510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Ziehm Vision RFD 3D

K231701510(k) SpecialCleared Jul 10, 202328-day reviewClass II21 CFR 892.1650Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ziehm Imaging GmbH · OWB Interventional Fluoroscopic X-Ray System

Cleared on Jul 10, 2023 as substantially equivalent to K202360 (Ziehm Imaging GmbH, 2020) after a 28-day review. Its family tree reaches back 8 generations to 2000. 3 of the 19 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled once.

Decided
Jul 10, 2023
Review
28 days
Class
Class II
Product code
OWB
Recalls
1
Predicates
1

Its longest line of descent runs back 8 generations to K003266 (2000). 15 ancestors in all. 3 later devices descend from it, 1 citing it directly. 3 of the 19 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 2 other ancestors2001: 3 other ancestors, 1 with a Class I recall12003: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor, 1 with a Class I recall12025: 1 descendant cleared2026: 2 descendants clearedK003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 2000K032280 · Siremobil Iso-C 3D · Siemens AG · 2003K061310 · Ziehm Vario 3D Digital Mobile Imaging System · Ziehm Imaging, Inc. · 2006K073346 · Ziehm Vision 2 · Ziehm Imaging, Inc. · 2008 · recalledK083545 · Ziehm Vision RFD · Ziehm Imaging, Inc. · 2009 · recalledK132904 · Ziehm Vision RFD · Ziehm Imaging GmbH · 2013 · recalledK142740 · Ziehm Vision RFD 3D · Ziehm Imaging GmbH · 2015 · recalledK202360 · Ziehm Vision RFD 3D · Ziehm Imaging GmbH · 2020200020052010201520202025
  1. Root · 8 generations backK003266 Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc. · 2000
  2. 7 generations backK032280 Siremobil Iso-C 3DSiemens AG · 2003
  3. 6 generations backK061310 Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc. · 2006
  4. 5 generations backK073346 Ziehm Vision 2Ziehm Imaging, Inc. · 2008 · Recalled
  5. 4 generations backK083545 Ziehm Vision RFDZiehm Imaging, Inc. · 2009 · Recalled
  6. 3 generations backK132904 Ziehm Vision RFDZiehm Imaging GmbH · 2013 · Recalled
  7. 2 generations backK142740 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2015 · Recalled
  8. 1 generation backK202360 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2020
  9. This deviceK231701 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2023
  10. Descendant · cites it directly · 2 of its ownK243303 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2025
  11. Descendant · 2 generations on · 0 of its ownK253296 Digital C-ARM with 3DAllengers Medical Systems Limited · 2026
  12. Descendant · 2 generations on · 0 of its ownK252735 C-beamPulmera, Inc. · 2026

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K243303Ziehm Vision RFD 3DZiehm Imaging GmbH · 2025 · cites it directly2descendants
  • K253296Digital C-ARM with 3DAllengers Medical Systems Limited · 2026 · 2 generations on0descendants
  • K252735C-beamPulmera, Inc. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K003266Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc.2000—
−7K032280Siremobil Iso-C 3DSiemens AG2003—
−6K061310Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc.2006—
−5K073346Ziehm Vision 2Ziehm Imaging, Inc.20085
−4K083545Ziehm Vision RFDZiehm Imaging, Inc.20093
−3K132904Ziehm Vision RFDZiehm Imaging GmbH20131
−2K142740Ziehm Vision RFD 3DZiehm Imaging GmbH20151
−1K202360Ziehm Vision RFD 3DZiehm Imaging GmbH2020—
0 (this device)K231701Ziehm Vision RFD 3DZiehm Imaging GmbH2023—
+1K243303Ziehm Vision RFD 3DZiehm Imaging GmbH2025—
+2K253296Digital C-ARM with 3DAllengers Medical Systems Limited2026—
+2K252735C-beamPulmera, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200131
200211
200311
200620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
202620
FIG. 1 — Lineage of K231701, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K202360Ziehm Vision RFD 3D
Summary page 4
…ber 892.1650 Product Code owB I Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K202360 Ziehm Vision RFD 3D OwB Device Description Summary 21 CFR 807.92(a)(4) The device Ziehm Vision RFD 3D is a medical Fluoroscopic…
Ziehm Imaging GmbH2020—

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K243303Ziehm Vision RFD 3DZiehm Imaging GmbH20252

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 8, 2024Class IIZ-0579-2025Radiation Control for Health and Safety ActZiehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003 · 6 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · 6 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 6 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 2.8 years on median.
  • 9 of the 15 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K243303 (Ziehm Imaging GmbH, 2025) has gone furthest, with 2 descendants of its own.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Ziehm Vision RFD 3D system is intended for use in providing both 2D and 3D pulsed and continuous fluoroscopic medical imaging for adult and pediatric populations. The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine loops during diagnostic interventional and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher contrast density are…

Read the full statement (256 words)

The Ziehm Vision RFD 3D system is intended for use in providing both 2D and 3D pulsed and

continuous fluoroscopic medical imaging for adult and pediatric populations.

The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine

loops during diagnostic interventional and surgical procedures where intra-operative imaging and

visualization of complex anatomical structures of both lower and higher contrast density are

required. Such procedures may include but are not limited to those of interventional cardiology, heart

surgery, hybrid procedures, interventional radiology, interventional angiography, electrophysiology,

pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurology,

neurosurgery, critical care, emergency room procedures, and those procedures visualizing

structures of the cervical, thoracic, and lumbar regions of the spine and joint fractures of the upper

and lower extremities, and where digital image data is required for Computer-Assisted Surgery

procedures.

The device is also intended to provide 3D medical imaging of patients during orthopedic,

neurological, intra-operative surgical procedures and where the clinician benefits from 3D

visualization of complex anatomical structures, such as but not limited to those of high contrast

objects, bones, joints, maxillofacial, cervical, thoracic, and lumbar regions of the spine, pelvis,

acetabulum and joint fractures of the upper and lower extremities, and where digital image and C-

arm positioning data is required for Computer-Assisted Surgery procedures.

The visualization of such anatomical structures assists the clinician in the clinical outcome. This

device does not support direct radiographic film exposures and is not intended for use in performing

mammography. The system is not intended for use near MRI systems.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K231701: Ziehm Vision RFD 3D.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K231701

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026