510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Philips Integris Allura Flat Detector Release 1.2

K031333510(k) TraditionalCleared May 13, 200315-day reviewClass II21 CFR 892.1600Recalls 25

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America, Inc. · IZI System, X-Ray, Angiographic

Cleared on May 13, 2003 as substantially equivalent to K022899 (Philips Medical Systems North America Co., 2002) after a 15-day review. Its family tree reaches back 4 generations to 1989. 10 of the 121 devices in its family, counting itself, have a Class I recall. 113 later devices descend from it. It has been recalled 25 times.

Decided
May 13, 2003
Review
15 days
Class
Class II
Product code
IZI
Recalls
25
Predicates
1

Its longest line of descent runs back 4 generations to K890348 (1989). 7 ancestors in all. 113 later devices descend from it, 2 citing it directly. 10 of the 121 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1999: 1 other ancestor1995: 1 other ancestor1993: 1 other ancestor2004: 1 descendant cleared2003: 1 descendant cleared2006: 1 descendant cleared2009: 3 descendants cleared2013: 6 descendants cleared, 1 with a Class I recall12010: 5 descendants cleared2008: 1 descendant cleared2017: 5 descendants cleared, 2 with a Class I recall22014: 9 descendants cleared2018: 7 descendants cleared2011: 2 descendants cleared2021: 4 descendants cleared2016: 7 descendants cleared, 1 with a Class I recall12025: 15 descendants cleared, 1 with a Class I recall12012: 6 descendants cleared2015: 2 descendants cleared2022: 6 descendants cleared2019: 2 descendants cleared, 1 with a Class I recall12020: 5 descendants cleared, 1 with a Class I recall12023: 6 descendants cleared, 1 with a Class I recall12024: 8 descendants cleared, 2 with a Class I recall22026: 11 descendants clearedK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK020055 · Philips Integris Allura 9 with FD Option · Philips Medical Systems · 2002 · recalledK022899 · Philips Integris Allura Flat Detector Release 1.0 · Philips Medical Systems North America Co. · 2002 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor4 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 8 descendants, 1 with a Class I recall13 generations forward: 9 descendants4 generations forward: 13 descendants5 generations forward: 15 descendants6 generations forward: 12 descendants, 3 with a Class I recall37 generations forward: 10 descendants, 5 with a Class I recall58 generations forward: 10 descendants, 1 with a Class I recall19 generations forward: 8 descendants10 generations forward: 7 descendants11 generations forward: 2 descendants12 generations forward: 3 descendants13 generations forward: 6 descendants14 generations forward: 4 descendants15 generations forward: 3 descendants16 generations forward: 1 descendant K022899 · Philips Integris Allura Flat Detector Release 1.0 · Philips Medical Systems North America Co. · 2002 · recalledK020055 · Philips Integris Allura 9 with FD Option · Philips Medical Systems · 2002 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalled0+5+10+15
  1. Root · 4 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 3 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. 2 generations backK020055 Philips Integris Allura 9 with FD OptionPhilips Medical Systems · 2002 · Recalled
  4. 1 generation backK022899 Philips Integris Allura Flat Detector Release 1.0Philips Medical Systems North America Co. · 2002 · Recalled
  5. This deviceK031333 Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc. · 2003
  6. Descendant · cites it directly · 64 of its ownK041949 Allura Xper FD10Philips Medical Systems North America Co. · 2004 · Recalled
  7. Descendant · cites it directly · 51 of its ownK031836 Philips Fresco Release 1Philips Medical Systems North America Co. · 2003
  8. Descendant · 2 generations on · 49 of its ownK062650 EP Navigator SoftwarePhilips Medical Systems · 2006
  9. Descendant · 3 generations on · 41 of its ownK101311 EP Navigator R3Philips Medical Systems North America Co. · 2010
  10. Descendant · 4 generations on · 39 of its ownK111245 Heartnavigator Release 1Philips Medical Systems Nederland B.V. · 2011

Showing 5 of 113 descendants. See all descendants

Descendants cleared per year

2003: 12004: 12006: 12008: 12009: 32010: 52011: 22012: 62013: 62014: 92015: 22016: 72017: 52018: 72019: 22020: 52021: 42022: 62023: 62024: 82025: 152026: 11
2003high 15 (2025) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 113 · See all
  • K041949Allura Xper FD10Philips Medical Systems North America Co. · 2004 · cites it directly64descendants
  • K031836Philips Fresco Release 1Philips Medical Systems North America Co. · 2003 · cites it directly51descendants
  • K062650EP Navigator SoftwarePhilips Medical Systems · 2006 · 2 generations on49descendants
  • K101311EP Navigator R3Philips Medical Systems North America Co. · 2010 · 3 generations on41descendants
  • K111245Heartnavigator Release 1Philips Medical Systems Nederland B.V. · 2011 · 4 generations on39descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.19895
−3K993037LCV+ Version 2Ge Medical Systems, Inc.20006
−2K020055Philips Integris Allura 9 with FD OptionPhilips Medical Systems20022
−1K022899Philips Integris Allura Flat Detector Release 1.0Philips Medical Systems North America Co.20028
0 (this device)K031333Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc.2003—
+1K041949Allura Xper FD10Philips Medical Systems North America Co.200448
+1K031836Philips Fresco Release 1Philips Medical Systems North America Co.2003—
+2K062650EP Navigator SoftwarePhilips Medical Systems2006—
+3K101311EP Navigator R3Philips Medical Systems North America Co.2010—
+4K111245Heartnavigator Release 1Philips Medical Systems Nederland B.V.2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199510
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200410
200610
200810
200930
201050
201120
201260
201361
201490
201520
201671
201752
201870
201921
202051
202140
202260
202361
202482
2025151
2026110
FIG. 1 — Lineage of K031333, 1989–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

10 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZI device

  1. #1 K230727 SPY Portable Handheld Imaging (SPY-PHI) Systemscore 23.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022899Philips Integris Allura Flat Detector Release 1.0Philips Medical Systems North America Co.20028
*—GE Innova 4100 system
Named without a submission number
—

What descends from it? (113)

See all 113

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041949Allura Xper FD10Philips Medical Systems North America Co.200464
K031836Philips Fresco Release 1Philips Medical Systems North America Co.200351

Has anything in its family been recalled? (10 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 13, 2025Class IIZ-1077-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 2, 2024Class IIZ-1151-2024Process designDisk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Feb 19, 2019Class IIZ-1140-2019Component design/selectionAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or…
Feb 19, 2019Class IIZ-1141-2019Component design/selectionAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or…
Jul 16, 2018Class IIZ-3213-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3214-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3217-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3220-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3219-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3215-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3218-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Jul 16, 2018Class IIZ-3216-2018Device DesignThere is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the…
Mar 22, 2017Class IIZ-1820-2017Under Investigation by firmDue to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to…
Mar 22, 2017Class IIZ-1822-2017Under Investigation by firmDue to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to…
Sep 24, 2014Class IIZ-1280-2016Device DesignLoss of key image functionality due to a bent pedal of the Footswitch.
Apr 2, 2012Not classifiedZ-1462-2012Process controlThe monitor Ceiling suspension may fail, which may cause the monitor to drop.
Jul 8, 2011Not classifiedZ-2976-2011Software designSystem freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.
Feb 4, 2010Not classifiedZ-2449-2010Software designA lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator…
Dec 10, 2008Not classifiedZ-0344-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Oct 23, 2008Not classifiedZ-2471-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Dec 14, 2006Not classifiedZ-0370-2007OtherMovement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.
Oct 12, 2006Not classifiedZ-0165-2007OtherPotential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
Feb 6, 2006Not classifiedZ-0581-06OtherPotential lock-up of system requiring reboot of system.
Oct 12, 2005Not classifiedZ-0206-06OtherPotential for high voltage exposure involving the injector interface connector of patient support tables.
Oct 12, 2005Not classifiedZ-0207-06OtherPotential for high voltage exposure involving the injector interface connector of patient support tables.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K130842Allura Xper FD or Table Series; Allura Xper FD10 Series; Allura Xper FD20 Series; Allura Xper FD Biplane…Philips Medical Systems Nederland B.V. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 111 Class I recalls
  • K190106Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve Catheter ; Angled Tip Guidewire…Centerline Biomedical, Inc. · 2019 · 6 generations on · ◆ in the 2019 descendants bar of FIG. 11 Class I recall
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 6 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 7 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K242133Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad); Intra-Operative Positioning System…Centerline Biomedical, Inc. · 2024 · 7 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K241243Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters); Intra-Operative Positioning…Centerline Biomedical, Inc. · 2024 · 7 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K230309Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter (SCC-1) and IOPS Reverse Curve…Centerline Biomedical, Inc. · 2023 · 7 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K243842Intra-Operative Positioning System (IOPS®)Centerline Biomedical, Inc. · 2025 · 7 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 8 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 14 years and passes through 5 companies; a new generation was cleared every 2.1 years on median.
  • 5 of the 7 ancestors have a recall of any class.
  • Of its descendants, K041949 (Philips Medical Systems North America Co., 2004) has gone furthest, with 64 descendants of its own.

How does its category compare?

All IZI devices

IZI System, X-Ray, Angiographic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IZI

2in 2025

2007: 72008: 62009: 22010: 22011: 42012: 12013: 12014: 42016: 12017: 22018: 22019: 22020: 12021: 12022: 12023: 32024: 22025: 22026 (partial year): 2
2007high 7 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZI

81in 2023

2007: 28 days median2008: 68 days median2011: 104 days median2014: 111 days median2023: 81 days median
2007high 111 (2014) · low 28 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZI

1in 2021

2007: 62008: 522009: 182010: 112011: 62012: 62013: 62014: 52015: 142016: 72017: 12018: 22019: 22020: 22021: 12026 (partial year): 1
2007high 52 (2008) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZI

2,171in 2025

2007: 922008: 1012009: 1322010: 2522011: 3772012: 772013: 582014: 1232015: 362016: 1022017: 882018: 492019: 302020: 142021: 292022: 2652023: 1,2042024: 1,3322025: 2,171
2007high 2,171 (2025) · low 14 (2020)2025

MAUDE reports naming product code IZI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007728692
200866852101
20092—18132
20102—11252
201141046377
20121—677
20131—658
201441115123
20150—1436
20161—7102
20172—188
20182—249
20192—230
20201—214
20211—129
20221—0265
202338101,204
20242—01,332
20252—02,171
20262—1—

What is it indicated for?

The Philips lntegris Allura Flat Detector release 1.2 system is intended for use in cardiovascular and vascular x-ray imaging applications, including diagnostic, intewentional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031333: Philips Integris Allura Flat Detector Release 1.2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K031333

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026