510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Allura Xper FD Series / Allura Xper or Table Series

K141979510(k) SpecialCleared Aug 19, 201429-day reviewClass II21 CFR 892.1650Recalls 81

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Netherland BV · OWB Interventional Fluoroscopic X-Ray System

Cleared on Aug 19, 2014 as substantially equivalent to K133292 (Philips Medical Systems Nederland B.V., 2014) after a 29-day review. Its family tree reaches back 11 generations to 1980. 5 of the 66 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 81 times.

Decided
Aug 19, 2014
Review
29 days
Class
Class II
Product code
OWB
Recalls
81
Predicates
1

Its longest line of descent runs back 11 generations to K801623 (1980). 52 ancestors in all. 13 later devices descend from it, 3 citing it directly. 5 of the 66 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 3 other ancestors2003: 2 other ancestors, 1 with a Class I recall12002: 5 other ancestors2000: 4 other ancestors2001: 1 other ancestor1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 2 other ancestors1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2016: 4 descendants cleared, 1 with a Class I recall12017: 2 descendants cleared, 2 with a Class I recall22022: 1 descendant cleared2018: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall12025: 2 descendants cleared2026: 2 descendants clearedReference device: K133819 · Veradius · Philips Medical Systems Nederland B.V. · 2014K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK102005 · Allura Xper or Table Series · Philips Medical Systems North America, Inc. · 2010 · recalledK130638 · Alluraclarity Xper FD Series X-Ray System · Philips Medical Systems Nederland B.V. · 2013 · recalledK133292 · Allura Xper FD; or Table · Philips Medical Systems Nederland B.V. · 2014 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 4 other ancestors, 1 with a Class I recall14 generations back: 3 other ancestors5 generations back: 5 other ancestors6 generations back: 4 other ancestors7 generations back: 12 other ancestors8 generations back: 9 other ancestors9 generations back: 4 other ancestors 1 generation forward: 3 descendants2 generations forward: 2 descendants, 2 with a Class I recall23 generations forward: 2 descendants, 1 with a Class I recall14 generations forward: 2 descendants, 1 with a Class I recall15 generations forward: 2 descendants6 generations forward: 2 descendants K133292 · Allura Xper FD; or Table · Philips Medical Systems Nederland B.V. · 2014 · recalledK130638 · Alluraclarity Xper FD Series X-Ray System · Philips Medical Systems Nederland B.V. · 2013 · recalledK102005 · Allura Xper or Table Series · Philips Medical Systems North America, Inc. · 2010 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−10−50+5
  1. Root · 11 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 7 generations between
  3. 3 generations backK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  4. 2 generations backK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · Recalled
  5. 1 generation backK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
  6. This deviceK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014
  7. Descendant · cites it directly · 8 of its ownK162148 Allura Xper R9Philips Medical Systems Nethrlands BV · 2016 · Recalled
  8. Descendant · 2 generations on · 7 of its ownK163715 Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · Recalled (Class I)
  9. Descendant · 3 generations on · 6 of its ownK172822 Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · Recalled (Class I)
  10. Descendant · 4 generations on · 5 of its ownK181830 Azurion R2.0Philips Medical Systems Nederland B.V. · 2018 · Recalled
  11. Descendant · 4 generations on · 4 of its ownK200917 Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · Recalled (Class I)
Show all 11 generations
  1. Root · 11 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 10 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 9 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 8 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 7 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 6 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 5 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 4 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. 3 generations backK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  10. 2 generations backK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · Recalled
  11. 1 generation backK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
  12. This deviceK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2016: 42017: 22018: 12020: 12022: 12025: 22026: 2
2016high 4 (2016) · low 0 (2019)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K162148Allura Xper R9Philips Medical Systems Nethrlands BV · 2016 · cites it directly8descendants
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 2 generations on7descendants
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 3 generations on6descendants
  • K181830Azurion R2.0Philips Medical Systems Nederland B.V. · 2018 · 4 generations on5descendants
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 4 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−10K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−9K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−8K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−7K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−6K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−5K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
−4K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
−3K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201039
−2K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V.201339
−1K133292Allura Xper FD; or TablePhilips Medical Systems Nederland B.V.201445
0 (this device)K141979Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV2014—
+1K162148Allura Xper R9Philips Medical Systems Nethrlands BV20163
+2K163715Allura Xper R9Philips Medical Systems Nederland B.V.2017Class I
+3K172822Azurion series R1.2Philips Medical Systems Nederland B.V.2017Class I
+4K181830Azurion R2.0Philips Medical Systems Nederland B.V.201842
+4K200917Azurion R2.1Philips Medical Systems Nederland B.V.2020Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199920
200040
200110
200250
200321
200430

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201641
201722
201810
202011
202210
202520
202620
FIG. 1 — Lineage of K141979, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133292Allura Xper FD; or Table
Summary page 4
…rer Philips Medical Systems Nederland B.V. 510(k) number K133292 Classification Regulation 21 CFR 892.1650 Device Class: II…
Philips Medical Systems Nederland B.V.201445
†K133819VeradiusPhilips Medical Systems Nederland B.V.2014—

Also cleared against K133292: 1 other device →

What descends from it? (13)

See all 13

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162148Allura Xper R9Philips Medical Systems Nethrlands BV20168
K161563Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V.20162
K1618392D Quantitative AnalysisPhilips Medical Systems Nederland B.V.20160

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 29, 2026Class IIZ-2928-2026Software designFirmware issue potentially leading to loss of geometry movements.
Jun 17, 2026Class IIZ-2891-2026Under Investigation by firmPotential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Jun 17, 2026Class IIZ-2893-2026Under Investigation by firmPotential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
May 18, 2026Class IIZ-2412-2026Radiation Control for Health and Safety ActPhilips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
Apr 30, 2026Class IIZ-2258-2026Under Investigation by firmIt was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the…
Apr 30, 2026Class IIZ-2259-2026Under Investigation by firmIt was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the…
Dec 15, 2025Class IIZ-1062-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1071-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1075-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1064-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1076-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1070-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1073-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1066-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1063-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1068-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1067-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1069-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1072-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1074-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1065-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Dec 15, 2025Class IIZ-1077-2026OtherPhilips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not…
Sep 3, 2025Class IIZ-2624-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2634-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2626-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2629-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2625-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2635-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2628-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2627-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2633-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2623-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2631-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2632-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Sep 3, 2025Class IIZ-2630-2025Device DesignSystems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become…
Aug 1, 2025Class IIZ-2504-2025Under Investigation by firmPotential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
Aug 1, 2025Class IIZ-2503-2025Under Investigation by firmPhilips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Jun 30, 2025Class IIZ-2189-2025Software designPotential for temporary loss of imaging (X-ray) functionality due to software issue.
Jan 13, 2025Class IIZ-1080-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Aug 22, 2024Class IIZ-3166-2024Device DesignPotential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Jan 2, 2024Class IIZ-1151-2024Process designDisk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Jan 2, 2024Class IIZ-1144-2024Process designFramegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may…
Jan 2, 2024Class IIZ-1156-2024Process designDual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Oct 4, 2023Class IIZ-0228-2024Device DesignPhilips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.
Feb 2, 2022Class IIZ-0874-2022Under Investigation by firmDue to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
Nov 24, 2021Class IIZ-0724-2022Process controlDue to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to…
Nov 24, 2021Class IIZ-0725-2022Process controlDue to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to…
Nov 19, 2021Class IIZ-0416-2022OtherSet screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure…
Aug 13, 2021Class IIZ-2475-2021Device DesignEarly life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU.…
Aug 13, 2021Class IIZ-2474-2021Device DesignEarly life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU.…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K033737Allura Xper FD20Philips Medical Systems North America, Inc. · 2003 · 3 generations back · ◆ in the 2003 ancestors bar of FIG. 113 Class I recalls
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 4 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 34 years and passes through 8 companies; a new generation was cleared every 1.5 years on median.
  • 18 of the 52 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K162148 (Philips Medical Systems Nethrlands BV, 2016) has gone furthest, with 8 descendants of its own.
  • 1 other device also cites its predicate K133292.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform: • Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. • Cardiac imaging…

Read the full statement (168 words)

The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform:

• Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis.

• Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP).

• Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

Additionally:

• The Allura Xper and Allura Xper OR Table series is compatible with a hybrid Operating Room.

• Allura Xper FD10 is compatible with specified magnetic navigation systems.

• Combined with a qualified, compatible OR table, the Allura Xper OR Table series can be used for imaging in the Hybrid OR within the applications domains Neuro, Vascular, Non Vascular and Cardiac. The OR table can also be used standalone for surgical use in the OR.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141979: Allura Xper FD Series / Allura Xper or Table Series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141979

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026