510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Modular Angiography System Axiom Artis

K021021510(k) TraditionalCleared Jun 6, 200269-day reviewClass II21 CFR 892.1650Recalls 21

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · JAA System, X-Ray, Fluoroscopic, Image-Intensified

Cleared on Jun 6, 2002 as substantially equivalent to K010721 (Siemens Medical Corp., 2001) and 1 other predicate after a 69-day review. Its family tree reaches back 2 generations to 1989. 4 of the 224 devices in its family, counting itself, have a Class I recall. 218 later devices descend from it. It has been recalled 21 times.

Decided
Jun 6, 2002
Review
69 days
Class
Class II
Product code
JAA
Recalls
21
Predicates
2

Its longest line of descent runs back 2 generations to K890348 (1989). 5 ancestors in all. 218 later devices descend from it, 5 citing it directly. 4 of the 224 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1995: 1 other ancestor1993: 1 other ancestor2008: 8 descendants cleared2003: 1 descendant cleared2010: 7 descendants cleared2007: 1 descendant cleared2006: 2 descendants cleared2005: 1 descendant cleared2015: 12 descendants cleared2013: 11 descendants cleared2009: 5 descendants cleared2011: 8 descendants cleared2018: 21 descendants cleared2012: 6 descendants cleared2014: 14 descendants cleared2016: 15 descendants cleared, 1 with a Class I recall12022: 13 descendants cleared2019: 9 descendants cleared2017: 9 descendants cleared, 2 with a Class I recall22021: 13 descendants cleared2024: 11 descendants cleared2025: 16 descendants cleared2020: 15 descendants cleared, 1 with a Class I recall12023: 9 descendants cleared2026: 11 descendants clearedK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors 1 generation forward: 5 descendants2 generations forward: 20 descendants3 generations forward: 39 descendants4 generations forward: 44 descendants5 generations forward: 35 descendants, 2 with a Class I recall26 generations forward: 24 descendants, 1 with a Class I recall17 generations forward: 19 descendants, 1 with a Class I recall18 generations forward: 8 descendants9 generations forward: 9 descendants10 generations forward: 7 descendants11 generations forward: 3 descendants12 generations forward: 2 descendants13 generations forward: 2 descendants14 generations forward: 1 descendant K993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalled0+5+10
  1. Root · 2 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 1 generation backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. This deviceK021021 Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc. · 2002
  4. Descendant · cites it directly · 78 of its ownK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  5. Descendant · cites it directly · 75 of its ownK033244 Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt…Ge Medical Systems, Inc. · 2003 · Recalled
  6. Descendant · 2 generations on · 74 of its ownK052412 Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova…GE Healthcare · 2005 · Recalled
  7. Descendant · cites it directly · 72 of its ownK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. Descendant · 2 generations on · 69 of its ownK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled

Showing 5 of 218 descendants. See all descendants

Descendants cleared per year

2003: 12005: 12006: 22007: 12008: 82009: 52010: 72011: 82012: 62013: 112014: 142015: 122016: 152017: 92018: 212019: 92020: 152021: 132022: 132023: 92024: 112025: 162026: 11
2003high 21 (2018) · low 0 (2004)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 218 · See all
  • K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · cites it directly78descendants
  • K033244Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table OptionGe Medical Systems, Inc. · 2003 · cites it directly75descendants
  • K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare · 2005 · 2 generations on74descendants
  • K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · cites it directly72descendants
  • K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · 2 generations on69descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.19895
−1K993037LCV+ Version 2Ge Medical Systems, Inc.20006
0 (this device)K021021Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc.2002—
+1K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc.200720
+1K033244Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table OptionGe Medical Systems, Inc.20038
+2K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare200550
+1K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
+2K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199510
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200510
200620
200710
200880
200950
201070
201180
201260
2013110
2014140
2015120
2016151
201792
2018210
201990
2020151
2021130
2022130
202390
2024110
2025160
2026110
FIG. 1 — Lineage of K021021, 1989–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAA device

  1. #1 K253103 Nexus DRF Digital X-ray Imaging Systemscore 13.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K010721Axiom Artis Modular Angiography System
Summary page 1
…etween Thales, Philips and Siemens. It may be used with the Siemens AXIOM Artis and the Philips Integris Allura. The AXIOM Artis was described in premarket notification K010721 and received FDA clearance on March 30, 2001. The Philips Integris Allura Flat Detector system with the Trixell Pixium 4800 was d…
Siemens Medical Corp.200126
*K993037LCV+ Version 2
Summary page 1
…s described in premarket notification (510(k) number not yet published) and received FDA clearance on March 15, 2002. INNOVA 2000 was described in premarket notification K993037 and received FDA clearance on Fcbruary 14, 2000. Information that substantiates this claim of equivalence is provided throughout…
Ge Medical Systems, Inc.20006
*—Philips Integris Allura Flat Detector system
Named without a submission number
—

Also cleared against K010721: 1 other device →

What descends from it? (218)

See all 218

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc.200778
K033244Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table OptionGe Medical Systems, Inc.200375
K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200872
K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201024
K100102CS-Series-FPOmega Medical Imaging, Inc.20108

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 25, 2018Class IIZ-2525-2018Process controlDue to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure
Jan 5, 2018Class IIZ-1150-2018Process controlCertain systems may have leaking of the plastic coolant couplings, the liquid can enter into the flat panel detector leading to a failure of the detector. It may be necessary to cancel or restart the treatment or transfer the patient to…
Sep 28, 2017Class IIZ-0711-2018Device DesignThe possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release…
Feb 19, 2016Class IIZ-1310-2016OtherPossible leak in the cooling system, fluids could potentially seep into the equipment cabinet. This leak can occur sporadically in affected systems.
Apr 26, 2013Class IIZ-2108-2013Nonconforming Material/ComponentThe firm became aware of a potential issue with Axiom Artis systems installed by a third party installer that may involve incorrect channel nuts being used.
Sep 9, 2010Not classifiedZ-0144-2011Software designSporadic system crash while Perivision is performed.
Jul 13, 2010Not classifiedZ-2233-2010OtherThe Axiom Artis MP, Artis dMP or Artis zee Multipurpose C-arm gearbox can become damaged. If the gearbox is damaged, the C-arm may tilt away when being repositioned and collide with nearby objects in its range of motion.
Apr 12, 2010Not classifiedZ-1732-2010Device Designsporadic malfunctions of the Artis Systems with Motor Controller caused by tolerances on electronic parts that might increase during the lifetime of the system.
Mar 4, 2010Not classifiedZ-1237-2010Nonconforming Material/ComponentDue to a contact problem at the low voltage power supply of the generator controller, in rare cases where a user has invoked radiation, it may occur that no radiation is released. Under certain circumstances, radiation is possible by…
Feb 16, 2010Not classifiedZ-1312-2010Device DesignCommunication problems can occur which may result in unavailability of Bypass Fluoro.
Oct 26, 2009Not classifiedZ-0484-2010Otherc-arm may tilt
May 1, 2009Not classifiedZ-1440-2009OtherImproper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.
Sep 2, 2008Not classifiedZ-0128-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Sep 2, 2008Not classifiedZ-0121-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Sep 2, 2008Not classifiedZ-0122-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Sep 2, 2008Not classifiedZ-0123-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Sep 2, 2008Not classifiedZ-0125-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Sep 2, 2008Not classifiedZ-0127-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Sep 2, 2008Not classifiedZ-0120-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Sep 2, 2008Not classifiedZ-0124-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Sep 2, 2008Not classifiedZ-0126-2009Software designIncorrect contour finding: Image may be calibrated to the wrong size catheter.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 7 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 11.0 years on median.
  • 3 of the 5 ancestors have a recall of any class.
  • Of its descendants, K062623 (Siemens Medical Solutions USA, Inc., 2007) has gone furthest, with 78 descendants of its own.
  • 1 other device also cites its predicate K010721.

How does its category compare?

All JAA devices

JAA System, X-Ray, Fluoroscopic, Image-Intensified: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAA

2in 2025

2007: 12008: 62010: 12011: 42012: 12013: 62014: 62015: 32016: 22017: 12018: 32019: 62020: 42021: 62022: 12023: 22024: 32025: 22026 (partial year): 1
2007high 6 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAA

89in 2024

2008: 18 days median2011: 48 days median2013: 129 days median2014: 276 days median2015: 197 days median2018: 30 days median2019: 49 days median2020: 39 days median2021: 104 days median2024: 89 days median
2007high 276 (2014) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAA

4in 2025

2007: 352008: 212009: 182010: 62011: 72012: 22013: 32014: 12015: 12016: 72017: 32018: 12019: 22020: 22021: 22022: 52023: 92024: 22025: 4
2007high 35 (2007) · low 1 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAA

33in 2025

2007: 9,5172008: 9,4842009: 12,0402010: 5,8032011: 14,0002012: 11,6842013: 13,7062014: 10,2652015: 2,9222016: 352017: 562018: 362019: 492020: 392021: 352022: 342023: 412024: 252025: 33
2007high 14,000 (2011) · low 25 (2024)2025

MAUDE reports naming product code JAA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—359,517
2008618219,484
20090—1812,040
20101—65,803
2011448714,000
20121—211,684
20136129313,706
20146276110,265
2015319712,922
20162—735
20171—356
2018330136
2019649249
2020439239
20216104235
20221—534
20232—941
2024389225
20252—433
20261—0—

What is it indicated for?

AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro-angiography, general angiography, operating room angiography, multipurpose angiography and radiographic/fluoroscopic procedures e.g. Gastro-intestinal imaging, Skeletal imaging ctc. AXIOM Arttis can also support the…

Read the full statement (100 words)

AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures.

Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro-angiography, general angiography, operating room angiography, multipurpose angiography and radiographic/fluoroscopic procedures e.g. Gastro-intestinal imaging, Skeletal imaging ctc.

AXIOM Arttis can also support the acquisition of position triggered imaging for spatial data synthesis.

The intended use and indications for use of the Modular Angiography System AXIOM Artis with Flat Panel Detector as described in its labeling have not changed from its predicate device AXIOM Artis with the image intensifier.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K021021: Modular Angiography System Axiom Artis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K021021

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026