510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

1.5T Signa HDXT 3.0T Signa HDXT Signa Vibrant

K121676510(k) TraditionalCleared Sep 20, 2012105-day reviewClass II21 CFR 892.1000Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Sep 20, 2012 as substantially equivalent to K052293 (Ge Medical Systems, LLC, 2005) and 1 other predicate after a 105-day review. Its family tree reaches back 10 generations to 1997. 5 of the 31 devices in its family, counting itself, have a Class I recall. 5 later devices descend from it. It has been recalled 10 times.

Decided
Sep 20, 2012
Review
105 days
Class
Class II
Product code
LNH
Recalls
10
Predicates
2

Its longest line of descent runs back 10 generations to K971684 (1997). 25 ancestors in all, the oldest K962061 (1996). 5 later devices descend from it, 2 citing it directly. 5 of the 31 devices in this family, counting this one, have a Class I recall.

  1. Root · 10 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 6 generations between
  3. 3 generations backK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
  4. 2 generations backK091536 Optima MR450WGe Medical Systems, LLC · 2009 · Recalled
  5. 1 generation backK103327 Discovery MR750W 3.0TGe Healthcare Japan Corporation · 2011 · Recalled
  6. This deviceK121676 1.5T Signa HDXT 3.0T Signa HDXT Signa VibrantGe Medical Systems, LLC · 2012
  7. Descendant · cites it directly · 2 of its ownK151212 Ziostation2Qi Imaging, LLC · 2015
  8. Descendant · cites it directly · 1 of its ownK200327 EvrySynaptive Medical, Inc. · 2020
  9. Descendant · 2 generations on · 1 of its ownK181892 Ziostation2Ziosoft, Inc. · 2018
  10. Descendant · 2 generations on · 0 of its ownK232981 Synq Software Version 1.3Synaptive Medical, Inc. · 2023
  11. Descendant · 3 generations on · 0 of its ownK200315 Ziostation2Ziosoft, Inc. · 2020
Show all 10 generations
  1. Root · 10 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 9 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  3. 8 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 7 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
  5. 6 generations backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
  6. 5 generations backK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
  7. 4 generations backK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
  8. 3 generations backK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
  9. 2 generations backK091536 Optima MR450WGe Medical Systems, LLC · 2009 · Recalled
  10. 1 generation backK103327 Discovery MR750W 3.0TGe Healthcare Japan Corporation · 2011 · Recalled
  11. This deviceK121676 1.5T Signa HDXT 3.0T Signa HDXT Signa VibrantGe Medical Systems, LLC · 2012

Class I recalls in this family

Most influential descendants

Showing 5 of 5 · See all
  • K151212Ziostation2Qi Imaging, LLC · 2015 · cites it directly2descendants
  • K200327EvrySynaptive Medical, Inc. · 2020 · cites it directly1descendant
  • K181892Ziostation2Ziosoft, Inc. · 2018 · 2 generations on1descendant
  • K232981Synq Software Version 1.3Synaptive Medical, Inc. · 2023 · 2 generations on0descendants
  • K200315Ziostation2Ziosoft, Inc. · 2020 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K971684Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc.19974
−9K032428Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc.20034
−8K041112Magnetom Espree SystemSiemens Medical Solutions USA, Inc.20044
−7K052164Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc.2005—
−6K062454Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc.2006—
−5K072237Magnetom VerioSiemens Medical Solutions USA, Inc.2007Class I
−4K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC200811
−3K083147Discovery MR450Ge Medical Systems, LLC200810
−2K091536Optima MR450WGe Medical Systems, LLC20098
−1K103327Discovery MR750W 3.0TGe Healthcare Japan Corporation20116
0 (this device)K1216761.5T Signa HDXT 3.0T Signa HDXT Signa VibrantGe Medical Systems, LLC2012—
+1K151212Ziostation2Qi Imaging, LLC2015—
+1K200327EvrySynaptive Medical, Inc.2020—
+2K181892Ziostation2Ziosoft, Inc.2018—
+2K232981Synq Software Version 1.3Synaptive Medical, Inc.2023—
+3K200315Ziostation2Ziosoft, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199810
199920
200141
200221
200311
200411
200530

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201810
202020
202310
FIG. 1 — Lineage of K121676, 1997–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052293GE 1.5T and 3.0T Signa HDX MR System
Summary page 1
…m Classification Names: Magnetic resonance diagnostic device Product Code: LNH Predicate Device(s): l.5T and 3.OT Signa HDx MR System (K052293) Discovery MR75Ow 3.OT (K103327) Device Description: The l.5T Signa HDx family and 3.OT…
Ge Medical Systems, LLC200521
*K103327Discovery MR750W 3.0T
Summary page 1
…evice Product Code: LNH Predicate Device(s): l.5T and 3.OT Signa HDx MR System (K052293) Discovery MR75Ow 3.OT (K103327) Device Description: The l.5T Signa HDx family and 3.OT Signa HDx family systems are a w…
Ge Healthcare Japan Corporation20116

Also cleared against K052293: 6 other devices →

What descends from it? (5)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K151212Ziostation2Qi Imaging, LLC20152
K200327EvrySynaptive Medical, Inc.20201

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 14, 2024Class IIZ-2463-2024Device DesignGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2…
Jun 14, 2024Class IIZ-2471-2024Device DesignGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2…
Feb 18, 2022Class IIZ-0860-2022Device DesignThere is potential for the images to be flipped left to right.
Dec 23, 2021Class IIZ-0603-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Aug 26, 2021Class IIZ-0141-2022Software designUnder certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Jun 4, 2021Class IIZ-2065-2021Software designGE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed…
Dec 23, 2019Class IIZ-0877-2020Software designIt was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an…
Nov 15, 2018Class IIZ-0693-2019Software designPossible incorrect software version loaded.
May 11, 2017Class IIZ-3028-2017Mixed-up of materials/componentsPossible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Apr 9, 2013Class IIZ-2266-2013Software designGE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products. For ClariView and Screen Save from GVTK (Gems Visualization ToolKit) Viewer, the…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · 2 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K030874GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co. · 2003 · 3 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · 3 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · 4 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 12 months on median.
  • 18 of the 25 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K151212 (Qi Imaging, LLC, 2015) has gone furthest, with 2 descendants of its own.
  • 6 other devices also cite its predicate K052293.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head…

Read the full statement (143 words)

The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the 1.5T Signa HDx family and 3.0T Signa HDx family reflects the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121676: 1.5T Signa HDXT 3.0T Signa HDXT Signa Vibrant.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K121676

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026