510(k) Report

510(k) · Cleared

Recalled · 2007

Medex 3000 Series Syringe Infusion Pump

K982640510(k) TraditionalCleared Apr 6, 1999251-day reviewClass II21 CFR 880.5725Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medex, Inc. · FRN Pump, Infusion

Cleared on Apr 6, 1999 as substantially equivalent to K890120 (Medfusion, Inc., 1989) and 3 other predicates after a 251-day review. Its family tree reaches back 1 generation to 1989. 7 of the 28 devices in its family, counting itself, have a Class I recall. 23 later devices descend from it. It has been recalled once.

Decided
Apr 6, 1999
Review
251 days
Class
Class II
Product code
FRN
Recalls
1
Predicates
4

Its longest line of descent runs back 1 generation to K890120 (1989). 4 ancestors in all. 23 later devices descend from it, 2 citing it directly. 7 of the 28 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1995: 1 other ancestor1990: 1 other ancestor2002: 1 descendant cleared, 1 with a Class I recall12008: 1 descendant cleared2006: 2 descendants cleared2014: 1 descendant cleared, 1 with a Class I recall12011: 3 descendants cleared, 1 with a Class I recall12012: 3 descendants cleared2009: 1 descendant cleared2018: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall12016: 2 descendants cleared2022: 2 descendants cleared, 1 with a Class I recall12023: 1 descendant cleared, 1 with a Class I recall12024: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant cleared2013: 1 descendant cleared2015: 1 descendant clearedK890120 · Syringe Infusion Pump · Medfusion, Inc. · 19891990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 4 descendants, 2 with a Class I recall23 generations forward: 9 descendants, 3 with a Class I recall34 generations forward: 4 descendants, 1 with a Class I recall15 generations forward: 1 descendant6 generations forward: 1 descendant7 generations forward: 1 descendant8 generations forward: 1 descendant K890120 · Syringe Infusion Pump · Medfusion, Inc. · 19890+5
  1. Root · 1 generation backK890120 Syringe Infusion PumpMedfusion, Inc. · 1989
  2. This deviceK982640 Medex 3000 Series Syringe Infusion PumpMedex, Inc. · 1999
  3. Descendant · cites it directly · 21 of its ownK023264 Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Recalled (Class I)
  4. Descendant · 2 generations on · 6 of its ownK053554 CompufloMilestone Scientific, Inc. · 2006
  5. Descendant · 3 generations on · 5 of its ownK101194 Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011
  6. Descendant · 4 generations on · 4 of its ownK112067 Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011
  7. Descendant · 2 generations on · 4 of its ownK111386 Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2…Smiths Medical MD, Inc. · 2011 · Recalled (Class I)

Showing 5 of 23 descendants. See all descendants

Descendants cleared per year

2002: 12006: 22008: 12009: 12011: 32012: 32013: 12014: 12015: 12016: 22018: 12020: 12022: 22023: 12024: 12025: 1
2002high 3 (2011) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 23 · See all
  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · cites it directly21descendants
  • K053554CompufloMilestone Scientific, Inc. · 2006 · 2 generations on6descendants
  • K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011 · 3 generations on5descendants
  • K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011 · 4 generations on4descendants
  • K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety…Smiths Medical MD, Inc. · 2011 · 2 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K890120Syringe Infusion PumpMedfusion, Inc.1989—
0 (this device)K982640Medex 3000 Series Syringe Infusion PumpMedex, Inc.1999—
+1K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc.2002Class I
+2K053554CompufloMilestone Scientific, Inc.2006—
+3K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc.2011—
+4K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc.2011—
+2K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication…Smiths Medical MD, Inc.2011Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199510
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200211
200620
200810
200910
201131
201230
201310
201411
201510
201620
201810
202011
202221
202311
202411
202510
FIG. 1 — Lineage of K982640, 1989–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K890120Syringe Infusion Pump
Summary page 1
…ump, for purposes as defined under Section 510(k) of the Federal Food, Drug and Cosmetic Act, is substantially equivalent to the Medex 2000 Series Syringe Infusion Pumps K890120, K901755 and K955231 and the Baxter AutoSyringe AS50 Infusion Pump K945942 (SE decision 7/10/95). For simplicity, the Medex 2010i…
Medfusion, Inc.1989—
*K901755Model 3001 Syringe Infusion Pump
Summary page 1
…purposes as defined under Section 510(k) of the Federal Food, Drug and Cosmetic Act, is substantially equivalent to the Medex 2000 Series Syringe Infusion Pumps K890120, K901755 and K955231 and the Baxter AutoSyringe AS50 Infusion Pump K945942 (SE decision 7/10/95). For simplicity, the Medex 2010i will be…
Medfusion, Inc.1990—
*K955231Medfusion Model 2001 Syringe Infusion Pump
Summary page 1
…defined under Section 510(k) of the Federal Food, Drug and Cosmetic Act, is substantially equivalent to the Medex 2000 Series Syringe Infusion Pumps K890120, K901755 and K955231 and the Baxter AutoSyringe AS50 Infusion Pump K945942 (SE decision 7/10/95). For simplicity, the Medex 2010i will be used through…
Medex, Inc.1996—
*K945942Auto Syringe (R) AS50 Infusion Pump
Summary page 1
…and Cosmetic Act, is substantially equivalent to the Medex 2000 Series Syringe Infusion Pumps K890120, K901755 and K955231 and the Baxter AutoSyringe AS50 Infusion Pump K945942 (SE decision 7/10/95). For simplicity, the Medex 2010i will be used throughout this 510(k) to represent the Medex 2000 Series Syr…
Baxter Healthcare Corp19953

Also cleared against K890120: 1 other device →

What descends from it? (23)

See all 23

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K023264Medley Syringe Pump Module System, Model 8110 Class IAlaris Medical Systems, Inc.200221
K072603Abc Syringe Infusion Pump, Model 4100Elixir Corporation20080

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 3, 2007Not classifiedZ-0768-2007OtherSoftware anomaly that causes the pump to experience a sudden and unexpected failure. This anomaly occurs during the pump's startup sequence when the pump's Preventative Maintenance Alarm goes off, preventing the pump from completing its…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 11 Class I recall
  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 128 Class I recalls
  • K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety Software and Supported…Smiths Medical MD, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 16 Class I recalls
  • K191910SpaceStation MRIB. Braun Melsugen AG · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls
  • K211125Novum IQ Syringe PumpBaxter Healthcare Corporation · 2022 · 3 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 3 generations on · ◆ in the 2023 descendants bar of FIG. 17 Class I recalls
  • K242390Novum IQ Syringe Pump (40800BAXUS)Baxter Healthcare Corporation · 2024 · 4 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K023264 (Alaris Medical Systems, Inc., 2002) has gone furthest, with 21 descendants of its own.
  • 1 other device also cites its predicate K890120.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Medex 3000 Series Syringe Infusion Pump indications for use are as follows: ¢ in the administration of fluids requiring precisely controlled infusion rates including blood or bfood products, lipids, drugs, antibiotics and other therapeutic fluids. ¢ in the following delivery routes: arterial, epidural, intravenous, spinal, and subcutaneous. ¢ in the following delivery modes: continuous, volume/time, mass, body weight, custom dilution, intermittent…

Read the full statement (118 words)

The Medex 3000 Series Syringe Infusion Pump indications for use are as follows:

¢ in the administration of fluids requiring precisely controlled infusion rates including blood or bfood products, lipids, drugs, antibiotics and other therapeutic fluids.

¢ in the following delivery routes: arterial, epidural, intravenous, spinal, and subcutaneous.

¢ in the following delivery modes: continuous, volume/time, mass, body weight, custom dilution, intermittent and bolus.

« in critical care, anesthesia, neonatal and pediatric applications or other healthcare settings where the use of the Syringe Infusion Pump can be monitored or supervised by a clinician.

¢ the Syringe Infusion Pump is contraindicated for use on the inlet side of extracorporeal membrane oxygenation (ECMO) systems where high negative pressures may occur.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K982640: Medex 3000 Series Syringe Infusion Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K982640

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026