownK062650 EP Navigator SoftwarePhilips Medical Systems · 2006
  9. Descendant · 4 generations on · 41 of its ownK101311 EP Navigator R3Philips Medical Systems North America Co. · 2010

Showing 5 of 114 descendants. See all descendants

Descendants cleared per year

2003: 22004: 12006: 12008: 12009: 32010: 52011: 22012: 62013: 62014: 92015: 22016: 72017: 52018: 72019: 22020: 52021: 42022: 62023: 62024: 82025: 152026: 11
2003high 15 (2025) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 114 · See all
  • K031333Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc. · 2003 · cites it directly113descendants
  • K041949Allura Xper FD10Philips Medical Systems North America Co. · 2004 · 2 generations on64descendants
  • K031836Philips Fresco Release 1Philips Medical Systems North America Co. · 2003 · 2 generations on51descendants
  • K062650EP Navigator SoftwarePhilips Medical Systems · 2006 · 3 generations on49descendants
  • K101311EP Navigator R3Philips Medical Systems North America Co. · 2010 · 4 generations on41descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.19895
−2K993037LCV+ Version 2Ge Medical Systems, Inc.20006
−1K020055Philips Integris Allura 9 with FD OptionPhilips Medical Systems20022
0 (this device)K022899Philips Integris Allura Flat Detector Release 1.0Philips Medical Systems North America Co.2002—
+1K031333Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc.200325
+2K041949Allura Xper FD10Philips Medical Systems North America Co.200448
+2K031836Philips Fresco Release 1Philips Medical Systems North America Co.2003—
+3K062650EP Navigator SoftwarePhilips Medical Systems2006—
+4K101311EP Navigator R3Philips Medical Systems North America Co.2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199510
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200320
200410
200610
200810
200930
201050
201120
201260
201361
201490
201520
201671
201752
201870
201921
202051
202140
202260
202361
202482
2025151
2026110
FIG. 1 — Lineage of K022899, 1989–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

10 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAA device

  1. #1 K253103 Nexus DRF Digital X-ray Imaging Systemscore 13.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020055Philips Integris Allura 9 with FD Option
Summary page 1
…m with FD Option manufactured by Philips Medical Systems. The Philips Integris Allura 9 system with FD Option received a 510(k) substantially cquivalent determination in K020055 on March 15, 2002. The solid state x-ray imaging device is the same product in both systems. Device description: The Philips I…
Philips Medical Systems20022

Also cleared against K020055: 1 other device →

What descends from it? (114)

See all 114

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K031333Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc.2003113

Has anything in its family been recalled? (10 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 7, 2010Not classifiedZ-1654-2011Nonconforming Material/ComponentA wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown.
Feb 4, 2010Not classifiedZ-2448-2010Software designA lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem could only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator…
Dec 10, 2008Not classifiedZ-0347-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Dec 10, 2008Not classifiedZ-0350-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Oct 23, 2008Not classifiedZ-2474-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Oct 23, 2008Not classifiedZ-2477-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Dec 14, 2006Not classifiedZ-0369-2007OtherMovement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.
Dec 14, 2006Not classifiedZ-0368-2007OtherMovement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K130842Allura Xper FD or Table Series; Allura Xper FD10 Series; Allura Xper FD20 Series; Allura Xper FD Biplane…Philips Medical Systems Nederland B.V. · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 111 Class I recalls
  • K190106Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve Catheter ; Angled Tip Guidewire…Centerline Biomedical, Inc. · 2019 · 7 generations on · ◆ in the 2019 descendants bar of FIG. 11 Class I recall
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 7 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 7 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 8 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K242133Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad); Intra-Operative Positioning System…Centerline Biomedical, Inc. · 2024 · 8 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K241243Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters); Intra-Operative Positioning…Centerline Biomedical, Inc. · 2024 · 8 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K230309Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter (SCC-1) and IOPS Reverse Curve…Centerline Biomedical, Inc. · 2023 · 8 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K243842Intra-Operative Positioning System (IOPS®)Centerline Biomedical, Inc. · 2025 · 8 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 9 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.1 years on median.
  • 4 of the 6 ancestors have a recall of any class.
  • Of its descendants, K031333 (Philips Medical Systems North America, Inc., 2003) has gone furthest, with 113 descendants of its own.
  • 1 other device also cites its predicate K020055.

How does its category compare?

All JAA devices

JAA System, X-Ray, Fluoroscopic, Image-Intensified: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAA

2in 2025

2007: 12008: 62010: 12011: 42012: 12013: 62014: 62015: 32016: 22017: 12018: 32019: 62020: 42021: 62022: 12023: 22024: 32025: 22026 (partial year): 1
2007high 6 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAA

89in 2024

2008: 18 days median2011: 48 days median2013: 129 days median2014: 276 days median2015: 197 days median2018: 30 days median2019: 49 days median2020: 39 days median2021: 104 days median2024: 89 days median
2007high 276 (2014) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAA

4in 2025

2007: 352008: 212009: 182010: 62011: 72012: 22013: 32014: 12015: 12016: 72017: 32018: 12019: 22020: 22021: 22022: 52023: 92024: 22025: 4
2007high 35 (2007) · low 1 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAA

33in 2025

2007: 9,5172008: 9,4842009: 12,0402010: 5,8032011: 14,0002012: 11,6842013: 13,7062014: 10,2652015: 2,9222016: 352017: 562018: 362019: 492020: 392021: 352022: 342023: 412024: 252025: 33
2007high 14,000 (2011) · low 25 (2024)2025

MAUDE reports naming product code JAA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—359,517
2008618219,484
20090—1812,040
20101—65,803
2011448714,000
20121—211,684
20136129313,706
20146276110,265
2015319712,922
20162—735
20171—356
2018330136
2019649249
2020439239
20216104235
20221—534
20232—941
2024389225
20252—433
20261—0—

What is it indicated for?

The Philips Integris Allura Flat Detector release 1.0 system is intended for use in cardiovascular x-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K022899: Philips Integris Allura Flat Detector Release 1.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K022899

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.