510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

ARTIS icono (VE21) System

K220432510(k) TraditionalCleared Jun 29, 2022134-day reviewClass II21 CFR 892.1650Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Jun 29, 2022 as substantially equivalent to K190768 (Siemens Medical Solutions USA, Inc., 2019) after a 134-day review. Its family tree reaches back 12 generations to 1980. 4 later devices descend from it. It has been recalled 4 times.

Decided
Jun 29, 2022
Review
134 days
Class
Class II
Product code
OWB
Recalls
4
Predicates
1

Its longest line of descent runs back 12 generations to K801623 (1980). 55 ancestors in all. 4 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2017: 1 other ancestor2014: 1 other ancestor2016: 2 other ancestors2013: 1 other ancestor2004: 4 other ancestors2015: 1 other ancestor2001: 2 other ancestors2002: 3 other ancestors2000: 3 other ancestors1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 1 other ancestor1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2023: 1 descendant cleared2024: 1 descendant cleared2026: 2 descendants clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK123529 · Artis Q and Q.Zen- Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2013 · recalledK141574 · Artis Zee/Zeego SW VC21 · Siemens Medical Solutions USA, Inc. · 2014 · recalledK181407 · Artis zee/zeego & Artis Q/Q.zen · Siemens Medical Solution USA, Inc. · 2018 · recalledK190768 · ARTIS icono · Siemens Medical Solutions USA, Inc. · 2019 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 3 other ancestors4 generations back: 2 other ancestors5 generations back: 2 other ancestors6 generations back: 1 other ancestor7 generations back: 3 other ancestors8 generations back: 6 other ancestors9 generations back: 11 other ancestors10 generations back: 9 other ancestors11 generations back: 4 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 2 descendants K190768 · ARTIS icono · Siemens Medical Solutions USA, Inc. · 2019 · recalledK181407 · Artis zee/zeego & Artis Q/Q.zen · Siemens Medical Solution USA, Inc. · 2018 · recalledK141574 · Artis Zee/Zeego SW VC21 · Siemens Medical Solutions USA, Inc. · 2014 · recalledK123529 · Artis Q and Q.Zen- Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2013 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−10−50
  1. Root · 12 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 8 generations between
  3. 3 generations backK141574 Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc. · 2014 · Recalled
  4. 2 generations backK181407 Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc. · 2018 · Recalled
  5. 1 generation backK190768 ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · Recalled
  6. This deviceK220432 ARTIS icono (VE21) SystemSiemens Medical Solutions USA, Inc. · 2022
  7. Descendant · cites it directly · 3 of its ownK230950 ARTIS icono (VE30A)Siemens Medical Solutions USA, Inc. · 2023 · Recalled
  8. Descendant · 2 generations on · 2 of its ownK241572 ARTIS icono (VE40A) (floor); ARTIS icono (VE40A) (biplane); ARTIS…Siemens Medical Solutions USA, Inc. · 2024 · Recalled
  9. Descendant · 3 generations on · 0 of its ownK254173 ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS…Siemens Medical Solutions USA, Inc. · 2026
  10. Descendant · 3 generations on · 0 of its ownK253752 ARTIS genio floor; ARTIS icono.explore floorSiemens Medical Solutions USA, Inc. · 2026
Show all 12 generations
  1. Root · 12 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 11 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 10 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 9 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 8 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 7 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 6 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 5 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. 4 generations backK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · Recalled
  10. 3 generations backK141574 Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc. · 2014 · Recalled
  11. 2 generations backK181407 Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc. · 2018 · Recalled
  12. 1 generation backK190768 ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · Recalled
  13. This deviceK220432 ARTIS icono (VE21) SystemSiemens Medical Solutions USA, Inc. · 2022

Most influential descendants

Showing 4 of 4 · See all
  • K230950ARTIS icono (VE30A)Siemens Medical Solutions USA, Inc. · 2023 · cites it directly3descendants
  • K241572ARTIS icono (VE40A) (floor); ARTIS icono (VE40A) (biplane); ARTIS icono (VE40A)…Siemens Medical Solutions USA, Inc. · 2024 · 2 generations on2descendants
  • K254173ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoSiemens Medical Solutions USA, Inc. · 2026 · 3 generations on0descendants
  • K253752ARTIS genio floor; ARTIS icono.explore floorSiemens Medical Solutions USA, Inc. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−11K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−10K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−9K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−8K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−7K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−6K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
−5K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
−4K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc.201331
−3K141574Artis Zee/Zeego SW VC21Siemens Medical Solutions USA, Inc.201412
−2K181407Artis zee/zeego & Artis Q/Q.zenSiemens Medical Solution USA, Inc.201859
−1K190768ARTIS iconoSiemens Medical Solutions USA, Inc.20194
0 (this device)K220432ARTIS icono (VE21) SystemSiemens Medical Solutions USA, Inc.2022—
+1K230950ARTIS icono (VE30A)Siemens Medical Solutions USA, Inc.20231
+2K241572ARTIS icono (VE40A) (floor); ARTIS icono (VE40A) (biplane); ARTIS icono…Siemens Medical Solutions USA, Inc.202411
+3K254173ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoSiemens Medical Solutions USA, Inc.2026—
+3K253752ARTIS genio floor; ARTIS icono.explore floorSiemens Medical Solutions USA, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199910
200030
200120
200230
200440
201310
201410
201510
201620
201710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202410
202620
FIG. 1 — Lineage of K220432, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K190768ARTIS icono
Summary page 4
…OWB, IZI, JAA, JAK 4. Legally Marketed Primary Predicate Device Trade Name: ARTIS icono (VE20) System 510(k) Clearance K190768 Clearance Date September 12, 2019 Classification Name: Image-intensified fluoroscopic x-ray Sy…
Siemens Medical Solutions USA, Inc.20194

Also cleared against K190768: 3 other devices →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K230950ARTIS icono (VE30A)Siemens Medical Solutions USA, Inc.20233

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 26, 2024Class IIZ-2240-2024Software designA potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed…
Jun 19, 2023Class IIZ-2359-2023Process controlDue to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator…
Jun 19, 2023Class IIZ-2360-2023Process controlDue to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator…
Aug 10, 2022Class IIZ-1647-2022Device DesignSiemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 42 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
  • 18 of the 55 ancestors have a recall of any class.
  • Of its descendants, K230950 (Siemens Medical Solutions USA, Inc., 2023) has gone furthest, with 3 descendants of its own.
  • 3 other devices also cite its predicate K190768.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table…

Read the full statement (192 words)

ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography.

Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions.

ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.

The ARTIS family include also the software option DynaCT with following indications for use:

DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.

DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K220432: ARTIS icono (VE21) System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220432

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026