510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Artis Zee / Zeegos with SW VC14

K090745510(k) TraditionalCleared Jun 18, 200990-day reviewClass II21 CFR 892.1600Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IZI System, X-Ray, Angiographic

Cleared on Jun 18, 2009 as substantially equivalent to K073290 (Siemens Medical Solutions USA, Inc., 2008) after a 90-day review. Its family tree reaches back 7 generations to 1980. 4 of the 111 devices in its family, counting itself, have a Class I recall. 69 later devices descend from it. It has been recalled 7 times.

Decided
Jun 18, 2009
Review
90 days
Class
Class II
Product code
IZI
Recalls
7
Predicates
1

Its longest line of descent runs back 7 generations to K801623 (1980). 41 ancestors in all. 69 later devices descend from it, 5 citing it directly. 4 of the 111 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 3 other ancestors2004: 2 other ancestors2001: 1 other ancestor2000: 3 other ancestors1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 1 other ancestor1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2011: 1 descendant cleared2013: 7 descendants cleared2012: 1 descendant cleared2010: 1 descendant cleared2014: 7 descendants cleared2015: 2 descendants cleared2017: 4 descendants cleared, 2 with a Class I recall22018: 6 descendants cleared2021: 4 descendants cleared2016: 6 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared2026: 5 descendants cleared2020: 5 descendants cleared, 1 with a Class I recall12024: 3 descendants cleared2022: 5 descendants cleared2025: 8 descendants cleared2023: 3 descendants clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 3 other ancestors4 generations back: 6 other ancestors5 generations back: 11 other ancestors6 generations back: 9 other ancestors7 generations back: 4 other ancestors 1 generation forward: 5 descendants2 generations forward: 12 descendants3 generations forward: 17 descendants4 generations forward: 15 descendants5 generations forward: 11 descendants, 2 with a Class I recall26 generations forward: 3 descendants, 1 with a Class I recall17 generations forward: 2 descendants, 1 with a Class I recall18 generations forward: 2 descendants9 generations forward: 2 descendants K073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−50+5
  1. Root · 7 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 6 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 5 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 4 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 3 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 2 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 1 generation backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. This deviceK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009
  9. Descendant · cites it directly · 33 of its ownK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · Recalled
  10. Descendant · cites it directly · 30 of its ownK122644 Artis Zee/Zeego with CSX-10 Detector SW VC21Siemens Medical Solutions USA, Inc. · 2013 · Recalled
  11. Descendant · cites it directly · 24 of its ownK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  12. Descendant · 2 generations on · 23 of its ownK163286 ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · Recalled
  13. Descendant · 2 generations on · 18 of its ownK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · Recalled

Showing 5 of 69 descendants. See all descendants

Descendants cleared per year

2010: 12011: 12012: 12013: 72014: 72015: 22016: 62017: 42018: 62019: 12020: 52021: 42022: 52023: 32024: 32025: 82026: 5
2010high 8 (2025) · low 1 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 69 · See all
  • K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · cites it directly33descendants
  • K122644Artis Zee/Zeego with CSX-10 Detector SW VC21Siemens Medical Solutions USA, Inc. · 2013 · cites it directly30descendants
  • K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · cites it directly24descendants
  • K163286ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · 2 generations on23descendants
  • K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · 2 generations on18descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−6K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−5K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−4K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−3K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−2K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−1K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
0 (this device)K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.2009—
+1K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc.201331
+1K122644Artis Zee/Zeego with CSX-10 Detector SW VC21Siemens Medical Solutions USA, Inc.20131
+1K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201039
+2K163286ARTIS phenoSiemens Medical Solution USA, Inc.201714
+2K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V.201339

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199910
200030
200110
200230
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
201210
201370
201470
201520
201661
201742
201860
201910
202051
202140
202250
202330
202430
202580
202650
FIG. 1 — Lineage of K090745, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZI device

  1. #1 K230727 SPY Portable Handheld Imaging (SPY-PHI) Systemscore 23.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K073290Artis Zee and Zeego
Summary page 2
…ynthesis. The intended use and indications for use of the Artis zee / zeego VC 14 have minor changes from its predicate device the Artis zee / zeego VC13, K073290 4. Device Description: The Artis zee / zeego Modular Angiography System is designed as a set of components tha…
Siemens Medical Solutions USA, Inc.200839

Also cleared against K073290: 2 other devices →

What descends from it? (69)

See all 69

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc.201333
K122644Artis Zee/Zeego with CSX-10 Detector SW VC21Siemens Medical Solutions USA, Inc.201330
K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201024
K113403GE Discovery Igs Angiographic,Fluoroscopic X-Ray SystemGE Medical Systems SCS20126
K113034GE Innova Solid State Xray Imager Flurosopic X Ray SystemGE Healthcare20116

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 28, 2017Class IIZ-0710-2018Device DesignThe possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release…
Jun 26, 2013Class IIZ-1782-2013Software Design ChangeSiemens issued a customer safety advisory notice and field safety corrective action for a potential issue on Artis systems with Artis VC20x/VC21A that have Patch 7 and a 3D and DSA license installed. The issue affects-DSA rotational…
May 13, 2013Class IIZ-1945-2013Software designSiemens issued a customer safety notice about the overlay of system information after a functional check of the brakes on the Artis Zeego system.
Mar 8, 2013Class IIZ-0997-2013Software designThe firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may…
Jan 31, 2013Class IIZ-1750-2013Device DesignDuring the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail.
Jan 21, 2013Class IIZ-0172-2014Device DesignThe firm became aware of a potential issue with Artis zee systems bearing model numbers 10094135, 10094137, and 10094141 equipped with a card collimator from a specific lot where a dislocation of the scatter radiation filter inside the…
Nov 26, 2012Class IIZ-0992-2013Software designIn the course of the firm's product monitoring activities, a potential risk for patient or operators was determined during the operation of an Artis zeego system with software revision VC14, VC20 and VC21 in conjunction with a specific…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 7 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 29 years and passes through 5 companies; a new generation was cleared every 1.9 years on median.
  • 7 of the 41 ancestors have a recall of any class.
  • Of its descendants, K123529 (Siemens Medical Solutions USA, Inc., 2013) has gone furthest, with 33 descendants of its own.
  • 2 other devices also cite its predicate K073290.

How does its category compare?

All IZI devices

IZI System, X-Ray, Angiographic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · IZI

2in 2025

2007: 72008: 62009: 22010: 22011: 42012: 12013: 12014: 42016: 12017: 22018: 22019: 22020: 12021: 12022: 12023: 32024: 22025: 22026 (partial year): 2
2007high 7 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZI

81in 2023

2007: 28 days median2008: 68 days median2011: 104 days median2014: 111 days median2023: 81 days median
2007high 111 (2014) · low 28 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZI

1in 2021

2007: 62008: 522009: 182010: 112011: 62012: 62013: 62014: 52015: 142016: 72017: 12018: 22019: 22020: 22021: 12026 (partial year): 1
2007high 52 (2008) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZI

2,171in 2025

2007: 922008: 1012009: 1322010: 2522011: 3772012: 772013: 582014: 1232015: 362016: 1022017: 882018: 492019: 302020: 142021: 292022: 2652023: 1,2042024: 1,3322025: 2,171
2007high 2,171 (2025) · low 14 (2020)2025

MAUDE reports naming product code IZI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007728692
200866852101
20092—18132
20102—11252
201141046377
20121—677
20131—658
201441115123
20150—1436
20161—7102
20172—188
20182—249
20192—230
20201—214
20211—129
20221—0265
202338101,204
20242—01,332
20252—02,171
20262—1—

What is it indicated for?

Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as…

Read the full statement (129 words)

Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.

Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions.

Artis zee / zeego can also support the acquisition of position triggered imaging for spatial data synthesis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K090745: Artis Zee / Zeegos with SW VC14.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K090745

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026