Artis Q and Q.Zen- Modular Angiographic System

K123529510(k) TraditionalCleared Feb 26, 2013102-day reviewClass II21 CFR 892.1650Recalls 31

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Feb 26, 2013 as substantially equivalent to K090745 (Siemens Medical Solutions USA, Inc., 2009) after a 102-day review. Its family tree reaches back 8 generations to 1980. 33 later devices descend from it. It has been recalled 31 times.

Decided
Feb 26, 2013
Review
102 days
Class
Class II
Product code
OWB
Recalls
31
Predicates
1

Its longest line of descent runs back 8 generations to K801623 (1980). 42 ancestors in all. 33 later devices descend from it, 5 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 3 other ancestors2004: 2 other ancestors2001: 1 other ancestor2000: 3 other ancestors1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 1 other ancestor1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2018: 3 descendants cleared2017: 2 descendants cleared2014: 2 descendants cleared2013: 1 descendant cleared2015: 1 descendant cleared2019: 1 descendant cleared2026: 3 descendants cleared2020: 4 descendants cleared2024: 3 descendants cleared2022: 4 descendants cleared2016: 1 descendant cleared2021: 3 descendants cleared2025: 3 descendants cleared2023: 2 descendants clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor4 generations back: 3 other ancestors5 generations back: 6 other ancestors6 generations back: 11 other ancestors7 generations back: 9 other ancestors8 generations back: 4 other ancestors 1 generation forward: 5 descendants2 generations forward: 10 descendants3 generations forward: 9 descendants4 generations forward: 8 descendants5 generations forward: 1 descendant K090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−50+5
  1. Root · 8 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 7 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 6 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 5 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 4 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 3 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 2 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 1 generation backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. This deviceK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013
  10. Descendant · cites it directly · 23 of its ownK163286 ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · Recalled
  11. Descendant · cites it directly · 14 of its ownK141574 Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc. · 2014 · Recalled
  12. Descendant · cites it directly · 13 of its ownK181407 Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc. · 2018 · Recalled
  13. Descendant · cites it directly · 11 of its ownK133580 Artis OneSiemens Medical Solutions USA, Inc. · 2014 · Recalled
  14. Descendant · 2 generations on · 10 of its ownK190768 ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · Recalled

Showing 5 of 33 descendants. See all descendants

Descendants cleared per year

2013: 12014: 22015: 12016: 12017: 22018: 32019: 12020: 42021: 32022: 42023: 22024: 32025: 32026: 3
2013high 4 (2020) · low 1 (2013)2026

Most influential descendants

Showing 5 of 33 · See all
  • K163286ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · cites it directly23descendants
  • K141574Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc. · 2014 · cites it directly14descendants
  • K181407Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc. · 2018 · cites it directly13descendants
  • K133580Artis OneSiemens Medical Solutions USA, Inc. · 2014 · cites it directly11descendants
  • K190768ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · 2 generations on10descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−7K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−6K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−5K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−4K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−3K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−2K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
−1K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
0 (this device)K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc.2013—
+1K163286ARTIS phenoSiemens Medical Solution USA, Inc.201714
+1K141574Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc.201412
+1K181407Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc.201859
+1K133580Artis OneSiemens Medical Solutions USA, Inc.20147
+2K190768ARTIS iconoSiemens Medical Solutions USA, Inc.20194

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199910
200030
200110
200230
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201420
201510
201610
201720
201830
201910
202040
202130
202240
202320
202430
202530
202630
FIG. 1 — Lineage of K123529, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090745Artis Zee / Zeegos with SW VC14
Summary page 2
…e Device Trade Name: Artis zee and Artis zeego - Modular Angiographic System 510(k) Clearance K090745 Clearance Date June 18, 2009 Classification Name: Angiographic X-Ray System Classification…
Siemens Medical Solutions USA, Inc.20097

Also cleared against K090745: 4 other devices →

What descends from it? (33)

See all 33

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K163286ARTIS phenoSiemens Medical Solution USA, Inc.201723
K141574Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc.201414
K181407Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc.201813
K133580Artis OneSiemens Medical Solutions USA, Inc.201411
K132558Sync-RX SystemVolcano Corporation20133

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 9, 2019Class IIZ-1341-2019Software designIn affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
Apr 9, 2019Class IIZ-1342-2019Software designIn affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
Oct 19, 2018Class IIZ-0458-2019Device DesignVibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
Oct 19, 2018Class IIZ-0459-2019Device DesignVibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
Jul 29, 2018Class IIZ-3189-2018Device DesignTemperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g.…
Jun 22, 2018Class IIZ-2747-2018Nonconforming Material/ComponentPotential hardware issue which may cause thermal effects and possible damage to system parts.
May 1, 2018Class IIZ-2307-2018Device DesignAfter the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem…
May 1, 2018Class IIZ-2306-2018Device DesignAfter the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem…
May 1, 2018Class IIZ-2305-2018Device DesignAfter the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem…
May 1, 2018Class IIZ-2303-2018Device DesignAfter the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem…
May 1, 2018Class IIZ-2304-2018Device DesignAfter the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem…
Sep 28, 2017Class IIZ-0709-2018Device DesignThe possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release…
Aug 24, 2017Class IIZ-0508-2018Employee errorWithin a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be…
Jun 6, 2017Class IIZ-0706-2018Device DesignPotential for liquid entry
Feb 27, 2017Class IIZ-1466-2017Process controlA gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. Such liquids may include disinfectants, cleaning agents and/ or bodily fluids. In rare cases, the introduction of liquids in…
Feb 22, 2017Class IIZ-1534-2017OtherIn Artis systems with A100G generators, a component can be affected by aging which could result in the potential failure of a module in the high-voltage generator.
Dec 8, 2016Class IIZ-0972-2017Software in the Use EnvironmentSiemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: - In Artis Systems with A100 Plus or A100G generators, an attempt to resume operation following…
Nov 1, 2016Class IIZ-0601-2017Software in the Use EnvironmentThe possibility exists that the semi-automatic focus switch may not work as intended when pressing the footswitch multiple times.
Jul 13, 2016Class IIZ-2395-2016Under Investigation by firmThe possibility exists for contaminants, in the form of biomass, to develop in the cooling system of Artis systems resulting in damage to the pump system.
Jul 7, 2016Class IIZ-2344-2016Nonconforming Material/ComponentDue to an error in the 19 Live Display, image reproduction may fail in the examination room and the potential exists for the loss of images immediately after system startup.
Feb 18, 2016Class IIZ-3089-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3088-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-1283-2016Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can result in a cooling system failure and subsequent system failure. This leak can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3091-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Feb 18, 2016Class IIZ-3090-2017Unknown/Undetermined by firmDue to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Oct 30, 2015Class IIZ-0481-2016Software designDue to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have…
Jun 30, 2015Class IIZ-2404-2015Software designin case a system error occurs and the system enters the "Bypass Fluoro" mode while the X-ray locking function is active, the only way to exit the X-ray locking function, would be to either resolve the root cause of the system being in…
Jun 29, 2015Class IIZ-2341-2015Device DesignThe possibility exists for a large amount of liquid to seep through the top cover of the table and penetrate the components inside the table. If this occurs, table movements may no longer be possible. No injuries reported.
Mar 19, 2015Class IIZ-1445-2015Employee errorpotential issue with the Artis Q/ Q zen systems where it is possible that an electrical connection in the equipment cabinet has not been installed correctly. The potentially defective connection in the equipment cabinet will be inspected…
Dec 3, 2014Class IIZ-0972-2015Device DesignSuboptimal routing of the cable may result in increased wear over time. Without additional measures, damage of a cable may result in limited functionality and in rare cases, failure of a system. An ongoing procedure could be terminated in…
Dec 5, 2013Class IIZ-1202-2014Device DesignThere exists the possibility that during clinical operation, a breakdown of the x-ray tube could result in loss of x-ray imaging for the Artis Q and Artis Q biplane systems with a flat detector as 40 HDR (large 30x40).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 8 of the 42 ancestors have a recall of any class.
  • Of its descendants, K163286 (Siemens Medical Solution USA, Inc., 2017) has gone furthest, with 23 descendants of its own.
  • 4 other devices also cite its predicate K090745.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table…

Read the full statement (187 words)

Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Procedures that can be performed with the Artis family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.

Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions.

Artis can also support the acquisition of position triggered imaging for spatial data synthesis.

The Artis Q and Artis Q.zen include also the software option DynaCT with following IFU:

DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.

DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123529: Artis Q and Q.Zen- Modular Angiographic System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123529

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026