510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Alluraclarity Xper FD Series X-Ray System

K130638510(k) TraditionalCleared Jun 28, 2013109-day reviewClass II21 CFR 892.1650Recalls 39

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Nederland B.V. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Jun 28, 2013 as substantially equivalent to K102005 (Philips Medical Systems North America, Inc., 2010) and 2 other predicates after a 109-day review. Its family tree reaches back 9 generations to 1980. 5 of the 69 devices in its family, counting itself, have a Class I recall. 18 later devices descend from it. It has been recalled 39 times.

Decided
Jun 28, 2013
Review
109 days
Class
Class II
Product code
OWB
Recalls
39
Predicates
3

Its longest line of descent runs back 9 generations to K801623 (1980). 50 ancestors in all. 18 later devices descend from it, 2 citing it directly. 5 of the 69 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 3 other ancestors2003: 2 other ancestors, 1 with a Class I recall12002: 5 other ancestors2000: 4 other ancestors2001: 1 other ancestor1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 2 other ancestors1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2014: 4 descendants cleared2025: 3 descendants cleared2016: 4 descendants cleared, 1 with a Class I recall12017: 2 descendants cleared, 2 with a Class I recall22022: 1 descendant cleared2018: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall12026: 2 descendants clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK102005 · Allura Xper or Table Series · Philips Medical Systems North America, Inc. · 2010 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 3 other ancestors3 generations back: 3 other ancestors4 generations back: 5 other ancestors5 generations back: 4 other ancestors6 generations back: 11 other ancestors7 generations back: 9 other ancestors8 generations back: 4 other ancestors 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 3 descendants4 generations forward: 2 descendants, 2 with a Class I recall25 generations forward: 2 descendants, 1 with a Class I recall16 generations forward: 2 descendants, 1 with a Class I recall17 generations forward: 2 descendants8 generations forward: 2 descendants K102005 · Allura Xper or Table Series · Philips Medical Systems North America, Inc. · 2010 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−50+5
  1. Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 5 generations between
  3. 3 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  4. 2 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  5. 1 generation backK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  6. This deviceK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013
  7. Descendant · cites it directly · 15 of its ownK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
  8. Descendant · 2 generations on · 13 of its ownK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · Recalled
  9. Descendant · 3 generations on · 8 of its ownK162148 Allura Xper R9Philips Medical Systems Nethrlands BV · 2016 · Recalled
  10. Descendant · 4 generations on · 7 of its ownK163715 Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · Recalled (Class I)
  11. Descendant · 5 generations on · 6 of its ownK172822 Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · Recalled (Class I)
Show all 9 generations
  1. Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 8 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 7 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 6 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 5 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 4 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 3 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 2 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. 1 generation backK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  10. This deviceK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2014: 42016: 42017: 22018: 12020: 12022: 12025: 32026: 2
2014high 4 (2014) · low 0 (2015)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 18 · See all
  • K133292Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · cites it directly15descendants
  • K141979Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · 2 generations on13descendants
  • K162148Allura Xper R9Philips Medical Systems Nethrlands BV · 2016 · 3 generations on8descendants
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 4 generations on7descendants
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 5 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−8K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−7K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−6K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−5K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−4K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−3K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
−2K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
−1K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201039
0 (this device)K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V.2013—
+1K133292Allura Xper FD; or TablePhilips Medical Systems Nederland B.V.201445
+2K141979Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV201481
+3K162148Allura Xper R9Philips Medical Systems Nethrlands BV20163
+4K163715Allura Xper R9Philips Medical Systems Nederland B.V.2017Class I
+5K172822Azurion series R1.2Philips Medical Systems Nederland B.V.2017Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199920
200040
200110
200250
200321
200430

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201440
201641
201722
201810
202011
202210
202530
202620
FIG. 1 — Lineage of K130638, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.2 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K102005Allura Xper or Table Series
Summary page 1
…R 892.1650 Regulation Name: Image Intensified fluoroscopic x-ray system Regulatory Class: Class 1 Product Code: OWB Predicate Devices: Allura Xper FD OR Tables Series (K102005) Allura Xper FD10 (K041949) Allura Xper FD20 (K033737) Device Description: The AlluraClarity Xper FD Series X-Ray System (Allur…
Philips Medical Systems North America, Inc.201039
*K041949Allura Xper FD10
Summary page 1
…Image Intensified fluoroscopic x-ray system Regulatory Class: Class 1 Product Code: OWB Predicate Devices: Allura Xper FD OR Tables Series (K102005) Allura Xper FD10 (K041949) Allura Xper FD20 (K033737) Device Description: The AlluraClarity Xper FD Series X-Ray System (AlluraClarity system) is a modul…
Philips Medical Systems North America Co.200448
*K033737Allura Xper FD20
Summary page 1
…opic x-ray system Regulatory Class: Class 1 Product Code: OWB Predicate Devices: Allura Xper FD OR Tables Series (K102005) Allura Xper FD10 (K041949) Allura Xper FD20 (K033737) Device Description: The AlluraClarity Xper FD Series X-Ray System (AlluraClarity system) is a modular angiographic X-ray syste…
Philips Medical Systems North America, Inc.2003Class I

Also cleared against K102005: 3 other devices →

What descends from it? (18)

See all 18

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133292Allura Xper FD; or TablePhilips Medical Systems Nederland B.V.201415
K140138Heartnavigator Release 2.0Philips Medical Systems Nederland B.V.20142

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 17, 2026Class IIZ-2891-2026Under Investigation by firmPotential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Jun 17, 2026Class IIZ-2893-2026Under Investigation by firmPotential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
May 18, 2026Class IIZ-2412-2026Radiation Control for Health and Safety ActPhilips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
Apr 30, 2026Class IIZ-2259-2026Under Investigation by firmIt was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the…
Apr 30, 2026Class IIZ-2258-2026Under Investigation by firmIt was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the…
Dec 15, 2025Class IIZ-1062-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1063-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1065-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1066-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1067-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1071-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1072-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1074-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1077-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1076-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1064-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1075-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1069-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1070-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1073-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1068-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Sep 3, 2025Class IIZ-2624-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2634-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2626-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2625-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2629-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2635-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2628-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2633-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2623-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2630-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2632-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2631-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2627-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Aug 1, 2025Class IIZ-2504-2025Under Investigation by firmPotential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
Jan 2, 2024Class IIZ-1156-2024Process designDual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Jan 2, 2024Class IIZ-1144-2024Process designFramegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may…
Jan 2, 2024Class IIZ-1151-2024Process designDisk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Jun 12, 2023Class IIZ-2283-2023Component design/selectionThe bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K033737Allura Xper FD20Philips Medical Systems North America, Inc. · 2003 · Direct predicate · ◆ in the 2003 ancestors bar of FIG. 113 Class I recalls
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 4 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 4 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 6 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 33 years and passes through 7 companies; a new generation was cleared every 1.9 years on median.
  • 16 of the 50 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K133292 (Philips Medical Systems Nederland B.V., 2014) has gone furthest, with 15 descendants of its own.
  • 3 other devices also cite its predicate K102005.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The AlluraClarity Xper FD Series X-ray System is intended for use on human patients to perform: - Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. - Cardiac imaging applications including diagnostics, interventional and…

Read the full statement (105 words)

The AlluraClarity Xper FD Series X-ray System is intended for use on human patients to perform:

- Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis.

- Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP).

- Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

Additionally:

- The AlluraClarity Xper FD Series X-ray System is compatible with a hybrid Operating Room.

- FDI10 is compatible with specified magnetic navigation systems.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K130638: Alluraclarity Xper FD Series X-Ray System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130638

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history