510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Artis Zee and Zeego

K073290510(k) TraditionalCleared Feb 11, 200882-day reviewClass II21 CFR 892.1600Recalls 39

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IZI System, X-Ray, Angiographic

Cleared on Feb 11, 2008 as substantially equivalent to K021021 (Siemens Medical Solutions USA, Inc., 2002) and 1 other predicate after an 82-day review. Its family tree reaches back 6 generations to 1980. 4 of the 113 devices in its family, counting itself, have a Class I recall. 72 later devices descend from it. It has been recalled 39 times.

Decided
Feb 11, 2008
Review
82 days
Class
Class II
Product code
IZI
Recalls
39
Predicates
2

Its longest line of descent runs back 6 generations to K801623 (1980). 40 ancestors in all. 72 later devices descend from it, 3 citing it directly. 4 of the 113 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 3 other ancestors2004: 2 other ancestors2001: 1 other ancestor2000: 3 other ancestors1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 1 other ancestor1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2009: 2 descendants cleared2010: 1 descendant cleared2011: 2 descendants cleared2013: 7 descendants cleared2012: 1 descendant cleared2014: 7 descendants cleared2015: 2 descendants cleared2017: 4 descendants cleared, 2 with a Class I recall22018: 6 descendants cleared2021: 4 descendants cleared2016: 6 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared2026: 5 descendants cleared2025: 8 descendants cleared2020: 5 descendants cleared, 1 with a Class I recall12024: 3 descendants cleared2022: 5 descendants cleared2023: 3 descendants clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 6 other ancestors4 generations back: 11 other ancestors5 generations back: 9 other ancestors6 generations back: 4 other ancestors 1 generation forward: 3 descendants2 generations forward: 9 descendants3 generations forward: 14 descendants4 generations forward: 15 descendants5 generations forward: 13 descendants, 2 with a Class I recall26 generations forward: 11 descendants, 1 with a Class I recall17 generations forward: 3 descendants, 1 with a Class I recall18 generations forward: 2 descendants9 generations forward: 2 descendants K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−50+5
  1. Root · 6 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 5 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 4 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 3 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 2 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 1 generation backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. This deviceK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008
  8. Descendant · cites it directly · 69 of its ownK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. Descendant · 2 generations on · 33 of its ownK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · Recalled
  10. Descendant · 2 generations on · 30 of its ownK122644 Artis Zee/Zeego with CSX-10 Detector SW VC21Siemens Medical Solutions USA, Inc. · 2013 · Recalled
  11. Descendant · cites it directly · 24 of its ownK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  12. Descendant · 3 generations on · 23 of its ownK163286 ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · Recalled

Showing 5 of 72 descendants. See all descendants

Descendants cleared per year

2009: 22010: 12011: 22012: 12013: 72014: 72015: 22016: 62017: 42018: 62019: 12020: 52021: 42022: 52023: 32024: 32025: 82026: 5
2009high 8 (2025) · low 1 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 72 · See all
  • K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · cites it directly69descendants
  • K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · 2 generations on33descendants
  • K122644Artis Zee/Zeego with CSX-10 Detector SW VC21Siemens Medical Solutions USA, Inc. · 2013 · 2 generations on30descendants
  • K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · cites it directly24descendants
  • K163286ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · 3 generations on23descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−5K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−4K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−3K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−2K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−1K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
0 (this device)K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.2008—
+1K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
+2K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc.201331
+2K122644Artis Zee/Zeego with CSX-10 Detector SW VC21Siemens Medical Solutions USA, Inc.20131
+1K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201039
+3K163286ARTIS phenoSiemens Medical Solution USA, Inc.201714

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199910
200030
200110
200230
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200920
201010
201120
201210
201370
201470
201520
201661
201742
201860
201910
202051
202140
202250
202330
202430
202580
202650
FIG. 1 — Lineage of K073290, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZI device

  1. #1 K230727 SPY Portable Handheld Imaging (SPY-PHI) Systemscore 23.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K021021Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc.200221
*K052202Axiom Artis - Modular Angiographic System
Summary page 2
…510(k) Number Date of Clearance Device Name K(02 1021 June 06, 2002 MODLA ANOIOGRAPIIY SYSTEM AXIOM ARTIS K052202 Murch 07, 2006 AXIOM ARTIS MODULAR ANGIOGRAPHY SYSTEM The Indication for use of the Artis zee/zeego d…
Siemens Medical Solutions USA, Inc.200629

Also cleared against K021021: 4 other devices →

What descends from it? (72)

See all 72

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.200969
K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201024
K091166IMRXImris, Inc.20091

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 9, 2019Class IIZ-1339-2019Software designIn affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
Apr 9, 2019Class IIZ-1340-2019Software designIn affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
Oct 26, 2018Class IIZ-0558-2019Process controlThis recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the…
Oct 19, 2018Class IIZ-0457-2019Device DesignVibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
Sep 28, 2017Class IIZ-0710-2018Device DesignThe possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release…
Aug 24, 2017Class IIZ-0508-2018Employee errorWithin a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be…
Jun 6, 2017Class IIZ-0707-2018Device DesignPotential for liquid entry
Mar 28, 2017Class IIZ-1824-2017Nonconforming Material/ComponentSoftware error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system functions properly until…
Feb 27, 2017Class IIZ-1467-2017Process controlA gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. Such liquids may include disinfectants, cleaning agents and/ or bodily fluids. In rare cases, the introduction of liquids in…
Feb 3, 2017Class IIZ-2686-2017Component design/selectionFailures can result due to a component defect on Artis zee systems with an A100 Plus generator of a certain delivery lot and 2-focus Megalix Cat Plus tube unit.
Jul 13, 2016Class IIZ-2394-2016Under Investigation by firmThe possibility exists for contaminants, in the form of biomass, to develop in the cooling system of Artis systems resulting in damage to the pump system.
Jul 7, 2016Class IIZ-2344-2016Nonconforming Material/ComponentDue to an error in the 19 Live Display, image reproduction may fail in the examination room and the potential exists for the loss of images immediately after system startup.
Feb 26, 2016Class IIZ-1299-2016Device DesignIn Artis zeego systems, angulations in the vicinity of the C-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. In Artis systems with A100Plus generators, an attempt to resume operation…
Feb 19, 2016Class IIZ-1311-2016OtherPossible leak in the cooling system, fluids could potentially seep into the equipment cabinet. This leak can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-1284-2016Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can result in a cooling system failure and subsequent system failure. This leak can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3093-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3095-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3099-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3094-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3096-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3092-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3097-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3098-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Oct 30, 2015Class IIZ-0481-2016Software designDue to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have…
Jun 29, 2015Class IIZ-2340-2015Device DesignThe possibility exists for a large amount of liquid to seep through the top cover of the table and penetrate the components inside the table. If this occurs, table movements may no longer be possible. No injuries reported.
Sep 23, 2014Class IIZ-0183-2015Device DesignA cable inside the C-arm systems may have been routed in an inappropriate manner. Improper routing of the cable may result in increased wear over time. Without additional measures, the potential exists for a cable inside the C-arm to…
Sep 17, 2014Class IIZ-0854-2015Process controlThe Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed…
Jun 5, 2014Class IIZ-2587-2014Device DesignThere is a potential problem with the AXIOM Artis or Artis zee system in connection with the hand switch operated at the patient table which may result in uncontrolled release of an x-ray image.
Jun 5, 2014Class IIZ-2569-2014Software designThere is a potential problem with Artis zee and Artis zeego systems running software version VC21B and being used in conjunction with the Large Display, in that under certain circumstances, the release of radiation can become blocked…
Feb 7, 2014Class IIZ-1272-2014Device DesignCertain units of the AXIOM Artis Systems are using a flat detector cooling unit with inappropriate connectors that may possibly leak, causing a system failure and possible drop out of the gantry system.
Feb 7, 2014Class IIZ-1309-2014Device DesignAn inappropriate cable connection was assembled for certain Artis zee systems. which may cause a system failure and possibly lead to a restricted maneuverability of the gantry system.
Dec 16, 2013Class IIZ-1080-2014Device DesignDuring the lifetime of Artis zee Ceiling systems, there is an increased wear due to potential sub-optimal cable guidance that can lead to cable damage.
Sep 11, 2013Class IIZ-0486-2015Device DesignThere is a potential issue on running Artis systems running software VC1x software if a network problem arises, the function cannot be deactivated again by pressing the "Block Radiation" key on the touch screen control. If this behavior…
Jun 26, 2013Class IIZ-1782-2013Software Design ChangeSiemens issued a customer safety advisory notice and field safety corrective action for a potential issue on Artis systems with Artis VC20x/VC21A that have Patch 7 and a 3D and DSA license installed. The issue affects-DSA rotational…
Mar 8, 2013Class IIZ-0997-2013Software designThe firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may…
Jul 13, 2010Not classifiedZ-2233-2010OtherThe Axiom Artis MP, Artis dMP or Artis zee Multipurpose C-arm gearbox can become damaged. If the gearbox is damaged, the C-arm may tilt away when being repositioned and collide with nearby objects in its range of motion.
Oct 26, 2009Not classifiedZ-0485-2010Otherc-arm may tilt
May 1, 2009Not classifiedZ-1443-2009OtherImproper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.
Aug 22, 2008Not classifiedZ-0194-2009Labeling designPatients could possibly fall off the table if not properly secured when: 1) moving them onto the patient table, 2) moving them on the patient table or 3) removing them from the patient table.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 7 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 28 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 6 of the 40 ancestors have a recall of any class.
  • Of its descendants, K090745 (Siemens Medical Solutions USA, Inc., 2009) has gone furthest, with 69 descendants of its own.
  • 4 other devices also cite its predicate K021021.

How does its category compare?

All IZI devices

IZI System, X-Ray, Angiographic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · IZI

2in 2025

2007: 72008: 62009: 22010: 22011: 42012: 12013: 12014: 42016: 12017: 22018: 22019: 22020: 12021: 12022: 12023: 32024: 22025: 22026 (partial year): 2
2007high 7 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZI

81in 2023

2007: 28 days median2008: 68 days median2011: 104 days median2014: 111 days median2023: 81 days median
2007high 111 (2014) · low 28 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZI

1in 2021

2007: 62008: 522009: 182010: 112011: 62012: 62013: 62014: 52015: 142016: 72017: 12018: 22019: 22020: 22021: 12026 (partial year): 1
2007high 52 (2008) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZI

2,171in 2025

2007: 922008: 1012009: 1322010: 2522011: 3772012: 772013: 582014: 1232015: 362016: 1022017: 882018: 492019: 302020: 142021: 292022: 2652023: 1,2042024: 1,3322025: 2,171
2007high 2,171 (2025) · low 14 (2020)2025

MAUDE reports naming product code IZI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007728692
200866852101
20092—18132
20102—11252
201141046377
20121—677
20131—658
201441115123
20150—1436
20161—7102
20172—188
20182—249
20192—230
20201—214
20211—129
20221—0265
202338101,204
20242—01,332
20252—02,171
20262—1—

What is it indicated for?

Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.

Artis zee / zeego can also support the acquisition of position triggered imaging for spatial data synthesis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073290: Artis Zee and Zeego.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K073290

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026