510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Artis zee/zeego & Artis Q/Q.zen

K181407510(k) TraditionalCleared Aug 15, 201877-day reviewClass II21 CFR 892.1650Recalls 59

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solution USA, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Aug 15, 2018 as substantially equivalent to K141574 (Siemens Medical Solutions USA, Inc., 2014) and 1 other predicate after a 77-day review. Its family tree reaches back 10 generations to 1980. 13 later devices descend from it. It has been recalled 59 times.

Decided
Aug 15, 2018
Review
77 days
Class
Class II
Product code
OWB
Recalls
59
Predicates
2

Its longest line of descent runs back 10 generations to K801623 (1980). 45 ancestors in all. 13 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2002: 3 other ancestors2004: 2 other ancestors2001: 1 other ancestor2000: 3 other ancestors1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 1 other ancestor1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2024: 2 descendants cleared2022: 3 descendants cleared2020: 3 descendants cleared2019: 1 descendant cleared2026: 2 descendants cleared2023: 2 descendants clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK123529 · Artis Q and Q.Zen- Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2013 · recalledK141574 · Artis Zee/Zeego SW VC21 · Siemens Medical Solutions USA, Inc. · 2014 · recalled19801990200020102020
  1. Root · 10 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 6 generations between
  3. 3 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  4. 2 generations backK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · Recalled
  5. 1 generation backK141574 Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc. · 2014 · Recalled
  6. This deviceK181407 Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc. · 2018
  7. Descendant · cites it directly · 10 of its ownK190768 ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · Recalled
  8. Descendant · cites it directly · 5 of its ownK201156 ARTIS pheno (VE2)Siemens Medical Solution USA, Inc. · 2020 · Recalled
  9. Descendant · 2 generations on · 4 of its ownK220432 ARTIS icono (VE21) SystemSiemens Medical Solutions USA, Inc. · 2022 · Recalled
  10. Descendant · 2 generations on · 4 of its ownK220409 ARTIS pheno (VE21) SystemSiemens Medical Solutions USA, Inc. · 2022 · Recalled
  11. Descendant · 3 generations on · 3 of its ownK230949 ARTIS pheno (VE30A)Siemens Medical Solutions USA, Inc. · 2023 · Recalled
Show all 10 generations
  1. Root · 10 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 9 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 8 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 7 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 6 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 5 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 4 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 3 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. 2 generations backK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · Recalled
  10. 1 generation backK141574 Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc. · 2014 · Recalled
  11. This deviceK181407 Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc. · 2018

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2019: 12020: 32022: 32023: 22024: 22026: 2
2019high 3 (2020) · low 0 (2021)2026

Most influential descendants

Showing 5 of 13 · See all
  • K190768ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · cites it directly10descendants
  • K201156ARTIS pheno (VE2)Siemens Medical Solution USA, Inc. · 2020 · cites it directly5descendants
  • K220432ARTIS icono (VE21) SystemSiemens Medical Solutions USA, Inc. · 2022 · 2 generations on4descendants
  • K220409ARTIS pheno (VE21) SystemSiemens Medical Solutions USA, Inc. · 2022 · 2 generations on4descendants
  • K230949ARTIS pheno (VE30A)Siemens Medical Solutions USA, Inc. · 2023 · 3 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−9K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−8K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−7K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−6K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−5K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−4K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
−3K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
−2K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc.201331
−1K141574Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc.201412
0 (this device)K181407Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc.2018—
+1K190768ARTIS iconoSiemens Medical Solutions USA, Inc.20194
+1K201156ARTIS pheno (VE2)Siemens Medical Solution USA, Inc.20206
+2K220432ARTIS icono (VE21) SystemSiemens Medical Solutions USA, Inc.20224
+2K220409ARTIS pheno (VE21) SystemSiemens Medical Solutions USA, Inc.20222
+3K230949ARTIS pheno (VE30A)Siemens Medical Solutions USA, Inc.20232

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199910
200030
200110
200230
200420
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202030
202230
202320
202420
202620
FIG. 1 — Lineage of K181407, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141574Artis Zee/Zeego SW VC21
Summary page 5
…WB, JAA, IZI 4. Legally Marketed Primary Predicate Device Trade Name: Artis zee/zeego SW VC21 510(k) Clearance K141574 Clearance Date September 05, 2014 Classification Name: Image-intensified fluoroscopic…
Siemens Medical Solutions USA, Inc.201412
*K123529Artis Q and Q.Zen- Modular Angiographic System
Summary page 5
…issues. Legally Marketed Secondary Predicate Device Trade Name: Artis Q and Artis Q.zen 510(k) Clearance K123529 Clearance Date February 26, 2013 Classification Name: Image-intensified fluoroscopic x-…
Siemens Medical Solutions USA, Inc.201331

What descends from it? (13)

See all 13

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K190768ARTIS iconoSiemens Medical Solutions USA, Inc.201910
K201156ARTIS pheno (VE2)Siemens Medical Solution USA, Inc.20205
K221516ARTIS icono (Ceiling Configuration)Siemens Medical Solutions USA, Inc.20223
K254173ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoSiemens Medical Solutions USA, Inc.20260
K233230See Factor CT3™Epica International, Inc.20240
K193293XV Lung Ventilation Analysis Software4Dx Limited20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 21, 2024Class IIZ-1598-2024Device DesignCoolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot…
Dec 16, 2022Class IIZ-1040-2023Software designIn rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an…
Aug 10, 2022Class IIZ-1645-2022Device DesignSiemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to…
May 24, 2022Class IIZ-1343-2022Component design/selectionIt may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only…
Feb 23, 2022Class IIZ-0972-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0971-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0976-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0979-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0968-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0975-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0969-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0970-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0974-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0973-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0980-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0967-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0977-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Feb 23, 2022Class IIZ-0978-2022Device DesignIf the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several…
Oct 20, 2021Class IIZ-0305-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0299-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0296-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0297-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0302-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0300-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0301-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0303-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0298-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Oct 20, 2021Class IIZ-0304-2022Software designIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Sep 24, 2021Class IIZ-0200-2022Software designSoftware error, the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4degrees in either direction. This can cause the 3D image reconstruction to be shown rotated by up to 5…
Jun 30, 2021Class IIZ-2404-2021Software designDue to a software error, the IAS (Image Acquisition System) may sporadically fail during startup/ restart, and result in delay of clinical treatment
Mar 30, 2021Class IIZ-1496-2021Equipment maintenanceDue to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the belts which may result in limited…
Mar 26, 2021Class IIZ-1495-2021Device DesignSome Artis zee ceiling systems show an increased abrasion of the cabling at the cable outlet of the inner C-arm, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where…
Nov 30, 2020Class IIZ-0656-2021Software designWhen the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
Oct 21, 2020Class IIZ-0469-2021Employee errorSome electronic components within the generator were equipped with an incorrect protection diode. As a result, protection against overvoltage may be inadequate, and the system may not function properly in the event an overvoltage…
Aug 3, 2020Class IIZ-0108-2021Software Manufacturing/Software DeploymentPotential exists for the collision supervision not to work properly following loss of the individual room configuration data settings causing danger of collision with fix mounted room equipment, walls or floor may occur. This might cause…
Apr 30, 2020Class IIZ-2302-2020Nonconforming Material/ComponentArtis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated…
Feb 14, 2020Class IIZ-1495-2020Software designA software issue could potentially cause the stand and table movements to be blocked.
Feb 14, 2020Class IIZ-1496-2020Software designA software issue could potentially cause the stand and table movements to be blocked.
Feb 14, 2020Class IIZ-1497-2020Software designA software issue could potentially cause the stand and table movements to be blocked.
Nov 14, 2019Class IIZ-0695-2020Software designIn affected Artis zee floor Interventional Fluoroscopic X-Ray Systems, the activation of a collision sensor will cause a block the movement of the floating tabletop and could result in a delay or interruption of a procedure.
Jul 30, 2019Class IIZ-2367-2019Process controlArtis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling…
Jun 12, 2019Class IIZ-2033-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2030-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2034-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2029-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2028-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2031-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2032-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 6, 2019Class IIZ-1903-2019Device DesignArtis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work…
Jun 6, 2019Class IIZ-1896-2019Device DesignArtis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 38 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
  • 11 of the 45 ancestors have a recall of any class.
  • Of its descendants, K190768 (Siemens Medical Solutions USA, Inc., 2019) has gone furthest, with 10 descendants of its own.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table…

Read the full statement (184 words)

Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Procedures that can be performed with the Artis family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.

Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions.

Artis can also support the acquisition of position triggered imaging for spatial data synthesis.

The Artis systems include also the software option DynaCT with following IFU:

DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.

DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K181407: Artis zee/zeego & Artis Q/Q.zen.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K181407

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026