510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

IQon Spectral CT

K163711510(k) TraditionalCleared Apr 5, 201796-day reviewClass II21 CFR 892.1750Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Apr 5, 2017 as substantially equivalent to K133674 (Philips Healthcare (Cleveland), 2014) after a 96-day review. Its family tree reaches back 8 generations to 1993. 7 later devices descend from it. It has been recalled 2 times.

Decided
Apr 5, 2017
Review
96 days
Class
Class II
Product code
JAK
Recalls
2
Predicates
1

Its longest line of descent runs back 8 generations to K930090 (1993). 13 ancestors in all. 7 later devices descend from it, 1 citing it directly.

  1. Root · 8 generations backK930090 HelicatElscint, Inc. · 1993
  2. 7 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 6 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 5 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 4 generations backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. 3 generations backK060937 Brilliance VolumePhilips Medical Systems (Cleveland), Inc. · 2006 · Recalled
  7. 2 generations backK131773 Brilliance IctPhilips Healthcare (Cleveland) · 2013
  8. 1 generation backK133674 Philips Iqon Spectral CTPhilips Healthcare (Cleveland) · 2014 · Recalled
  9. This deviceK163711 IQon Spectral CTPhilips Medical Systems (Cleveland), Inc. · 2017
  10. Descendant · cites it directly · 6 of its ownK193454 IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · Recalled
  11. Descendant · 2 generations on · 5 of its ownK203020 Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · Recalled
  12. Descendant · 3 generations on · 3 of its ownK240844 Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd. · 2024
  13. Descendant · 4 generations on · 0 of its ownK253649 Spectral CT Verida FamilyPhilips Medical Systems Technologies, Ltd. · 2026
  14. Descendant · 4 generations on · 0 of its ownK252992 CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd. · 2026

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2020: 12021: 12022: 12024: 12025: 12026: 2
2020high 2 (2026) · low 0 (2023)2026

Most influential descendants

Showing 5 of 7 · See all
  • K193454IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · cites it directly6descendants
  • K203020Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · 2 generations on5descendants
  • K240844Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd. · 2024 · 3 generations on3descendants
  • K253649Spectral CT Verida FamilyPhilips Medical Systems Technologies, Ltd. · 2026 · 4 generations on0descendants
  • K252992CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd. · 2026 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K930090HelicatElscint, Inc.1993—
−7K961464Select/SPElscint, Inc.19963
−6K982060Volumax CT Imaging SystemElscint, Inc.1998—
−5K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−4K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
−3K060937Brilliance VolumePhilips Medical Systems (Cleveland), Inc.200661
−2K131773Brilliance IctPhilips Healthcare (Cleveland)2013—
−1K133674Philips Iqon Spectral CTPhilips Healthcare (Cleveland)20146
0 (this device)K163711IQon Spectral CTPhilips Medical Systems (Cleveland), Inc.2017—
+1K193454IQon Spectral CTPhilips Medical Systems Nederland B.V.20206
+2K203020Spectral CTPhilips Medical Systems Nederland B.V.20218
+3K240844Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd.2024—
+4K253649Spectral CT Verida FamilyPhilips Medical Systems Technologies, Ltd.2026—
+4K252992CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202110
202210
202410
202510
202620
FIG. 1 — Lineage of K163711, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133674Philips Iqon Spectral CT
Summary page 5
…System Device: Manufacturer: Philips Medical Systems (Cleveland), Inc. 510(k) Clearance: K133674 (November 21, 2014) Classification Name: Computed tomography x-ray system Classification…
Philips Healthcare (Cleveland)20146
†K153444Philips Multislice CT System with Low Dose CT Lung CancerPhilips Medical Systems (Cleveland), Inc.2016—

Also cleared against K133674: 1 other device →

What descends from it? (7)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K193454IQon Spectral CTPhilips Medical Systems Nederland B.V.20206

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 13, 2025Class IIZ-2601-2025Process controlThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Jun 24, 2019Class IIZ-2192-2019OtherIn certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 5 companies; a new generation was cleared every 2.6 years on median.
  • 6 of the 13 ancestors have a recall of any class.
  • Of its descendants, K193454 (Philips Medical Systems Nederland B.V., 2020) has gone furthest, with 6 descendants of its own.
  • 1 other device also cites its predicate K133674.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a…

Read the full statement (247 words)

The IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectrum. The two spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.

This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.

The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.

The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K163711: IQon Spectral CT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K163711

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.